Literature DB >> 20432357

Repeatability of T1-quantification in dGEMRIC for three different acquisition techniques: two-dimensional inversion recovery, three-dimensional look locker, and three-dimensional variable flip angle.

Carl Siversson1, Carl-Johan Tiderius, Paul Neuman, Leif Dahlberg, Jonas Svensson.   

Abstract

PURPOSE: To evaluate the repeatability of the dGEMRIC (delayed gadolinium enhanced MRI of cartilage) method in osteoarthritis-prone knee joints for three different T1 quantification techniques: two-dimensional inversion recovery (2D-IR), three-dimensional Look-Locker (3D-LL), and three-dimensional variable flip angle (3D-VFA).
MATERIALS AND METHODS: Nine subjects were examined twice, with a 2-week interval, using all three measurement techniques. Four regions of interest were defined in the central medial and lateral femoral cartilage. The repeatability was evaluated for each measurement technique. For the 3D techniques, the variation between different slices was also evaluated.
RESULTS: Repeatability expressed by root-mean-square coefficient of variation (CV(RMS)) showed similar results for 2D-IR and 3D-LL (5.4-8.4%). For 3D-VFA CV(RMS) was higher (9.3-15.2%). Intraclass correlation coefficient showed both 2D-IR and 3D-LL reliability to be moderate, while 3D-VFA reliability was low. Inter-slice CV(RMS) and ICC was of the same magnitude as the repeatability. No clear differences could be interpreted between the condyles.
CONCLUSION: Both 2D-IR and 3D-LL perform well in generating repeatable dGEMRIC results, while 3D-VFA results are somewhat inferior. Furthermore, repeatability results in this study are similar to previously published results for healthy subjects. Finally, the positioning of the analyzed images is crucial to generate reliable repeatability results. Copyright 2010 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2010        PMID: 20432357     DOI: 10.1002/jmri.22159

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  14 in total

1.  Myocardial T1 mapping with MRI: comparison of look-locker and MOLLI sequences.

Authors:  Marcelo Souto Nacif; Evrim B Turkbey; Neville Gai; Saman Nazarian; Rob J van der Geest; Radwa A Noureldin; Christopher T Sibley; Martin Ugander; Songtao Liu; Andrew E Arai; João A C Lima; David A Bluemke
Journal:  J Magn Reson Imaging       Date:  2011-09-23       Impact factor: 4.813

2.  Impact of uncertainty in longitudinal T1 measurements on quantification of dynamic contrast-enhanced MRI.

Authors:  Madhava P Aryal; Thomas L Chenevert; Yue Cao
Journal:  NMR Biomed       Date:  2016-01-21       Impact factor: 4.044

Review 3.  MR imaging of articular cartilage physiology.

Authors:  Jung-Ah Choi; Garry E Gold
Journal:  Magn Reson Imaging Clin N Am       Date:  2011-05       Impact factor: 2.266

Review 4.  Quantitative MRI of articular cartilage and its clinical applications.

Authors:  Xiaojuan Li; Sharmila Majumdar
Journal:  J Magn Reson Imaging       Date:  2013-10-02       Impact factor: 4.813

5.  Repeatability of magnetic resonance fingerprinting T1 and T2 estimates assessed using the ISMRM/NIST MRI system phantom.

Authors:  Yun Jiang; Dan Ma; Kathryn E Keenan; Karl F Stupic; Vikas Gulani; Mark A Griswold
Journal:  Magn Reson Med       Date:  2016-10-27       Impact factor: 4.668

6.  Measurement of T1 relaxation time of osteochondral specimens using VFA-SWIFT.

Authors:  Mikko J Nissi; Lauri J Lehto; Curtis A Corum; Djaudat Idiyatullin; Jutta M Ellermann; Olli H J Gröhn; Miika T Nieminen
Journal:  Magn Reson Med       Date:  2014-08-08       Impact factor: 4.668

7.  Quantitative MRI techniques of cartilage composition.

Authors:  Stephen J Matzat; Jasper van Tiel; Garry E Gold; Edwin H G Oei
Journal:  Quant Imaging Med Surg       Date:  2013-06

8.  Flip angle profile correction for T₁ and T₂ quantification with look-locker inversion recovery 2D steady-state free precession imaging.

Authors:  Mitchell A Cooper; Thanh D Nguyen; Pascal Spincemaille; Martin R Prince; Jonathan W Weinsaft; Yi Wang
Journal:  Magn Reson Med       Date:  2012-01-31       Impact factor: 4.668

Review 9.  Quantitative radiologic imaging techniques for articular cartilage composition: toward early diagnosis and development of disease-modifying therapeutics for osteoarthritis.

Authors:  Edwin H G Oei; Jasper van Tiel; William H Robinson; Garry E Gold
Journal:  Arthritis Care Res (Hoboken)       Date:  2014-08       Impact factor: 4.794

10.  Reproducibility of 3D delayed gadolinium enhanced MRI of cartilage (dGEMRIC) of the knee at 3.0 T in patients with early stage osteoarthritis.

Authors:  J van Tiel; E E Bron; C J Tiderius; P K Bos; M Reijman; S Klein; J A N Verhaar; G P Krestin; H Weinans; G Kotek; E H G Oei
Journal:  Eur Radiol       Date:  2012-08-11       Impact factor: 5.315