Design, development, and implementation of the radiological physics center's pelvis and thorax anthropomorphic quality assurance phantoms.

David S Followill1, DeeAnn Radford Evans, Christopher Cherry, Andrea Molineu, Gary Fisher, William F Hanson, Geoffrey S Ibbott.   

Abstract

The Radiological Physics Center (RPC) developed two heterogeneous anthropomorphic quality assurance phantoms for use in verifying the accuracy of radiation delivery: one for intensity-modulated radiation therapy (IMRT) to the pelvis and the other for stereotactic body radiation therapy (SBRT) to the thorax. The purpose of this study was to describe the design and development of these two phantoms and to demonstrate the reproducibility of measurements generated with them. The phantoms were built to simulate actual patient anatomy. They are lightweight and water-fillable, and they contain imageable targets and organs at risk of radiation exposure that are of similar densities to their human counterparts. Dosimetry inserts accommodate radiochromic film for relative dosimetry and thermoluminesent dosimetry capsules for absolute dosimetry. As a part of the commissioning process, each phantom was imaged, treatment plans were developed, and radiation was delivered at least three times. Under these controlled irradiation conditions, the reproducibility of dose delivery to the target TLD in the pelvis and thorax phantoms was 3% and 0.5%, respectively. The reproducibility of radiation-field localization was less than 2.5 mm for both phantoms. Using these anthropomorphic phantoms, pelvic IMRT and thoracic SBRT radiation treatments can be verified with a high level of precision. These phantoms can be used to effectively credential institutions for participation in specific NCI-sponsored clinical trials.

Entities:  

Mesh:

Year:  2007        PMID: 17654910     DOI: 10.1118/1.2737158

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  35 in total

1.  Methodology for determining doses to in-field, out-of-field and partially in-field organs for late effects studies in photon radiotherapy.

Authors:  Rebecca M Howell; Sarah B Scarboro; Phillip J Taddei; Sunil Krishnan; Stephen F Kry; Wayne D Newhauser
Journal:  Phys Med Biol       Date:  2010-11-12       Impact factor: 3.609

2.  Technical note: Heterogeneity dose calculation accuracy in IMRT: study of five commercial treatment planning systems using an anthropomorphic thorax phantom.

Authors:  Scott E Davidson; Richard A Popple; Geoffrey S Ibbott; David S Followill
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

3.  Angular dependence of the nanoDot OSL dosimeter.

Authors:  James R Kerns; Stephen F Kry; Narayan Sahoo; David S Followill; Geoffrey S Ibbott
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

4.  Adaptation and validation of a commercial head phantom for cranial radiosurgery dosimetry end-to-end audit.

Authors:  Alexis Dimitriadis; Antony L Palmer; Russell A S Thomas; Andrew Nisbet; Catharine H Clark
Journal:  Br J Radiol       Date:  2017-05-25       Impact factor: 3.039

5.  On the feasibility of comprehensive high-resolution 3D remote dosimetry.

Authors:  Titania Juang; Ryan Grant; John Adamovics; Geoffrey Ibbott; Mark Oldham
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

6.  An evaluation of three commercially available metal artifact reduction methods for CT imaging.

Authors:  Jessie Y Huang; James R Kerns; Jessica L Nute; Xinming Liu; Peter A Balter; Francesco C Stingo; David S Followill; Dragan Mirkovic; Rebecca M Howell; Stephen F Kry
Journal:  Phys Med Biol       Date:  2015-01-14       Impact factor: 3.609

7.  Approaches to reducing photon dose calculation errors near metal implants.

Authors:  Jessie Y Huang; David S Followill; Rebecca M Howell; Xinming Liu; Dragan Mirkovic; Francesco C Stingo; Stephen F Kry
Journal:  Med Phys       Date:  2016-09       Impact factor: 4.071

8.  Modification and validation of an analytical source model for external beam radiotherapy Monte Carlo dose calculations.

Authors:  Scott E Davidson; Jing Cui; Stephen Kry; Joseph O Deasy; Geoffrey S Ibbott; Milos Vicic; R Allen White; David S Followill
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

9.  Developing and characterizing MR/CT-visible materials used in QA phantoms for MRgRT systems.

Authors:  Angela Steinmann; R Jason Stafford; Gabriel Sawakuchi; Zhifei Wen; Laurence Court; Clifton D Fuller; David Followill
Journal:  Med Phys       Date:  2017-12-21       Impact factor: 4.071

10.  Treatment Planning System Calculation Errors Are Present in Most Imaging and Radiation Oncology Core-Houston Phantom Failures.

Authors:  James R Kerns; Francesco Stingo; David S Followill; Rebecca M Howell; Adam Melancon; Stephen F Kry
Journal:  Int J Radiat Oncol Biol Phys       Date:  2017-04-04       Impact factor: 7.038

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.