Intensity modulated proton therapy treatment planning using single-field optimization: the impact of monitor unit constraints on plan quality.

X R Zhu1, N Sahoo, X Zhang, D Robertson, H Li, S Choi, A K Lee, M T Gillin.   

Abstract

PURPOSE: To investigate the effect of monitor unit (MU) constraints on the dose distribution created by intensity modulated proton therapy (IMPT) treatment planning using single-field optimization (SFO).
METHODS: Ninety-four energies between 72.5 and 221.8 MeV are available for scanning beam IMPT delivery at our institution. The minimum and maximum MUs for delivering each pencil beam (spot) are 0.005 and 0.04, respectively. These MU constraints are not considered during optimization by the treatment planning system; spots are converted to deliverable MUs during postprocessing. Treatment plans for delivering uniform doses to rectangular volumes with and without MU constraints were generated for different target doses, spot spacings, spread-out Bragg peak (SOBP) widths, and ranges in a homogeneous phantom. Four prostate cancer patients were planned with and without MU constraints using different spot spacings. Rounding errors were analyzed using an in-house software tool.
RESULTS: From the phantom study, the authors have found that both the number of spots that have rounding errors and the magnitude of the distortion of the dose distribution from the ideally optimized distribution increases as the field dose, spot spacing, and range decrease and as the SOBP width increases. From our study of patient plans, it is clear that as the spot spacing decreases the rounding error increases, and the dose coverage of the target volume becomes unacceptable for very small spot spacings.
CONCLUSIONS: Constraints on deliverable MU for each spot could create a significant distortion from the ideally optimized dose distributions for IMPT fields using SFO. To eliminate this problem, the treatment planning system should incorporate the MU constraints in the optimization process and the delivery system should reliably delivery smaller minimum MUs.

Entities:  

Mesh:

Year:  2010        PMID: 20384258      PMCID: PMC2837732          DOI: 10.1118/1.3314073

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  15 in total

1.  Methodologies and tools for proton beam design for lung tumors.

Authors:  M F Moyers; D W Miller; D A Bush; J D Slater
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-04-01       Impact factor: 7.038

2.  Intensity modulation methods for proton radiotherapy.

Authors:  A Lomax
Journal:  Phys Med Biol       Date:  1999-01       Impact factor: 3.609

3.  Direct aperture optimization: a turnkey solution for step-and-shoot IMRT.

Authors:  D M Shepard; M A Earl; X A Li; S Naqvi; C Yu
Journal:  Med Phys       Date:  2002-06       Impact factor: 4.071

4.  Treatment planning and verification of proton therapy using spot scanning: initial experiences.

Authors:  Antony J Lomax; Terence Böhringer; Alessandra Bolsi; Doelf Coray; Frank Emert; Gudrun Goitein; Martin Jermann; Shixiong Lin; Eros Pedroni; Hanspeter Rutz; Otto Stadelmann; Beate Timmermann; Jorn Verwey; Damien C Weber
Journal:  Med Phys       Date:  2004-11       Impact factor: 4.071

5.  A technique for the quantitative evaluation of dose distributions.

Authors:  D A Low; W B Harms; S Mutic; J A Purdy
Journal:  Med Phys       Date:  1998-05       Impact factor: 4.071

6.  Dosimetric comparison of four target alignment methods for prostate cancer radiotherapy.

Authors:  Jennifer C O'Daniel; Lei Dong; Lifei Zhang; Renaud de Crevoisier; He Wang; Andrew K Lee; Rex Cheung; Susan L Tucker; Rajat J Kudchadker; Mark D Bonnen; James D Cox; Radhe Mohan; Deborah A Kuban
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-11-01       Impact factor: 7.038

7.  The 200-MeV proton therapy project at the Paul Scherrer Institute: conceptual design and practical realization.

Authors:  E Pedroni; R Bacher; H Blattmann; T Böhringer; A Coray; A Lomax; S Lin; G Munkel; S Scheib; U Schneider
Journal:  Med Phys       Date:  1995-01       Impact factor: 4.071

8.  Effect of anatomic motion on proton therapy dose distributions in prostate cancer treatment.

Authors:  Xiaodong Zhang; Lei Dong; Andrew K Lee; James D Cox; Deborah A Kuban; Ron X Zhu; Xiaochun Wang; Yupeng Li; Wayne D Newhauser; Michael Gillin; Radhe Mohan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-02-01       Impact factor: 7.038

9.  Spot scanning proton beam therapy for prostate cancer: treatment planning technique and analysis of consequences of rotational and translational alignment errors.

Authors:  Jeff Meyer; Jaques Bluett; Richard Amos; Larry Levy; Seungtaek Choi; Quynh-Nhu Nguyen; X Ron Zhu; Michael Gillin; Andrew Lee
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-01-25       Impact factor: 7.038

10.  Comparison of direct machine parameter optimization versus fluence optimization with sequential sequencing in IMRT of hypopharyngeal carcinoma.

Authors:  Barbara Dobler; Fabian Pohl; Ludwig Bogner; Oliver Koelbl
Journal:  Radiat Oncol       Date:  2007-09-06       Impact factor: 3.481

View more
  27 in total

1.  Verification of proton range, position, and intensity in IMPT with a 3D liquid scintillator detector system.

Authors:  L Archambault; F Poenisch; N Sahoo; D Robertson; A Lee; M T Gillin; R Mohan; S Beddar
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

2.  Robust optimization in IMPT using quadratic objective functions to account for the minimum MU constraint.

Authors:  Jie Shan; Yu An; Martin Bues; Steven E Schild; Wei Liu
Journal:  Med Phys       Date:  2017-12-05       Impact factor: 4.071

3.  Reducing Dose Uncertainty for Spot-Scanning Proton Beam Therapy of Moving Tumors by Optimizing the Spot Delivery Sequence.

Authors:  Heng Li; X Ronald Zhu; Xiaodong Zhang
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-06-18       Impact factor: 7.038

4.  Incorporating deliverable monitor unit constraints into spot intensity optimization in intensity-modulated proton therapy treatment planning.

Authors:  Wenhua Cao; Gino Lim; Xiaoqiang Li; Yupeng Li; X Ronald Zhu; Xiaodong Zhang
Journal:  Phys Med Biol       Date:  2013-07-09       Impact factor: 3.609

5.  Proton energy optimization and reduction for intensity-modulated proton therapy.

Authors:  Wenhua Cao; Gino Lim; Li Liao; Yupeng Li; Shengpeng Jiang; Xiaoqiang Li; Heng Li; Kazumichi Suzuki; X Ronald Zhu; Daniel Gomez; Xiaodong Zhang
Journal:  Phys Med Biol       Date:  2014-10-08       Impact factor: 3.609

6.  Use of treatment log files in spot scanning proton therapy as part of patient-specific quality assurance.

Authors:  Heng Li; Narayan Sahoo; Falk Poenisch; Kazumichi Suzuki; Yupeng Li; Xiaoqiang Li; Xiaodong Zhang; Andrew K Lee; Michael T Gillin; X Ronald Zhu
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

7.  Impact of Spot Size and Spacing on the Quality of Robustly Optimized Intensity Modulated Proton Therapy Plans for Lung Cancer.

Authors:  Chenbin Liu; Steven E Schild; Joe Y Chang; Zhongxing Liao; Shawn Korte; Jiajian Shen; Xiaoning Ding; Yanle Hu; Yixiu Kang; Sameer R Keole; Terence T Sio; William W Wong; Narayan Sahoo; Martin Bues; Wei Liu
Journal:  Int J Radiat Oncol Biol Phys       Date:  2018-02-14       Impact factor: 7.038

8.  Commissioning dose computation models for spot scanning proton beams in water for a commercially available treatment planning system.

Authors:  X R Zhu; F Poenisch; M Lii; G O Sawakuchi; U Titt; M Bues; X Song; X Zhang; Y Li; G Ciangaru; H Li; M B Taylor; K Suzuki; R Mohan; M T Gillin; N Sahoo
Journal:  Med Phys       Date:  2013-04       Impact factor: 4.071

9.  Comparison of Monte Carlo and analytical dose computations for intensity modulated proton therapy.

Authors:  Pablo Yepes; Antony Adair; David Grosshans; Dragan Mirkovic; Falk Poenisch; Uwe Titt; Qianxia Wang; Radhe Mohan
Journal:  Phys Med Biol       Date:  2018-02-09       Impact factor: 3.609

10.  A novel energy layer optimization framework for spot-scanning proton arc therapy.

Authors:  Wenbo Gu; Dan Ruan; Qihui Lyu; Wei Zou; Lei Dong; Ke Sheng
Journal:  Med Phys       Date:  2020-03-13       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.