Filmless quality assurance of a Leksell Gamma Knife® Icon™.

Borna Maraghechi1, Taeho Kim1, Timothy J Mitchell1, S Murty Goddu1, Joe Dise1, James A Kavanaugh1, Jacqueline E Zoberi1, Sasa Mutic1, Nels C Knutson1.   

Abstract

PURPOSE: The annual quality assurance (QA) of Leksell Gamma Knife® (LGK) systems are typically performed using films. Film is a good candidate for small field dosimetry due to its high spatial resolution and availability. However, there are multiple challenges with using film; film does not provide real-time measurement and requires batch-specific calibration. Our findings show that active detector-based QA can simplify the procedure and save time without loss of accuracy.
METHODS: Annual QA tests for a LGK Icon™ system were performed using both film-based and filmless techniques. Output calibration, relative output factors (ROF), radiation profiles, sector uniformity/source counting, and verification of the unit center point (UCP) and radiation focal point (RFP) coincidence tests were performed. Radiochromic films, two ionization chambers, and a synthetic diamond detector were used for the measurements. Results were compared and verified with the treatment planning system (TPS).
RESULTS: The measured dose rate of the LGK Icon was within 0.4% of the TPS value set at the time of commissioning using an ionization chamber. ROF for the 8 and 4-mm collimators were found to be 0.3% and 1.8% different from TPS values using the MicroDiamond detector and 2.6% and 1.9% different for film, respectively. Excellent agreement was found between TPS and measured dose profiles using the MicroDiamond detector which was within 1%/1 mm vs 2%/1 mm for film. Sector uniformity was found to be within 1% for all eight sectors measured using an ionization chamber. Verification of UCP and RFP coincidence using the MicroDiamond detector and pinprick film test was within 0.3 mm at isocenter for both.
CONCLUSION: The annual QA of a LGK Icon was successfully performed by employing filmless techniques. Comparable results were obtained using radiochromic films. Utilizing active detectors instead of films simplifies the QA process and saves time without loss of accuracy.
© 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  Gamma Knife® Icon™; film dosimetry; quality assurance; radiation profile; small field dosimetry

Mesh:

Substances:

Year:  2020        PMID: 33300664      PMCID: PMC7856498          DOI: 10.1002/acm2.13070

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  15 in total

1.  Verification of source and collimator configuration for Gamma Knife Perfexion using panoramic imaging.

Authors:  Young-Bin Cho; Monique van Prooijen; David A Jaffray; Mohammad K Islam
Journal:  Med Phys       Date:  2010-03       Impact factor: 4.071

2.  Commissioning of the Leksell Gamma Knife® Icon™.

Authors:  Michele Zeverino; Maud Jaccard; David Patin; Nick Ryckx; Maud Marguet; Constantin Tuleasca; Luis Schiappacasse; Jean Bourhis; Marc Levivier; Francois O Bochud; Raphaël Moeckli
Journal:  Med Phys       Date:  2017-01-30       Impact factor: 4.071

3.  Response characterization of EBT-XD radiochromic films in megavoltage photon and electron beams.

Authors:  E Yazmin León-Marroquín; Daniel Mulrow; Arash Darafsheh; Rao Khan
Journal:  Med Phys       Date:  2019-08-01       Impact factor: 4.071

4.  Dosimetric comparison of fractionated radiosurgery plans using frameless Gamma Knife ICON and CyberKnife systems with linear accelerator-based radiosurgery plans for multiple large brain metastases.

Authors:  Eun Young Han; He Wang; Dershan Luo; Jing Li; Xin Wang
Journal:  J Neurosurg       Date:  2019-04-05       Impact factor: 5.115

5.  Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification.

Authors:  Valeria Casanova Borca; Massimo Pasquino; Giuliana Russo; Pierangelo Grosso; Domenico Cante; Piera Sciacero; Giuseppe Girelli; Maria Rosa La Porta; Santi Tofani
Journal:  J Appl Clin Med Phys       Date:  2013-03-04       Impact factor: 2.102

6.  Evaluation of stability of stereotactic space defined by cone-beam CT for the Leksell Gamma Knife Icon.

Authors:  Ismail AlDahlawi; Dheerendra Prasad; Matthew B Podgorsak
Journal:  J Appl Clin Med Phys       Date:  2017-04-17       Impact factor: 2.102

7.  First year experience with newly developed Leksell Gamma Knife Perfexion.

Authors:  Jagdish P Bhatnagar; Josef Novotny; Ajay Niranjan; Douglas Kondziolka; John Flickinger; Dade Lunsford; M Saiful Huq
Journal:  J Med Phys       Date:  2009-07

8.  Fractionated Stereotactic Gamma Knife Radiosurgery for Large Brain Metastases: A Retrospective, Single Center Study.

Authors:  Joo Whan Kim; Hye Ran Park; Jae Meen Lee; Jin Wook Kim; Hyun-Tai Chung; Dong Gyu Kim; Hee-Won Jung; Sun Ha Paek
Journal:  PLoS One       Date:  2016-09-23       Impact factor: 3.240

9.  Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density.

Authors:  Elsa Y León Marroquin; José A Herrera González; Miguel A Camacho López; José E Villarreal Barajas; Olivia A García-Garduño
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

10.  Characterization and validation of an intra-fraction motion management system for masked-based radiosurgery.

Authors:  Nels C Knutson; Brad J Hawkins; Douglas Bollinger; S Murty Goddu; James A Kavanaugh; Lakshmi Santanam; Timothy J Mitchell; Jacqueline E Zoberi; Christina Tsien; Jiayi Huang; Clifford G Robinson; Stephanie M Perkins; Joshua L Dowling; Michael R Chicoine; Keith M Rich; Gavin P Dunn; Sasa Mutic
Journal:  J Appl Clin Med Phys       Date:  2019-05-04       Impact factor: 2.102

View more
  1 in total

1.  Replacing gamma knife beam-profiles on film with point-detector scans.

Authors:  Benedikt Rudek; Kenneth Bernstein; Sunshine Osterman; Tanxia Qu
Journal:  J Appl Clin Med Phys       Date:  2022-01-09       Impact factor: 2.102

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.