Literature DB >> 28306334

Impact of prone, supine and oblique patient positioning on CBCT image quality, contrast-to-noise ratio and figure of merit value in the maxillofacial region.

Juha Koivisto1, Maureen van Eijnatten1, Jorma Järnstedt2, Kirsi Holli-Helenius2, Prasun Dastidar2,3, Jan Wolff1.   

Abstract

OBJECTIVES: To assess the impact of supine, prone and oblique patient imaging positions on the image quality, contrast-to-noise ratio (CNR) and figure of merit (FOM) value in the maxillofacial region using a CBCT scanner. Furthermore, the CBCT supine images were compared with supine multislice CT (MSCT) images.
METHODS: One fresh frozen cadaver head was scanned in prone, supine and oblique imaging positions using a mobile CBCT scanner. MSCT images of the head were acquired in a supine position. Two radiologists graded the CBCT and MSCT images at ten different anatomical sites according to their image quality using a six-point scale. The CNR and FOM values were calculated at two different anatomical sites on the CBCT and MSCT images.
RESULTS: The best image quality was achieved in the prone imaging position for sinus, mandible and maxilla, followed by the supine and oblique imaging positions. 12-mA prone images presented high delineation scores for all anatomical landmarks, except for the ear region (carotid canal), which presented adequate to poor delineation scores for all studied head positions and exposure parameters. The MSCT scanner offered similar image qualities to the 7.5-mA supine images acquired using the mobile CBCT scanner. The prone imaging position offered the best CNR and FOM values on the mobile CBCT scanner.
CONCLUSIONS: Head positioning has an impact on CBCT image quality. The best CBCT image quality can be achieved using the prone and supine imaging positions. The oblique imaging position offers inadequate image quality except in the sinus region.

Entities:  

Keywords:  CBCT; CNR; FOM; MSCT; image quality

Mesh:

Year:  2017        PMID: 28306334      PMCID: PMC5606279          DOI: 10.1259/dmfr.20160418

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  38 in total

1.  3D-imaging of the facial skeleton with an isocentric mobile C-arm system (Siremobil Iso-C3D).

Authors:  M Heiland; D Schulze; G Adam; R Schmelzle
Journal:  Dentomaxillofac Radiol       Date:  2003-01       Impact factor: 2.419

2.  Zygomatic bone: anatomic bases for osseointegrated implant anchorage.

Authors:  Maurício Bruhns Rigolizzo; José Angelo Camilli; Carlos Eduardo Francischone; Carlos Roberto Padovani; Per-Ingvar Brånemark
Journal:  Int J Oral Maxillofac Implants       Date:  2005 May-Jun       Impact factor: 2.804

3.  Cone beam CT: a current overview of devices.

Authors:  A Nemtoi; C Czink; D Haba; A Gahleitner
Journal:  Dentomaxillofac Radiol       Date:  2013       Impact factor: 2.419

4.  Imaging Optimization of Temporal Bones With Cochlear Implant Using a High-resolution Cone Beam CT and the Corresponding Effective Dose.

Authors:  Jing Zou; Juha Koivisto; Jaakko Lähelmä; Antti Aarnisalo; Jan Wolff; Ilmari Pyykkö
Journal:  Ann Otol Rhinol Laryngol       Date:  2014-12-22       Impact factor: 1.547

5.  The effect of decreasing mAs on image quality and patient dose in sinus CT.

Authors:  S A Sohaib; P D Peppercorn; J A Horrocks; M H Keene; G S Kenyon; R H Reznek
Journal:  Br J Radiol       Date:  2001-02       Impact factor: 3.039

Review 6.  Technical aspects of dental CBCT: state of the art.

Authors:  R Pauwels; K Araki; J H Siewerdsen; S S Thongvigitmanee
Journal:  Dentomaxillofac Radiol       Date:  2015       Impact factor: 2.419

7.  The optimization of signal detectability in digital fluoroscopy.

Authors:  R G Zamenhof
Journal:  Med Phys       Date:  1982 Sep-Oct       Impact factor: 4.071

8.  CT arthrography of the wrist using a novel, mobile, dedicated extremity cone-beam CT (CBCT).

Authors:  Seppo K Koskinen; Ville V Haapamäki; Jari Salo; Nina C Lindfors; Mika Kortesniemi; Lauri Seppälä; Kimmo T Mattila
Journal:  Skeletal Radiol       Date:  2012-09-19       Impact factor: 2.199

9.  Dosimetric characterization of a cone-beam O-arm imaging system.

Authors:  Jie Zhang; Victor Weir; Liliosa Fajardo; Jingying Lin; Hsiang Hsiung; E Russell Ritenour
Journal:  J Xray Sci Technol       Date:  2009       Impact factor: 1.535

10.  Comparison of manual and automatic MR-CT registration for radiotherapy of prostate cancer.

Authors:  Anne Sofie Korsager; Jesper Carl; Lasse Riis Østergaard
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

View more
  2 in total

1.  Influence of head positioning during cone-beam CT imaging on the accuracy of virtual 3D models.

Authors:  Maureen van Eijnatten; Jan Wolff; Ruben Pauwels; Kalle Karhu; Ari Hietanen; Henry der Sarkissian; Juha H Koivisto
Journal:  Dentomaxillofac Radiol       Date:  2022-07-13       Impact factor: 3.525

2.  Comparative dosimetry of radiography device, MSCT device and two CBCT devices in the elbow region.

Authors:  Juha Koivisto; Maureen van Eijnatten; John Ludlow; Timo Kiljunen; Xie-Qi Shi; Jan Wolff
Journal:  J Appl Clin Med Phys       Date:  2021-04-03       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.