| Literature DB >> 20098561 |
Jagdish P Bhatnagar1, Josef Novotny, Ajay Niranjan, Douglas Kondziolka, John Flickinger, Dade Lunsford, M Saiful Huq.
Abstract
A new model of Leksell Gamma Knife(R) (LGK), known as Perfexion (LGK PFX), was introduced by Elekta Instrument, AB, Sweden, in 2006. This model has a radically different design from the earlier models U, B, C and 4C. Dosimetric characteristics of LGK PFX, technical differences between LGK PFX and LGK 4C, experience gained with acceptance testing and commissioning of the LGK PFX, and comparison between LGK PFX and LGK 4C are presented in this study. Excellent agreement is found between the manufacturers recommended values of absorbed dose rate, relative output factors for 4 and 8 mm collimators, coincidence of mechanical and dosimetric isocenter, FWHM for beam profiles for various collimators and those reported in the present study. Excellent agreement is also found between the dosimetric characteristics of LGK PFX and LGK 4C for the 4 and 8 mm collimators. Examples of clinical cases treated with LGK PFX and impact of LGK PFX on workflow and dosimetric conformity of treatment planning is also given. The set up and treatment of patients on the LGK PFX is much more efficient since it is a fully automated system. The system also provides more options to generate plan with high dosimetric conformity.Entities:
Keywords: Leksell Gamma Knife PERFEXION; Leksell Gamma Knife dosimetry; Leksell Gamma Knife treatment planning
Year: 2009 PMID: 20098561 PMCID: PMC2807679 DOI: 10.4103/0971-6203.54848
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1LGK PFX radiation unit and collimator system. A) Cross section of the LGK PFX radiation unit. B) Detailed view of sectors; each sector holds 24 60Co sources and can be moved independent of other sectors in desired position to define a collimator size or to block beams. C) Sector position which defines a 4 mm collimator. D) Sector position which defines a 8 mm collimator. E) Sector position which defines a 16 mm collimator. (With permission of Elekta Instrument AB, Stockholm, Sweden.)
Figure 2Comparison of LGK PFX and LGK 4C profiles calculated for 4, 8, 14, 16, and 18 mm collimators for all stereotactic X, Y, Z axes. Profiles for LGK 4C are given as solid lines whereas those for LGK PFX are given as dotted lines
Figure 3Example of using composite shots for the LGK PFX. To shape 50% isodose line and cover irregular target volume (meningioma) different combination of composite shots was used (see combination of eight sectors for each of eight different shots used). Different sector colors indicate different collimator size used: yellow – 4 mm, green – 8 mm, red – 16 mm and blue – blocked beams. Cross hatched sectors represent sectors that were automatically blocked by the treatment planning system to reduce dose to critical structures
Figure 4Example of using dynamic shaping for the LGK PFX. To shape 50% isodose line and cover irregular target volume (meningioma) and minimize dose to brainstem different sectors were blocked for three shots located close to brainstem by dynamic shaping. Please see elongated shape of these three shots and blocking pattern of sectors. Different sector colors indicate different collimator size used: yellow – 4 mm, green – 8 mm and red – 16 mm. Cross hatched sectors represent sectors that were automatically blocked by the treatment planning system to reduce dose to critical structures
Comparison of different technical parameters between LGK PFX and LGK 4C
| 192 Co-60 sources | 201 Co-60 sources |
| Sources move | Sources stationary and fixed in position |
| Source-to-focus distance varies from 374 mm to 433 mm | Source-to-focus distance fixed at 400 mm |
| 8 moving sectors holding 24 sources each | Sources fixed in central radiation unit |
| Collimators fixed - built in the unit | Final external collimators interchangeable |
| No collimator helmets | Four collimator helmets |
| Collimator sizes: 4, 8, 16 mm | Collimator sizes: 4, 8, 14, 18 mm |
| Automatic sector blocking | Individual manual beam blocking |
| Patient Positioning System (PPS) moves patient's whole body in X, Y, Z coordinates, requires patient frame adapter requires no patient frame adapter | Automatic Positioning System (APS) moves only patient's head in X, Y, Z coordinates' |
| No positioning tests required before treatment | Positioning tests required before treatment |
| No manual coordiantes set up | Manual coordinates set up possible |
| Treatment range in X/Y/Z (160/180/220) mm | Treatment range in X/Y/Z (100/120/165) mm |
| Gamma angles 70, 90 and 110 for patient's head fixation | Gamma angles 72, 90, 110, 125 (for APS) for patient's head fixation |
Leksell Gamma, Knife PERFEXION™
Acceptance Tests and Commissioning for the LGK PFX
| Acceptance Tests | |
| System initialization test | Functional |
| System normal treatment functionality test | Functional |
| Clearance test tool QA check | Functional |
| Safety and emergency tests | Functional |
| Uninterrupted power supply test | Functional |
| PPS radial precision test with central diode | 50 μm for 4 mm collimator, 150 μm for 8 and 16 mm collimators |
| PPS radial precision test with two diodes in offset positions | 300 μm for 4 mm |
| Radiological acceptance test (performed by ELEKTA): | |
| - dose rate measurement | Must be at least 3.000 Gy/min |
| - profiles | ± 1.0 mm for FWHM |
| - coincidence of mechanical and dosimetric isocenter | 0.4 mm |
| Commissioning Tests | |
| Absolute dose output at the center of 160 mm diameter polystyrene calibration phantom for 16 mm collimator | Must be at least 3.000 Gy/min ± 3.0% |
| Precision of beam alignment (measurement of X, Y, and Z dose profiles for 4, 8, and 16 mm collimators) | ± 1.0 mm for FWHM |
| Coincidence of radiological focal point with PPS mechanical isocenter measured by film | 0.4 mm |
| Measurement of Relative Output Factors (ROF) | ± 2.0% |
| Measurement of sector output uniformity | ± 1.0% |
| Timer-accuracy | ± 0.1% |
| -linearity | ± 2.0% |
| - constancy | ± 0.1% |
| - end-effect | ± 0.03 min |
| Sector transition dose | 0.05 Gy for 16 mm collimator |
| End-to-end test including all steps of the treatment procedure | Functional |
| Phantom delivered dose ± 4.0% | |
| RPC MD Anderson SRS quality audit | Phantom delivered dose ± 5.0% |
| Profiles offset ± 1.0 mm |
Results of relative output factor (ROF) measurements with EBT film dosimetry and their comparison with Monte Carlo (MC) calculations done by ELEKTA INSTRUMENT AB, Stockholm, Sweden for LGK PFX. Deviations between measured and calculated values are given in parenthesis. Relative output factor for 16 mm collimator size is by definition 1.000
| 8 mm | 0.924 | 0.917 ± 0.014 (−0.8%) |
| 4 mm | 0.805 | 0.810 ± 0.007 (+0.6%) |
Results of FWHM measurements with EBT film dosimetry for LGK PFX
| 4 mm | 6.0 | 6.0 | 5.0 | 6.4 | 6.3 | 5.2 |
| 8 mm | 10.7 | 10.7 | 9.7 | 11.0 | 10.9 | 9.9 |
| 16 mm | 21.2 | 21.3 | 17.1 | 21.7 | 21.7 | 17.4 |
Figure 5Comparison of patient distribution according to diagnosis between (A) LGK 4C and (B) LGK PFX treated during the first year after LGK PFX installation at the University of Pittsburgh Medical Center