Accuracy of AccuChek glucose measurement in intensive care patients.

Iwan A Meynaar1, Margot van Spreuwel, Peter L Tangkau, Lilian Dawson, Steven Sleeswijk Visser, Lode Rijks, Thea Vliet Vlieland.   

Abstract

OBJECTIVE: To evaluate the accuracy of the AccuChek Inform point-of-care glucose measurement device as compared with central laboratory glucose measurement.
DESIGN: Prospective, observational study.
SETTING: A ten-bed mixed closed format intensive care unit ina 500-bed general hospital. The unit has a computerized insulin protocol aiming for 81 to 135 mg/dL. PATIENTS: All intensive care unit patients were eligible.
INTERVENTIONS: None.
MEASUREMENTS AND MAIN RESULTS: Paired samples (AccuChek glucose in whole blood calibrated to give whole blood results and central laboratory glucose in serum) were taken simultaneously. In 32 critically ill patients, we obtained the following information: mean +/- standard deviation age 71.6 +/- 11.9 yrs; mean Acute Physiology and Chronic Health Evaluation II score at admission 17.8 +/- 6.7; 239 paired samples were taken from arterial catheters. Mean AccuChek whole blood glucose was 126 +/- 36 mg/dL (7.0 +/- 2.0 mmol/L); mean central laboratory serum glucose was 137 +/- 38 mg/dL (7.6 +/- 2.1 mmol/L). Mean difference was 11 mg/dL (0.61 mmol/L) (8%) (95% Confidence Interval 9-13 mg/dL, p < .001). ISO 15197 guideline requires 95% of point-of-care measurements to be within 15 mg/dL margins with reference <75 mg/dL or within 20% if reference is higher. In total, 216 (90.4%) of AccuChek measurements were within ISO 15197 margins. Because AccuChek was calibrated to give whole blood results, we calculated a correction factor of 1.086 from the two mean values to correct whole blood AccuChek into serum-like results. This is almost the same as the correction factor of 1.080 given by Roche Diagnostics. By multiplying AccuChek whole blood results with 1.086, 225 (94.1%) of results were within the ISO 15197 margins. Hematocrit did not influence AccuChek results in the 0.20 to 0.44 range. Beyond this range, there were not enough data to draw conclusions.
CONCLUSIONS: In critically ill patients, the accuracy of AccuChek glucose measurement calibrated to give serum-like results with blood samples derived from arterial catheters is acceptable but falls short by about 1% of complying with the ISO 15197 guideline.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19885987     DOI: 10.1097/ccm.0b013e3181a564fe

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  16 in total

Review 1.  Blood glucose measurements in critically ill patients.

Authors:  Tom Van Herpe; Dieter Mesotten
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

2.  Assessing inpatient glycemic control: what are the next steps?

Authors:  Curtiss B Cook; Kay E Wellik; Gail L Kongable; Jianfen Shu
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

Review 3.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

4.  Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus.

Authors:  David B Sacks; Mark Arnold; George L Bakris; David E Bruns; Andrea Rita Horvath; M Sue Kirkman; Ake Lernmark; Boyd E Metzger; David M Nathan
Journal:  Diabetes Care       Date:  2011-06       Impact factor: 19.112

Review 5.  Glucometrics and Insulinometrics.

Authors:  Bithika M Thompson; Curtiss B Cook
Journal:  Curr Diab Rep       Date:  2017-10-23       Impact factor: 4.810

6.  Hypoglycemia at admission is associated with inhospital mortality in Ugandan patients with severe sepsis.

Authors:  Richard Ssekitoleko; Shevin T Jacob; Patrick Banura; Relana Pinkerton; David B Meya; Steven J Reynolds; Nathan Kenya-Mugisha; Harriet Mayanja-Kizza; Rose Muhindo; Sanjay Bhagani; W Michael Scheld; Christopher C Moore
Journal:  Crit Care Med       Date:  2011-10       Impact factor: 7.598

Review 7.  Analytical and clinical performance of blood glucose monitors.

Authors:  Suzanne Austin Boren; William L Clarke
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

8.  Accuracy of bedside glucometry in critically ill children with peripheral hypoperfusion.

Authors:  Anant Khositseth; Onsuthi Pharadornuwat; Rojjanee Lertbunrian; Nattachai Anantasit; Jarin Vaewpanich
Journal:  J Clin Monit Comput       Date:  2014-05-07       Impact factor: 2.502

9.  Statistical transformation and the interpretation of inpatient glucose control data from the intensive care unit.

Authors:  George E Saulnier; Janna C Castro; Curtiss B Cook
Journal:  J Diabetes Sci Technol       Date:  2014-02-27

10.  Accuracy of Blood Glucose Measurement and Blood Glucose Targets.

Authors:  Gert-Jan Eerdekens; Steffen Rex; Dieter Mesotten
Journal:  J Diabetes Sci Technol       Date:  2020-02-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.