Commissioning 6 MV photon beams of a stereotactic radiosurgery system for Monte Carlo treatment planning.

Jun Deng1, C M Ma, Jenny Hai, Ravinder Nath.   

Abstract

The goal of this work is to implement a beam commissioning procedure to generate a multiple source model using a set of standard measurement data for possible Monte Carlo treatment planning in the clinic for a Cyberknife stereotactic radiosurgery system. The required measurement data include the central axis depth dose curve (PDD), the dose profile at dmax(= 1.5 cm) of 60 mm cone at 80 cm source-to-surface distance (SSD), and the cone output factors for cones of 5 mm to 60 mm at 80 cm source-to-axis distance (SAD). The employed dual source model has the same structure as the one that has been studied in our previous work while most of the parameters of each source are extracted from the measurement data rather than the beam phase space. The energy spectra will be extracted from the central axis PDD, the fluence distributions will be deconvoluted from the dose profile at dmax, and the source distributions will be determined from the measured cone output factors. Monte Carlo dose calculations in various water phantoms have been performed to verify the beam commissioning procedure. The agreement between the measurements and the commissioning results was within 2%/1 mm for the central axis PDDs and the dose profiles at various depths when an IC-3 chamber was used and within 2% for the cone output factors for various collimator sizes of 5 to 60 mm. Largest difference (9.5%) was observed for the 7.5 mm cone when an IC-10 chamber was used. The large differences can be attributed to the volumetric averaging effect of the IC-10 chamber, whose dimension is comparable to the field of the small cones. The overall agreement between the measurements and the commissioning results is clinically acceptable, which implies that our commissioning tool is adequate for clinical applications of Monte Carlo dose calculations for the Cyberknife stereotactic radiosurgery system.

Entities:  

Mesh:

Year:  2003        PMID: 14713079     DOI: 10.1118/1.1624753

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  Total scatter factors of small beams: a multidetector and Monte Carlo study.

Authors:  Paolo Francescon; Stefania Cora; Carlo Cavedon
Journal:  Med Phys       Date:  2008-02       Impact factor: 4.071

Review 2.  The evolution of stereotactic radiosurgery in neurosurgical practice.

Authors:  Daniel M Trifiletti; Henry Ruiz-Garcia; Alfredo Quinones-Hinojosa; Rohan Ramakrishna; Jason P Sheehan
Journal:  J Neurooncol       Date:  2021-02-21       Impact factor: 4.130

3.  A measurement-based generalized source model for Monte Carlo dose simulations of CT scans.

Authors:  Xin Ming; Yuanming Feng; Ransheng Liu; Chengwen Yang; Li Zhou; Hezheng Zhai; Jun Deng
Journal:  Phys Med Biol       Date:  2017-01-12       Impact factor: 3.609

4.  Determination of boundaries between ranges of high and low gradient of beam profile.

Authors:  Jacek Wendykier; Marcin Bieniasiewicz; Aleksandra Grządziel; Tadeusz Jedynak; Wiktor Kośniewski; Marta Reudelsdorf; Piotr Wendykier
Journal:  Rep Pract Oncol Radiother       Date:  2016-02-13

Review 5.  Stereotactic radiosurgery alone for multiple brain metastases? A review of clinical and technical issues.

Authors:  Arjun Sahgal; Mark Ruschin; Lijun Ma; Wilko Verbakel; David Larson; Paul D Brown
Journal:  Neuro Oncol       Date:  2017-04-01       Impact factor: 12.300

6.  Using beam profile inflection point in process of treatment planning system verification.

Authors:  Jacek Wendykier; Aleksandra Grządziel; Barbara Bekman; Marcin Bieniasiewicz; Adam Bekman; Piotr Wendykier; Bożena Woźniak; Marta Reudelsdorf; Krzysztof Ślosarek
Journal:  Rep Pract Oncol Radiother       Date:  2021-08-12

7.  Clinical implications of adopting Monte Carlo treatment planning for CyberKnife.

Authors:  Subhash C Sharma; Joseph T Ott; Jamone B Williams; Danny Dickow
Journal:  J Appl Clin Med Phys       Date:  2010-01-29       Impact factor: 2.102

8.  Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom.

Authors:  Yuxi Pan; Ruijie Yang; Jun Li; Xile Zhang; Lu Liu; Junjie Wang
Journal:  J Appl Clin Med Phys       Date:  2018-03-30       Impact factor: 2.102

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.