An MLC calibration method using a detector array.

Thomas A Simon1, Darren Kahler, William E Simon, Christopher Fox, Jonathan Li, Jatinder Palta, Chihray Liu.   

Abstract

PURPOSE: The authors have developed a quantitative calibration method for a multileaf collimator (MLC) which measures individual leaf positions relative to the MLC backup jaw on an Elekta Synergy linear accelerator.
METHODS: The method utilizes a commercially available two-axis detector array (Profiler 2; Sun Nuclear Corporation, Melbourne, FL). To calibrate the MLC bank, its backup jaw is positioned at the central axis and the opposing jaw is retracted to create a half-beam configuration. The position of the backup jaws field edge is then measured with the array to obtain what is termed the radiation defined reference line. The positions of the individual leaf ends relative to this reference line are then inferred by the detector response in the leaf end penumbra. Iteratively adjusting and remeasuring the leaf end positions to within specifications completes the calibration. Using the backup jaw as a reference for the leaf end positions is based on three assumptions: (1) The leading edge of an MLC leaf bank is parallel to its backup jaw's leading edge, (2) the backup jaw position is reproducible, and (3) the measured radiation field edge created by each leaf end is representative of that leaf's position. Data from an electronic portal imaging device (EPID) were used in a similar analysis to check the results obtained with the array.
RESULTS: The relative leaf end positions measured with the array differed from those measured with the EPID by an average of 0.11+/-0.09 mm per leaf. The maximum leaf positional change measured with the Profiler 2 over a 3 month period was 0.51 mm. A leaf positional accuracy of +/-0.4 mm is easily attainable through the iterative calibration process. The method requires an average of 40 min to measure both leaf banks.
CONCLUSIONS: This work demonstrates that the Profiler 2 is an effective tool for efficient and quantitative MLC quality assurance and calibration.

Entities:  

Mesh:

Year:  2009        PMID: 19928080      PMCID: PMC2771712          DOI: 10.1118/1.3218767

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  A fourier analysis of the dose grid resolution required for accurate IMRT fluence map optimization.

Authors:  James F Dempsey; H Edwin Romeijn; Jonathan G Li; Daniel A Low; Jatinder R Palta
Journal:  Med Phys       Date:  2005-02       Impact factor: 4.071

2.  The design and performance characteristics of a multileaf collimator.

Authors:  T J Jordan; P C Williams
Journal:  Phys Med Biol       Date:  1994-02       Impact factor: 3.609

3.  Evaluation of MLC leaf positioning using a scanning liquid ionization chamber EPID.

Authors:  Mohammad Mohammadi; Eva Bezak
Journal:  Phys Med Biol       Date:  2006-12-20       Impact factor: 3.609

4.  Evaluation of two methods of predicting MLC leaf positions using EPID measurements.

Authors:  Laure Parent; Joao Seco; Phil M Evans; David R Dance; Andrew Fielding
Journal:  Med Phys       Date:  2006-09       Impact factor: 4.071

5.  Impact of MLC leaf position errors on simple and complex IMRT plans for head and neck cancer.

Authors:  G Mu; E Ludlum; P Xia
Journal:  Phys Med Biol       Date:  2007-12-13       Impact factor: 3.609

6.  MLC quality assurance using EPID: a fitting technique with subpixel precision.

Authors:  Maria Mamalui-Hunter; Harold Li; Daniel A Low
Journal:  Med Phys       Date:  2008-06       Impact factor: 4.071

7.  A dosimetric calibration method for a double-focused multileaf collimator.

Authors:  M C Lopes; A Chaves; M Capela
Journal:  Med Phys       Date:  2007-09       Impact factor: 4.071

8.  Dosimetric considerations for a multileaf collimator system.

Authors:  J R Palta; D K Yeung; V Frouhar
Journal:  Med Phys       Date:  1996-07       Impact factor: 4.071

9.  Transmission, penumbra and leaf positional accuracy in commissioning and quality assurance program of a multileaf collimator for step-and-shoot IMRT treatments.

Authors:  Massimo Pasquino; Valeria Casanova Borca; Paola Catuzzo; Franca Ozzello; Santi Tofani
Journal:  Tumori       Date:  2006 Nov-Dec
  9 in total
  3 in total

1.  Evaluation of relative transmitted dose for a step and shoot head and neck intensity modulated radiation therapy using a scanning liquid ionization chamber electronic portal imaging device.

Authors:  Mohammad Mohammadi; Eva Bezak
Journal:  J Med Phys       Date:  2012-01

2.  BAER suppression during posterior fossa dural opening.

Authors:  Christopher B Shields; Lisa B E Shields; Yi Dan Jiang; Tom Yao; Yi Ping Zhang; David A Sun
Journal:  Surg Neurol Int       Date:  2015-04-09

3.  Quantifying isocenter measurements to establish clinically meaningful thresholds.

Authors:  Travis R Denton; Lisa B E Shields; Jonathan N Howe; Aaron C Spalding
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.