Literature DB >> 12702124

Progressive versus constant tapered shaft design using NiTi rotary instruments.

L Bergmans1, J Van Cleynenbreugel, M Beullens, M Wevers, B Van Meerbeek, P Lambrechts.   

Abstract

AIM: To evaluate the influence of a progressive versus constant tapered shaft design on canal preparation by NiTi rotary techniques.
METHODOLOGY: A XMCT-scanner and custom-made software were used to nondestructively analyse the mesial canals of 10 extracted mandibular molars in 3D with a spatial resolution of 12.5 microm. Specimens (n = 10 per group) were scanned before and after preparation using ProTaper (progressive tapered) or K3 (constant tapered) files. Numerical values for volume, curvature, dentine removal and centring ratio were obtained in addition to a visual inspection for canal aberrations. Data were analysed by (multiway factorial) anova, Wilcoxon tests and t-tests.
RESULTS: The volume of total dentine removal (mean +/- SD) was 1.21 +/- 0.66 mm(3) (ProTaper) and 1.06 +/- 0.23 mm(3) (K3) (P > 0.05), and the amount of dentine removal at all separate horizontal regions examined was comparable for both groups. The mean linear dentine removal (transportation) was in the range of 8-212 microm (ProTaper) and 4-187 microm (K3). The resultant centring ratio varied from 0.01 to 0.24 (ProTaper) and from 0.01 to 0.17 (K3), whilst different straightening patterns were observed. A centre displacement towards the furcation coronally was most pronounced for the ProTaper group whereas a centre displacement towards the outer side of the curvature more apically was only observed for the K3 group. No severe canal aberrations were found.
CONCLUSIONS: The progressive tapered shaft design of the ProTaper instrument was less influenced by the mid-root curvature than the constant tapered design of the K3 instrument thereby providing a good centred apical preparation. However, ProTaper instruments tended to transport towards the furcation in the coronal region.

Mesh:

Substances:

Year:  2003        PMID: 12702124     DOI: 10.1046/j.1365-2591.2003.00650.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  34 in total

1.  Evaluation of Single File Systems Reciproc, Oneshape, and WaveOne using Cone Beam Computed Tomography -An In Vitro Study.

Authors:  Annil Dhingra; Nidhi Ruhal; Anjali Miglani
Journal:  J Clin Diagn Res       Date:  2015-04-01

2.  Biomechanical preparation in primary molars using manual and three NiTi instruments: a cone-beam-computed tomographic in vitro study.

Authors:  A Manker; M Solanki; A Tripathi; M L Jain
Journal:  Eur Arch Paediatr Dent       Date:  2019-09-05

3.  Canal-centering ability: An endodontic challenge.

Authors:  Deivanayagam Kandaswamy; Nagendrababu Venkateshbabu; Ilango Porkodi; Gali Pradeep
Journal:  J Conserv Dent       Date:  2009-01

4.  Comparative evaluation of cyclic fatigue resistance of four different nickel-titanium rotary files with different cross-sectional designs and alloy properties.

Authors:  Mehmet Emin Kaval; Ismail Davut Capar; Hüseyin Ertas; Bilge Hakan Sen
Journal:  Clin Oral Investig       Date:  2016-07-26       Impact factor: 3.573

5.  Evaluation of apical deviation in root canals instrumented with K3 and ProTaper systems.

Authors:  Mariana Diniz Bisi dos Santos; Marília Fagury Marceliano; Patrícia Rodrigues de Almeida Silva E Souza
Journal:  J Appl Oral Sci       Date:  2006-12       Impact factor: 2.698

6.  Evaluation of Forces Generated on Three Different Rotary File Systems in Apical Third of Root Canal using Finite Element Analysis.

Authors:  Ashish Medha; Suvarna Patil; Upendra Hoshing; Siddhesh Bandekar
Journal:  J Clin Diagn Res       Date:  2013-12-09

7.  An ex vivo comparative analysis on shaping ability of four NiTi rotary endodontic instruments using spiral computed tomography.

Authors:  Nitin Maitin; D Arunagiri; Dexter Brave; Shipra Nangalia Maitin; Sandeep Kaushik; Saumya Roy
Journal:  J Conserv Dent       Date:  2013-05

8.  A comparative evaluation on the cleaning and shaping ability of three nickel titanium rotary instruments using computerized tomography - An ex vivo study.

Authors:  Gopal S Narayan; Sokkalingam Mothilal Venkatesan; C S Karumaran; Rajamani Indira; S Ramachandran; M R Srinivasan
Journal:  Contemp Clin Dent       Date:  2012-09

9.  Change of working length in curved molar root canals after preparation with different rotary nickel-titanium instruments.

Authors:  Pardeep Khurana; Mohan Thomas Nainan; Kiranmeet Kaur Sodhi; Bupinder Kaur Padda
Journal:  J Conserv Dent       Date:  2011-07

10.  Apical extrusion of intracanal debris using two engine driven and step-back instrumentation techniques: an in-vitro study.

Authors:  Alper Kustarci; Neslihan Akdemir; Seyda Herguner Siso; Demet Altunbas
Journal:  Eur J Dent       Date:  2008-10
View more