Evaluation of light-curing units in rural and urban areas.

Maan M Alshaafi1.   

Abstract

OBJECTIVES: To evaluate the distribution of light-curing units (LCU) used in an urban area (Riyadh) and a rural area (Kharj) of Saudi Arabia, and to compare their irradiance values.
METHODS: The study involved three dental centers in urban areas and two in rural areas, all of which were parts of a single healthcare institution providing dental services. The light outputs (power mW) from 140 LCUs were measured by laboratory-grade spectrometry, and the irradiance (mW/cm(2)) was calculated from the tip area of each LCU. The minimum acceptable irradiance outputs for the quartz-tungsten-halogen (QTH) and light-emitting diode (LED) units were set at 300 and 600 mW/cm(2), respectively. The ages of these units and the protocol used to light-cure the resins were also determined.
RESULTS: The total number of LCUs was 140, 112 (78%) in urban areas, and 28 (22%) in rural areas. In rural areas, only 7 of the 22 (32%) QTH units delivered irradiances greater than 300 mW/cm(2) and were therefore considered clinically acceptable, whereas 4 of the 6 (66.7%) LED units delivered values greater than 600 mW/cm(2). In urban centers, 43 of 61 (70.5%) LED units and 25 of 61 (49%) QTH units were considered clinically acceptable. Irradiance values for both QTH (P < 0.01) and LED (P < 0.05) units were significantly better in urban than in rural areas.
CONCLUSIONS: Urban areas had a greater distribution of LCUs than rural areas. Overall, irradiance values were significantly higher in urban areas.

Entities:  

Keywords:  Composite resin; LED; Light-curing unit; QTH; Resin restoration

Year:  2012        PMID: 23960546      PMCID: PMC3729293          DOI: 10.1016/j.sdentj.2012.07.001

Source DB:  PubMed          Journal:  Saudi Dent J        ISSN: 1013-9052


  33 in total

1.  The relationship between leachability of polymerization initiator and degree of conversion of visible light-cured resin.

Authors:  C Moin Jan; Y Nomura; H Urabe; M Okazaki; H Shintani
Journal:  J Biomed Mater Res       Date:  2001

Review 2.  Posterior resin-based composite restorations: clinical recommendations for optimal success.

Authors:  A V Ritter
Journal:  J Esthet Restor Dent       Date:  2001       Impact factor: 2.843

3.  Effects of light-curing time on the cytotoxicity of a restorative resin composite applied to an immortalized odontoblast-cell line.

Authors:  Carlos Alberto de Souza Costa; Josimeri Hebling; Carl Thomas Hanks
Journal:  Oper Dent       Date:  2003 Jul-Aug       Impact factor: 2.440

4.  Light energy matters.

Authors:  Richard B T Price
Journal:  J Can Dent Assoc       Date:  2010       Impact factor: 1.316

5.  Irradiance differences in the violet (405 nm) and blue (460 nm) spectral ranges among dental light-curing units.

Authors:  Richard B T Price; Daniel Labrie; Frederick A Rueggeberg; Christopher M Felix
Journal:  J Esthet Restor Dent       Date:  2010-12       Impact factor: 2.843

6.  Caries prevalence and its relation to water fluoride levels among schoolchildren in Central Province of Saudi Arabia.

Authors:  A M Al Dosari; A H Wyne; E S Akpata; N B Khan
Journal:  Int Dent J       Date:  2004-12       Impact factor: 2.512

7.  Intensity of quartz-tungsten-halogen light-curing units used in private practice in Toronto.

Authors:  Omar El-Mowafy; Wafa El-Badrawy; Donald W Lewis; Babak Shokati; Jaffer Kermalli; Osama Soliman; Avedis Encioiu; Avides Encioiu; Rema Zawi; Fatma Rajwani
Journal:  J Am Dent Assoc       Date:  2005-06       Impact factor: 3.634

8.  Composite resin microhardness: the influence of light curing method, composite shade, and depth of cure.

Authors:  Camila Silveira de Araújo; Marcelo Thome Schein; Cesar Henrique Zanchi; Sinval Adalberto Rodrigues; Flavio Fernando Demarco
Journal:  J Contemp Dent Pract       Date:  2008-05-01

9.  Evaluation of light intensity output of QTH and LED curing devices in various governmental health institutions.

Authors:  Mm Al Shaafi; Am Maawadh; Mq Al Qahtani
Journal:  Oper Dent       Date:  2011-08-11       Impact factor: 2.440

10.  Influence of radiant exposure on contraction stress, degree of conversion and mechanical properties of resin composites.

Authors:  Fernanda C Calheiros; Yoshio Kawano; Jeffrey W Stansbury; Roberto R Braga
Journal:  Dent Mater       Date:  2006-01-19       Impact factor: 5.304

View more
  3 in total

1.  Assessment of the radiant emittance of damaged/contaminated dental light-curing tips by spectrophotometric methods.

Authors:  Abdulrahman A Balhaddad; Isadora Garcia; Fabrício Collares; Cristopher M Felix; Nisha Ganesh; Qoot Alkabashi; Ward Massei; Howard Strassler; Mary Anne Melo
Journal:  Restor Dent Endod       Date:  2020-11-03

2.  Types of polymerisation units and their intensity output in private dental clinics of twin cities in eastern province, KSA; a pilot study.

Authors:  Theeb Alquria; Mohammed Al Gady; Abdul Khabeer; Saqib Ali
Journal:  J Taibah Univ Med Sci       Date:  2018-12-14

3.  Knowledge and Attitude of Dental Clinicians towards Light-Curing Units: A Cross-Sectional Study.

Authors:  Dana Al-Senan; Fatin Ageel; Amani Aldosari; Haifa Maktabi
Journal:  Int J Dent       Date:  2021-06-14
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.