| Literature DB >> 20160699 |
Subhash C Sharma1, Joseph T Ott, Jamone B Williams, Danny Dickow.
Abstract
It is documented that well-modeled Monte Carlo dose calculation algorithms are more accurate than traditional correction-based algorithms or convolution algorithms at predicting dose distributions delivered to heterogeneous volumes. This increased accuracy has clinical implications for CyberKnife, particularly when comparing dose distributions between the ray-tracing and Monte Carlo algorithms. Differences between ray-tracing and Monte Carlo calculations are exacerbated for highly heterogeneous volumes and small field sizes. In this study, the anthropomorphic thorax phantom from the Radiological Physics Center was used to validate the accuracy of the CyberKnife Monte Carlo dose calculation algorithm. Retrospective comparisons of dose distributions calculated by ray-tracing and Monte Carlo were made for a selection of CyberKnife treatment plans; comparisons were based on target coverage and conformality. For highly heterogeneous cases, such as those involving the lungs, the ray-tracing algorithm consistently overestimated the target dose and coverage. In our sample of lung treatment plans, the average target coverage for ray-tracing calculations was 97.7%, while for Monte Carlo, the average coverage dropped to 69.2%. In each plan comparison, the same beam orientations and monitor units were used for both calculations. Significant changes in conformality were also observed. Isodose prescription lines and subsequent target coverage selected for treatment plans calculated with the ray-tracing algorithm may be different from comparable treatment plans calculated with Monte Carlo, and as such, may have clinical implications for dose prescriptions.Entities:
Mesh:
Year: 2010 PMID: 20160699 PMCID: PMC5719782 DOI: 10.1120/jacmp.v11i1.3142
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Representative image from the Monte Carlo validation treatment plan, using the RPC thorax phantom. The target (PTV) is located in the left lung; 600 cGy is prescribed to the 79% isodose line.
TLD and film results for the Monte Carlo validation plan using the RPC thorax phantom.
|
|
|
|
|---|---|---|
| PTV TLD 1 | 1.00 | 0.92 – 1.02 |
| PTV TLD 2 | 0.96 | 0.92 – 1.02 |
| Left/Right Film | 0/3 mm |
|
| Post/Ant Film | 1/4 mm |
|
| Inf/Sup Film | 1/1 mm |
|
Summary of coverage and conformality comparisons between the ray‐tracing and Monte Carlo treatment plans. Significant degradation is observed for the Monte Carlo plans.
|
|
| ||||||
|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
| |
| Average | 84 | 97.7 | 1.27 | 1.30 | 69.2 | 1.12 | 1.72 |
| σ | 2 | 1.7 | 0.09 | 0.09 | 15.2 | 0.06 | 0.59 |
| %σ | 2.4 | 1.7 | 7.0 | 6.7 | 22.0 | 5.4 | 34.3 |
Comparison between the average isodose prescription lines required for comparable ray‐tracing and Monte Carlo target coverage, and subsequent conformality indexes for the Monte Carlo plans.
|
|
|
|
| |
|---|---|---|---|---|
| Average | 84 | 75 | 1.47 | 1.51 |
| σ | 2 | 4 | 0.22 | 0.21 |
| %σ | 2.4 | 5.3 | 15.0 | 13.9 |
Figure 2Comparison between ray‐tracing (left) and Monte Carlo (right) treatment plans exhibiting the importance of tumor position within the lung as a predictive factor. In each case, the prescription isodose line appears as the thick orange line: 87% for the peripheral target and 84% for the central target.
Figure 3Scatter plot indicating the difference in ray‐tracing and Monte Carlo coverage observed for the sample of 18 lung treatment plans