An in vitro analysis of a carotid artery stent with a protective porous membrane.

Stefan Müller-Hülsbeck1, Erhard J Hüsler, Silvio R Schaffner, Thomas Jahnke, Christoph Glass, Rüdiger Wenke, Martin Heller.   

Abstract

PURPOSE: To prove the effectiveness of a new stent concept with integrated protection (MembraX [MX]) by comparing it with five cerebral protection devices designed for carotid angioplasty in an in vitro model.
MATERIALS AND METHODS: Two simulation series of embolization from carotid angioplasty have been performed. In the first series, polyvinyl-alcohol particles (150-250 microm [small], 355-500 microm [medium], 710-1000 microm [large]; 5 mg each) were injected into a silicone flow model simulating the aortic arch with a carotid bifurcation. The particles were injected proximally to the partially deployed MX stent or one of the following protection devices: Angioguard (AG), FilterWire EX (EX), Trap, Neuroshield (NS), or GuardWire Plus (GW). Particles evading the protection device were caught in a filter at the end of the flow model and weighed. In the second series, human plaque material (8-12 particles; total weight 6.09 +/- 0.01 mg; 500-1500 microm) was injected into the model with the respective devices. MX was compared with the AG, EX, Trap, and NS devices.
RESULTS: MX had the most effective overall filtration performance for polyvinyl alcohol particles in the effluent of the internal carotid artery (ICA; 0.43 mg, 2.9%), compared with NS (0.53 mg, 3.5%), GW (1.10 mg, 7.0%), EX and AG (1.18 and 1.21 mg, respectively; 7.8% and 8.0%), and Trap (1.24 mg, 8.2%). MX performed best for the small particles (2.0% passed particles into ICA; P < .05 compared with all). Human plaque material was retained best in the in vitro model by MX (0.0%), followed by NS (0.8%), EX (1.3%), Trap (2.6%), and AG (4.4%).
CONCLUSIONS: In vitro, none of the tested devices had the ability to prevent embolization completely. Comparing current designs, the MX device captured the highest percentage of the three different particle groups. Tested with human plaque emboli, MX performed effectively in filtering the particles in the ICA.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15525750     DOI: 10.1097/01.RVI.0000141339.79538.0D

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  4 in total

Review 1.  The Angioguard embolic protection device.

Authors:  Gail M Siewiorek; Mark K Eskandari; Ender A Finol
Journal:  Expert Rev Med Devices       Date:  2008-05       Impact factor: 3.166

2.  In vitro performance assessment of distal protection filters: pulsatile flow conditions.

Authors:  Gail M Siewiorek; Mark H Wholey; Ender A Finol
Journal:  J Endovasc Ther       Date:  2009-12       Impact factor: 3.487

3.  Carotid artery stenting using a novel self-expanding braided nickel-titanium stent: feasibility and safety porcine trial.

Authors:  Frank Ahlhelm; Ralf Kaufmann; Dirk Ahlhelm; Mai Fang Ong; Christian Roth; Wolfgang Reith
Journal:  Cardiovasc Intervent Radiol       Date:  2009-06-16       Impact factor: 2.740

4.  The use of covered stents for the endovascular treatment of extracranial internal carotid artery stenosis: a prospective study with a 5-year follow-up.

Authors:  Alex Szólics; László K Sztriha; Péter Szikra; Miklós Szólics; András Palkó; Erika Vörös
Journal:  Eur Radiol       Date:  2009-12-22       Impact factor: 5.315

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.