A comparison of cup-to-disc ratio measurement in normal subjects using optical coherence tomography image analysis of the optic nerve head and stereo fundus biomicroscopy.

Patricia Hrynchak1, Natalie Hutchings, Deborah Jones, Trefford Simpson.   

Abstract

PURPOSE: To determine the relationships between morphometric optic nerve head (ONH) characteristics defined using optical coherence tomography (OCT) and clinical judgments of cup-to-disc (CD) ratios.
METHODS: CD ratios were estimated in 20 subjects by two experienced clinicians using stereo fundus biomicroscopy. Sagittal OCT images were obtained in positions defined by clinicians corresponding to the location of their horizontal and vertical CD ratio estimates. To estimate CD ratios using the OCT images, disc width was defined as the separation of the terminations of the retinal pigment epithelium on each side of the image. OCT cup widths were determined using two criteria: the first measured the separation of cup walls halfway down each side and the second estimated the separation 1/3 down each side. Cups that were less than 150 microm deep were defined as flat for the OCT images. Data were analysed using mean and standard deviation of the differences between clinical estimates and imaging techniques as well as using intraclass correlations (ICC).
RESULTS: Between techniques, the range of ICCs for the 1/2 depth criterion was 0.74-0.90 and for the 1/3 criterion was 0.59-0.81. The limits of agreement for the 1/2 criterion were also better than the 1/3 criterion.
CONCLUSION: Accurate measures of CD ratio may be derived from OCT ONH images. Estimates of clinical CD ratios in normal subjects appear to be made using depth judgments closer to 1/2 of the way down the cup than 1/3.

Entities:  

Mesh:

Year:  2004        PMID: 15491482     DOI: 10.1111/j.1475-1313.2004.00235.x

Source DB:  PubMed          Journal:  Ophthalmic Physiol Opt        ISSN: 0275-5408            Impact factor:   3.117


  3 in total

1.  Three dimensional optic disc visualisation from stereo images via dual registration and ocular media optical correction.

Authors:  J Xu; O Chutatape; C Zheng; P C T Kuan
Journal:  Br J Ophthalmol       Date:  2006-02       Impact factor: 4.638

2.  Automated assessment of the optic nerve head on stereo disc photographs.

Authors:  Juan Xu; Hiroshi Ishikawa; Gadi Wollstein; Richard A Bilonick; Kyung R Sung; Larry Kagemann; Kelly A Townsend; Joel S Schuman
Journal:  Invest Ophthalmol Vis Sci       Date:  2008-03-07       Impact factor: 4.799

3.  Automated volumetric evaluation of stereoscopic disc photography.

Authors:  Juan Xu; Hiroshi Ishikawa; Gadi Wollstein; Richard A Bilonick; Larry Kagemann; Jamie E Craig; David A Mackey; Alex W Hewitt; Joel S Schuman
Journal:  Opt Express       Date:  2010-05-24       Impact factor: 3.894

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.