Dose and image quality evaluation of a dedicated cone-beam CT system for high-contrast neurologic applications.

Lifeng Yu1, Thomas J Vrieze, Michael R Bruesewitz, James M Kofler, David R DeLone, John F Pallanch, E Paul Lindell, Cynthia H McCollough.   

Abstract

OBJECTIVE: The purpose of our study was to evaluate the dose and image quality performance of a dedicated cone-beam CT (CBCT) scanner in comparison with an MDCT scanner.
MATERIALS AND METHODS: The conventional dose metric, CT dose index (CTDI), is no longer applicable to CBCT scanners. We propose to use two dose metrics, the volume average dose and the mid plane average dose, to quantify the dose performance in a circular cone-beam scan. Under the condition of equal mid plane average dose, we evaluated the image quality of a CBCT scanner and an MDCT scanner, including high-contrast spatial resolution, low-contrast spatial resolution, noise level, CT number uniformity, and CT number accuracy.
RESULTS: For the sinus scanning protocol, the CBCT system had comparable high-contrast resolution and inferior low-contrast resolution to those obtained with the MDCT scanner when the doses were matched (mid plane average dose 9.2 mGy). The CT number uniformity and accuracy were worse on the CBCT scanner. The image artifacts caused by beam hardening and scattering were also much more severe on the CBCT system.
CONCLUSION: With a matched radiation dose, the CBCT system for sinus study has comparable high-contrast resolution and inferior low-contrast resolution relative to the MDCT scanner. Because of the more severe image artifacts on the CBCT system due to the small field of view and the lack of accurate scatter and beam-hardening correction, the utility of the CBCT system for diagnostic tasks related to soft tissue should be carefully assessed.

Mesh:

Year:  2010        PMID: 20093573     DOI: 10.2214/AJR.09.2951

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  14 in total

1.  Low-dose computed tomography of the paranasal sinus and facial skull using a high-pitch dual-source system--first clinical results.

Authors:  Boris Schell; Ralf W Bauer; Thomas Lehnert; J Matthias Kerl; Markus Hambek; Angelika May; Thomas J Vogl; Martin G Mack
Journal:  Eur Radiol       Date:  2010-07-20       Impact factor: 5.315

2.  MR-Imaging of teeth and periodontal apparatus: an experimental study comparing high-resolution MRI with MDCT and CBCT.

Authors:  Chiara Gaudino; Raluca Cosgarea; Sabine Heiland; Réka Csernus; Bruno Beomonte Zobel; Mirko Pham; Ti-Sun Kim; Martin Bendszus; Stefan Rohde
Journal:  Eur Radiol       Date:  2011-07-31       Impact factor: 5.315

Review 3.  Utility of cone-beam computed tomography in the assessment of the porto-spleno-mesenteric venous system.

Authors:  Karunakaravel Karuppasamy
Journal:  Cardiovasc Diagn Ther       Date:  2016-12

4.  Evaluation of different low-dose multidetector CT and cone beam CT protocols in maxillary sinus imaging: part I-an in vitro study.

Authors:  Abeer A Almashraqi; Eman A Ahmed; Nashwa S Mohamed; Imad H Barngkgei; Nader A Elsherbini; Esam S Halboub
Journal:  Dentomaxillofac Radiol       Date:  2017-04-12       Impact factor: 2.419

5.  Impact of prone, supine and oblique patient positioning on CBCT image quality, contrast-to-noise ratio and figure of merit value in the maxillofacial region.

Authors:  Juha Koivisto; Maureen van Eijnatten; Jorma Järnstedt; Kirsi Holli-Helenius; Prasun Dastidar; Jan Wolff
Journal:  Dentomaxillofac Radiol       Date:  2017-04-07       Impact factor: 2.419

6.  Monte Carlo study of the effects of system geometry and antiscatter grids on cone-beam CT scatter distributions.

Authors:  A Sisniega; W Zbijewski; A Badal; I S Kyprianou; J W Stayman; J J Vaquero; J H Siewerdsen
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

7.  Dual-energy cone-beam CT with a flat-panel detector: effect of reconstruction algorithm on material classification.

Authors:  W Zbijewski; G J Gang; J Xu; A S Wang; J W Stayman; K Taguchi; J A Carrino; J H Siewerdsen
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

Review 8.  How I do it: Cone-beam CT during transarterial chemoembolization for liver cancer.

Authors:  Vania Tacher; Alessandro Radaelli; MingDe Lin; Jean-François Geschwind
Journal:  Radiology       Date:  2015-02       Impact factor: 11.105

Review 9.  Therapy of Intermediate-Stage Hepatocellular Carcinoma: Current Evidence and Clinical Practice.

Authors:  Nathan X Chai; Julius Chapiro
Journal:  Semin Intervent Radiol       Date:  2020-12-11       Impact factor: 1.513

10.  A retrospective, semi-quantitative image quality analysis of cone beam computed tomography (CBCT) and MSCT in the diagnosis of distal radius fractures.

Authors:  H Lang; J Neubauer; B Fritz; E M Spira; J Strube; M Langer; E Kotter
Journal:  Eur Radiol       Date:  2016-03-22       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.