Tissue-mimicking bladder wall phantoms for evaluating acoustic radiation force-optical coherence elastography systems.

O'tega A Ejofodomi1, Vesna Zderic, Jason M Zara.   

Abstract

PURPOSE: Acoustic radiation force-optical coherence elastography (ARF-OCE) systems are novel imaging systems that have the potential to simultaneously quantify and characterize the optical and mechanical properties of in vivo tissues. This article presents the construction of bladder wall phantoms for use in ARF-OCE systems. Mechanical, acoustic, and optical properties are reported and compared to published values for the urinary bladder.
METHODS: The phantom consisted of 0.2000 +/- 0.0089 and 6.0000 +/- 0.2830 microm polystyrene microspheres (Polysciences Inc., Warrington, PA, Catalog Nos. 07304 and 07312), 7.5 +/- 1.5 microm copolymer microspheres composed of acrylonitrile and vinylidene chloride, (Expancel, Duluth, GA, Catalog No. 461 DU 20), and bovine serum albumin within a gelatin matrix. Young's modulus was measured by successive compression of the phantom and obtaining the slope of the resulting force-displacement data. Acoustic measurements were performed using the transmission method. The phantoms were submerged in a water bath and placed between transmitting and receiving 13 mm diameter unfocused transducers operating at a frequency of 3.5 MHz. A MATLAB algorithm to extract the optical scattering coefficient from optical coherence tomography (OCT) images of the phantom was used.
RESULTS: The phantoms possess a Young's modulus of 17.12 +/- 2.72 kPa, a mass density of 1.05 +/- 0.02 g/cm3, an acoustic attenuation coefficient of 0.66 +/- 0.08 dB/cm/MHz, a speed of sound of 1591 +/- 8.76 m/s, and an optical scattering coefficient of 1.80 +/- 0.23 mm(-1). Ultrasound and OCT images of the bladder wall phantom are presented.
CONCLUSIONS: A material that mimics the mechanical, optical, and acoustic properties of healthy bladder wall has been developed. This tissue-mimicking bladder wall phantom was developed as a control tool to investigate the feasibility of using ARF-OCE to detect the mechanical and optical changes that may be indicative of the onset or development of cancer in the urinary bladder. By following the methods used in this article, phantoms matching the optical, acoustic, and mechanical properties of other biological tissues can also be constructed.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20443465     DOI: 10.1118/1.3352686

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Characterization of deformation and physical force in uniform low contrast anatomy and its impact on accuracy of deformable image registration.

Authors:  Raj Varadhan; Taiki Magome; Susanta Hui
Journal:  Med Phys       Date:  2016-01       Impact factor: 4.071

2.  Three-dimensional, distendable bladder phantom for optical coherence tomography and white light cystoscopy.

Authors:  Kristen L Lurie; Gennifer T Smith; Saara A Khan; Joseph C Liao; Audrey K Ellerbee
Journal:  J Biomed Opt       Date:  2014-03       Impact factor: 3.170

3.  Multimodal 3D cancer-mimicking optical phantom.

Authors:  Gennifer T Smith; Kristen L Lurie; Dimitar V Zlatev; Joseph C Liao; Audrey K Ellerbee Bowden
Journal:  Biomed Opt Express       Date:  2016-01-25       Impact factor: 3.732

4.  Multilayered tissue mimicking skin and vessel phantoms with tunable mechanical, optical, and acoustic properties.

Authors:  Alvin I Chen; Max L Balter; Melanie I Chen; Daniel Gross; Sheikh K Alam; Timothy J Maguire; Martin L Yarmush
Journal:  Med Phys       Date:  2016-06       Impact factor: 4.071

5.  Automated, Depth-Resolved Estimation of the Attenuation Coefficient From Optical Coherence Tomography Data.

Authors:  Gennifer T Smith; Nicholas Dwork; Daniel O'Connor; Uzair Sikora; Kristen L Lurie; John M Pauly; Audrey K Ellerbee
Journal:  IEEE Trans Med Imaging       Date:  2015-12       Impact factor: 10.048

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.