Dosimetric considerations for a multileaf collimator system.

J R Palta1, D K Yeung, V Frouhar.   

Abstract

A commercial linear accelerator with a factory-fitted multileaf collimator (MLC) was commissioned for clinical use. Measurements made of dosimetric parameters included central axis depth-dose, field-size factors, wedge factors, penumbra, and leaf leakage for the 6-MV and 15-MV photon beams available on this accelerator. The depth-dose characteristics, output factors, and transmission factors were similar to those reported in the literature for a machine by the same manufacturer with a standard treatment head. Because of scalloping, the effective penumbra for the MLC was 3 to 4 mm wider than that for the conventional collimator jaws. The output for the fields shaped by the MLC was generally lower than that for similar fields shaped with Lipowitz's metal (Cerrobend). The magnitude of the difference was field-size dependent and ranged from 0.5% to 4.5% for open shaped fields, increasing to 1% to 5% in the presence of wedges. Further analysis of this observation has shown it to be primarily due to differences in the scattered radiation from the collimator head.

Mesh:

Year:  1996        PMID: 8839416     DOI: 10.1118/1.597678

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  An MLC calibration method using a detector array.

Authors:  Thomas A Simon; Darren Kahler; William E Simon; Christopher Fox; Jonathan Li; Jatinder Palta; Chihray Liu
Journal:  Med Phys       Date:  2009-10       Impact factor: 4.071

2.  Measurement and comparison of head scatter factor for 7 MV unflattened (FFF) and 6 MV flattened photon beam using indigenously designed columnar mini phantom.

Authors:  Sigamani Ashokkumar; Arunai Nambiraj; Sujit Nath Sinha; Girigesh Yadav; Kothanda Raman; Manindra Bhushan; Rajesh Thiyagarajan
Journal:  Rep Pract Oncol Radiother       Date:  2015-03-01

3.  Monitor unit calculations for external photon and electron beams: Report of the AAPM Therapy Physics Committee Task Group No. 71.

Authors:  John P Gibbons; John A Antolak; David S Followill; M Saiful Huq; Eric E Klein; Kwok L Lam; Jatinder R Palta; Donald M Roback; Mark Reid; Faiz M Khan
Journal:  Med Phys       Date:  2014-03       Impact factor: 4.071

4.  Physical and dosimetric characteristic of high-definition multileaf collimator (HDMLC) for SRS and IMRT.

Authors:  Dayananda Shamurailatpam Sharma; Prabhakar M Dongre; Vaibav Mhatre; Malhotra Heigrujam
Journal:  J Appl Clin Med Phys       Date:  2011-04-14       Impact factor: 2.102

5.  Comparison of Head Scatter Factor for 6MV and 10MV flattened (FB) and Unflattened (FFF) Photon Beam using indigenously Designed Columnar Mini Phantom.

Authors:  Sigamani Ashokkumar; N Arunai Nambi Raj; Sujit Nath Sinha; Girigesh Yadav; Rajesh Thiyagarajan; Kothanda Raman; Manindra Bhushan Mishra
Journal:  J Med Phys       Date:  2014-07

6.  Evaluation of dosimetric effect of leaf position in a radiation field of an 80 leaf multileaf collimator fitted to the LINAC head as tertiary collimator.

Authors:  Than S Kehwar; Anup K Bhardwaj; Shiv K Chakarvarti
Journal:  J Appl Clin Med Phys       Date:  2006-08-24       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.