Wear promoted in the apical third of simulated canals after instrumentation with ProTaper universal system.

Kathrein Tapia da Silva1, Renata Grazziotin-Soares, Orlando Limongi, Luis Eduardo Duarte Irala, Alexandre Azevedo Salles.   

Abstract

OBJECTIVE: This study evaluated the wear in the apical third of simulate canals after preparation with ProTaper Universal Rotary System.
MATERIAL AND METHODS: 24 sets of instruments were used in 24 simulated canals in transparent epoxy resin blocks with degree of curvature of either 20 degrees or 40 degrees. The canals were photographed preoperatively and after preparation of the apical stop with ProTaper F3, F4 and F5 instruments. The initial and final images were exported to Adobe Photoshop software and superimposed to detect the root canal wall differences (in mm) between them, in two points located 1 (A) and 5 (B) mm from the point where the working length was established. Data were subjected to analysis of variance to verify the existence of interaction among the factors: canal curvature, instrument size and curve location. Significant level was set at 5%.
RESULTS: Regardless of the location and the canal curvature, F4 and F5 instruments produced the greatest wear (p<0.05).
CONCLUSIONS: There was a deviation from the original pathway towards the outside of the root curvature in both analyzed points. All instruments produced canal transportation, but the F4 and F5 instruments produced more than the other instruments, and should thus be used with care in curved canals.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19936533      PMCID: PMC4327681          DOI: 10.1590/s1678-77572009000500027

Source DB:  PubMed          Journal:  J Appl Oral Sci        ISSN: 1678-7757            Impact factor:   2.698


INTRODUCTION

Root canal preparation is considered satisfactory when adequate disinfection is obtained. In addition to disinfection, it is important to maintain the continuous tapered shape development of the root canal with the largest diameter in the cervical third. Curved canals pose additional difficulties to endodontic treatment success, such as the maintenance of the original canal shape and the position of the apical foremen. Such difficulties may be attributed to limitations of the endodontic instruments, which may be unable to adapt to the anatomical variations of the root canal system1–3,27. The introduction of nickel-titanium (Ni-Ti) rotary instruments has allowed obtaining less canal transportation and more centered preparations6,7. The Ni-Ti alloy is superelastic, that is, by curving the instrument it misshapes and comes back to its original position when the load is removed. In curved canals, Ni-Ti instruments adapt easily to canal anatomy at the same time that they return almost all the force needed bend them to the canal walls3,19. The root canal system needs to be sufficiently enlarged to allow for removal of debris and microorganisms. It has been shown that apical third preparation with large-sized instruments allows that the irrigating chemical solution reaches this region and has a more effective action. Several authors15,21,22 have demonstrated that when apical preparation is done with size 40 and 45 files there is a significantly greater reduction of the bacterial content from infected canals compared to the use of smaller size instruments. However, the incidence of canal transportation may become more evident when instrument size increases4,8. Wear and canal morphology after instrumentation has been widely investigated in the root canals of extracted teeth and simulated canals prepared in resin blocks9,11,14 Griffiths, et al.9 analyzed superimposed images of simulated canals before and after instrumentation with Quantec SC nickel-titanium rotary instruments and observed that the instruments removed more material from the outer aspect of the curve. The prevalence and severity of wear increased proportionally to the increase of the size of the instruments used to prepare the apical stop. Miranzi, et al.14 compared the alterations occurred in simulated canals with 30° of curvature after preparation with stainless steel and nickel-titanium instruments. The authors evaluated the canal wear in three levels: 2, 4 and 11 mm from the simulated canal apical termination by the superimposition of digital images using computer software. At 2 and 4 mm, there was greater wear towards the outside of the curve when stainless steel instruments were used. The stainless steel instruments also promoted greater wear at 11 mm towards the inner side of the curve. The authors concluded that both types of instruments formed the same areas of wear: towards the outer aspect of the curve in the apical third and to inside the curve in the third middle from the simulated canal. Javaheri and Javaheri11 observed the occurrence of canal transportation and changes in the curvature of mesiobuccal canals from extracted molars after instrumentation with the rotary systems ProTaper Universal, Hero 642 and RaCe. Apical preparation was performed up to a 30 instrument in the three systems. Using a radiographic platform, the authors radiographed the tooth with a stainless steel file inserted in the canal, before and after preparation, and superimposed the two radiographs in the computer. Statistically significant difference in canal transportation was found in the ProTaper Universal group. The authors suggested that the ProTaper Universal System may be implemented in canals with curvatures of 25 to 35° in combination with other systems with greater flexibility and less taper, such as RaCe. Since the apical third of curved canals is the most challenging area for instrumentation, undesirable morphological alterations in this region are common. This study aimed to evaluate the wear promoted by the ProTaper Universal System in the apical third of simulated curved root canals and to determine whether the use of the complete set of instruments for apical stop preparation create inappropriate canal shape because of the increased stiffness of instruments with larger diamaters.

MATERIAL AND METHODS

Twenty-four ProTaper Universal System (Dentsply/Maillefer, Ballaigues, Switzerland) instrument sets, each set containing S1, S2, F1, F2, F3, F4 and F5 instruments, were used. All instruments applied in this study were new and were autoclaved (Cristófoli, São Paulo, SP, Brazil) at 1 atm and 127°C, for 20 min before use. Twenty-four simulated canals were prepared in clear epoxy resin blocks (Odontofix Comércio e Indústria de Materiais Odontológicos, Ribeirão Preto, SP, Brazil). Half of the canals were prepared with 20° curvature and 12 with 40° curvature, all of them presented 5 mm radius and 17 mm length (8 mm corresponded to upper straight portion and 9 mm corresponded to the final curved portion). Each instrument set was used to prepare one simulated canal. The simulated canals were fixed in a flask with the canal curvature always turned to the operator's right side. All clinical procedures were carried out by a single operator. Canals were prepared with the ProTaper instruments coupled to an electric engine (Endo Pro Torque; VK Driller Electrical equipments Ltda., São Paulo, SP, Brazil) at 300 rpm speed and 1 N/cm torque. Instrumentation started with S1 instrument followed by S2 instrument with a paitn brush movements against the canal walls. Subsequently, F1, F2, F3, F4 and F5 instruments were used in the same way until reaching working length (WL). All finishing instruments reached the WL. ProTaper 's manufacturer has suggested this sequence of instruments for clinical use. The WL of the simulated canals was set at 16 mm, corresponding to a distance of 1 mm from the apical foramen. Instrument penetration was limited by the placement of a rubber stopper prior to the start of treatment and was checked during the course of the experiment. The canal patency was permanently maintained with a size 10 stainless steel K-file (Dentsply/Maillefer), irrigation with a sodium lauryl sulfate-based detergent (Tergipol®; Biodinâmica Química e Farmacêutica Ltda., Ibiporã, PR, Brazil), aspiration and inundation at each instrument change. The irrigating solution was flushed with a plastic syringe with an adaptor for a 27 G long anesthesia needle. The simulated canals were positioned in a wooden photographic platform that maintained the same distance between the digital camera and the simulated canal. The canals were photographed before instrumentation and after use of the F3, F4 and F5 instruments. The images were stored in personal computer and transferred to Adobe Photoshop CS2 software version 9.0 (Adobe Systems Inc., San Jose, CA, USA). The image was converted into millimeters by associating the simulated canal original size with the image size in the computer screen. Image pixels were not reduced from the image, preserving the clearness. In Figure 1, the simulated canal image may be observed before the preparation, where gray tone inversion was applied for better visualization of canal anatomy.
FIGURE 1

Simulated canal before preparation with gray tone inversion in Adobe Photoshop CS2 program

Each instrumented canal image (surgical canal) was superimposed on the non-instrumented canal image (anatomic canal). Therefore, layers were created, the image was colored, and its opaque appearance was reduced. The two images were observed by transparency, one over the other. Figure 2 presents a photograph of an instrumented canal superimposed on the photograph of a non-instrumented canal.
FIGURE 2

Image of an instrumented canal superimposed onto the image of a non-instrumented canal

Later, the ruler tool was applied to locate exactly the areas in which wear would be measured in mm towards the outside of the root curvature. These locations were named A and B, which means, respective distances at 1 and 5 mm from the WL (Figure 3). Levels 1 and 5 were chosen for representing the apical third extremities.
FIGURE 3

Ruler tool being used to exactly locate points A and B. In this case, wear is being measured in point A, which may be observed at the right window

Data referring to the amount of wear in the simulated canals (in mm) were analyzed by ANOVA and Tukey's multiple-comparison test at significance level of 5%.

RESULTS

There was no statistically significant interaction (p>0.05) among the three factors (canal curvature, instrument size and curve location). In two-by-two interactions, only the interaction between canal curvature and curve location was significant (p<0.05), which means that, independently on the instrument size, in location A, canals at 20° presented mean wear smaller than canals at 40°, while in point B this difference was not significant (Table 2). In Table 2, the canals with 20° curvature showed no statistically significant difference (p>0.05) in the wear mean in relation to the location. However, in canals with 40° curvature it may be observed that in point A, the mean wear was significantly greater (p<0.05) than in point B. As the instrument size was significant in the interactions among the factors, its main effect was evaluated, that is, wear (in mm). In this way, independently on the curve location and canal curvature, the instruments sizes F4 and F5 presented the greatest wear means, without statistically significant difference from each other (p>0.05). Both instrument sizes produced significantly more wear than instrument F3 (p<0.05) (Table 3).
TABLE 2

Wear means (mm) according to the curve location and the degree of canal curvature

CurvatureLocation
AB
MeanStandard deviationMeanStandard deviation
2O°0.38 Ba0.130.47 Aa0.12
4O°0.80 Aa 0.120.46 Ab 0.18

Means followed by different uppercase letters in columns and different lowercase letters in rows differ significantly (p<0.05; ANOVA and Tukey's test).

TABLE 3

Wear means (mm) according to the instrument size

SizeMean (mm)Standard deviation
F30.40 B 0.20
F40.54 A 0.19
F50.63 A 0.20

Means followed by different letters differ significantly (p<0.05; ANOVA and Tukey's test).

S.D. = standard deviation Means followed by different uppercase letters in columns and different lowercase letters in rows differ significantly (p<0.05; ANOVA and Tukey's test). Means followed by different letters differ significantly (p<0.05; ANOVA and Tukey's test).

DISCUSSION

ProTaper instruments have multiple tapers within the same shaft. Shaping files instruments (Sx, S1 and S2) have progressive taper (increase from the tip towards the handle), which adds flexibility to the middle and apical portions of the files. The Sx, S1 and S2 instruments have tip diameters of 0.19 mm, 0.17 mm and 0.20 mm respectively. On the other hand, the decreasing taper of the Finishing instruments(F1, F2 and F3 with tip diameters of 0.20 mm, 0.25 mm and 0.30 mm respectively) affords a greater mechanical resistance and rigidity to the instruments3,15,16,28. In 2006, the manufacturer made some changes in the design of the instruments S2, F1, F2 and F3. In addition, new instruments were introduced (F4 and F5) and the system was renamed as ProTaper Universal. The S2 instrument had a mild taper modification along the blade. The tip transition angle was removed in all Finishing instruments. The crosssectional design of instrument F3 was reduced. The changes were complemented with the release of instruments F4 and F5 with tip diameter of 0.40 mm and 0.50 mm, respectively. The overall aim of this study was to analyze the wear promoted in the apical third after endodontic preparation with the ProTaperUniversal System, in simulated root canals under controlled laboratory conditions, in addition to determining whether the new instruments (F4 and F5) create inappropriate canal shapes. Simulated canals were chosen in order to standardize the groups, in accordance with previous studies14,25,29. Although simulated canals in acrylic blocks provide highly controlled in vitro conditions, such as standardized characteristics and resin hardness similar to that of root dentin, canal cross-sections differ from that of canals in extracted teeth16,26. In the present study, simulated canals were fabricated according to Pruett et al. method with curvatures of 20 and 40° and curve radius at 5 mm19. These curvatures, with mild and moderate degrees, were standardized as reported in previous studies2,9,25. The apical stop was prepared up to instrument F5 in both canal types in order to produce greater wear, which is desired in teeth with pulp necrosis. In addition, it allows for a more effective canal irrigation and filling. Other authors also used instruments with larger diameter to prepare the apical stop: Peters and Barbakow17 used the Profile System .04 until number 40 and Al-Sudani and Al-Shahrani1 used instruments with large taper (30/.06) in canals with curvature from 15 to 40°. Wear was measured in mm from the anatomical canal wall up to the surgical canal walls. In 72 images that were superimposed on the non-instrumented canal images wear was observed (in points A and B) only outside the curve. In the inner side of the curvature, there was a superpositioning of the anatomical and surgery canal walls. As no wear was observed in the inner side of the curvature, it is assumed that there was a canal transportation. During endodontic preparation of curved canals, the amount of wear must be similar in all walls (buccal, lingual, mesial and distal) in order to avoid the occurrence canal transportation. Finally, it should be kept in mind that the methodology used in the present study does not provide a threedimensional image of the canals, as obtained in studies that used computed tomography for wear analysis10,18. The present analysis was limited to the proximal aspect, which is considered a reliable, low-cost, interesting and easy technique. Adobe Photoshop computer software allows to superpose images and affords different visualizations, reducing the opaque appearance of images23. From the obtained results, there was significant interaction between the curve location and the degree of canal curvature, which means, independently on the instrument size (F3, F4 or F5), in point A, canals of 20° presented wear mean smaller than canals of 40°, but in point B this difference was not significant (Table 2). Such fact has been demonstrated in a previous12 study, where M4 System was used: canals with larger curvature degree showed greater apical wear. A possible explanation for not observing difference in the wear amount in point B, between the two canal types, was the fact that the curvature started at 8 mm from the canal entrance, becoming gradually curved from 7 mm, towards the apex. This may left the canals with similar curvature degree in point B (5 mm). The same wear values in point B were observed, after use of the instruments F3 and F5: 0.57 mm, in the canals of 20°and 40° and 0.33 mm in the canals of 20°and 40°. In canals with 20° curvature, there was no significant difference in the wear mean between points A and B, however in canals of 40°, the wear mean in A was larger than in B, what may be characterized as deviation. In agreement with that, in more curved canals, there is greater wear, and for this reason, caution must be taken during instrumentation to avoid having the instrument stopped during apical preparation5,25. Canal transportation is determined by the flexibility of the preparation instruments, the movement of the instruments in the canal, as well as the length of time the instrument is in contact with the canal wall during preparation20,24. As according to Loizides et al.13, it is of great clinical importance when working with ProTaper Finishing files not leaving the instruments preparing the root canal for longer than 1 second when reaching the required length. This means that finishing files should be immediately removed out of the canal once the working length is achieved. In this way, shaping aberrations in the apical part of the canals can be easily minimized. In both points A and B and regardless of the curvature degree, instruments F4 and F5 produced significantly greater wear than that produced by instrument F3, but did not differ from each other (Table 3). This occurs since, the larger metallic mass instrument quantity, the greater the resistance offered to file adaptation into the canal, because of its smaller flexibility2,3,19.

CONCLUSIONS

From the applied methodology and obtained results, it may be concluded that wear in points A and B occurred only in the outer side of the root curvature, producing canal transportation. In point A, canals with 20° curvature presented less wear than those with 40° curvature. In point B, this difference was not significant. In canals with 20° curvature, there was no significant difference in the mean wear between points A and B. In canals of 40°, the mean wear in point A was significantly greater than in point B. Independently on the location and the canal curvature, instruments F4 and F5, presented the greatest wear, significantly greater than the wear promoted by instrument F3. All instruments produced canal transportation, but the F4 and F5 instruments produced more than the other instruments, and should thus be used with care in curved canals.
TABLE 1

Wear means (mm) according to instrument size, degree of curvature and curve location

CurvatureLocationSizeTotal
F3F4F5
MeanS.D.MeanS.D.MeanS.D.MeanS.D.
2O°A0.230.060.400.100.500.000.380.13
B0.330.060.500.100.570.060.470.12
Total0.280.080.450.100.530.050.420.13
4O°A0.700.100.800.100.900.100.800.12
B0.330.060.470.150.570.250.460.18
Total0.520.210.630.220.730.250.630.23
TotalA0.470.270.600.240.700.230.590.25
B0.330.050.480.120.570.160.460.15
Total0.400.200.540.190.630.200.530.21

S.D. = standard deviation

  28 in total

1.  Canal shapes produced sequentially during instrumentation with Quantec LX rotary nickel-titanium instruments: a study in simulated canals.

Authors:  I T Griffiths; S T Bryant; P M Dummer
Journal:  Int Endod J       Date:  2000-07       Impact factor: 5.264

2.  Three-dimensional analysis of root canal geometry by high-resolution computed tomography.

Authors:  O A Peters; A Laib; P Rüegsegger; F Barbakow
Journal:  J Dent Res       Date:  2000-06       Impact factor: 6.116

3.  Reduction of intracanal bacteria using nickel-titanium rotary instrumentation and various medications.

Authors:  G B Shuping; D Orstavik; A Sigurdsson; M Trope
Journal:  J Endod       Date:  2000-12       Impact factor: 4.171

4.  Progressive versus constant tapered shaft design using NiTi rotary instruments.

Authors:  L Bergmans; J Van Cleynenbreugel; M Beullens; M Wevers; B Van Meerbeek; P Lambrechts
Journal:  Int Endod J       Date:  2003-04       Impact factor: 5.264

5.  Shaping ability of ProTaper nickel-titanium files in simulated resin root canals.

Authors:  F L G Calberson; C A J G Deroose; G M G Hommez; R J G De Moor
Journal:  Int Endod J       Date:  2004-09       Impact factor: 5.264

6.  The shaping effects of three nickel-titanium rotary instruments in simulated S-shaped canals.

Authors:  Y Yoshimine; M Ono; A Akamine
Journal:  J Endod       Date:  2005-05       Impact factor: 4.171

7.  Canal preparation with Hero 642 rotary Ni-Ti instruments compared with stainless steel hand K-file assessed using computed tomography.

Authors:  T Taşdemir; H Aydemir; U Inan; O Unal
Journal:  Int Endod J       Date:  2005-06       Impact factor: 5.264

8.  Determination of root canal curvatures before and after canal preparation (part II): A method based on numeric calculus.

Authors:  D Sonntag; S Stachniss-Carp; C Stachniss; V Stachniss
Journal:  Aust Endod J       Date:  2006-04       Impact factor: 1.659

9.  Analysis of Ni-Ti versus stainless steel instrumentation in resin simulated canals.

Authors:  C L Coleman; T A Svec
Journal:  J Endod       Date:  1997-04       Impact factor: 4.171

10.  The "balanced force" concept for instrumentation of curved canals.

Authors:  J B Roane; C L Sabala; M G Duncanson
Journal:  J Endod       Date:  1985-05       Impact factor: 4.171

View more
  2 in total

1.  Apical transportation associated with ProTaper Universal F1, F2 and F3 instruments in curved canals prepared by undergraduate students.

Authors:  Manoel Brito-Júnior; André Luis Faria-E-Silva; Carla Cristina Camilo; Rodrigo Dantas Pereira; Neilor Mateus Antunes Braga; Manoel Damião Sousa-Neto
Journal:  J Appl Oral Sci       Date:  2014-04       Impact factor: 2.698

2.  Wizard CD Plus and ProTaper Universal: analysis of apical transportation using new software.

Authors:  Daiana Giannastasio; Ricardo Abreu da Rosa; Bernardo Urbanetto Peres; Mirela Sangoi Barreto; Gustavo Nogara Dotto; Milton Carlos Kuga; Jefferson Ricardo Pereira; Marcus Vinícius Reis Só
Journal:  J Appl Oral Sci       Date:  2013 Sep-Oct       Impact factor: 2.698

  2 in total