Radiation exposure from fluoroscopy during fixation of hip fracture and fracture of ankle: Effect of surgical experience.

Rajesh Botchu1, Kassetti Ravikumar.   

Abstract

BACKGROUND: Over the years, there has been a tremendous increase in the use of fluoroscopy in orthopaedics. The risk of contracting cancer is significantly higher for an orthopedic surgeon. Hip and spine surgeries account for 99% of the total radiation dose. The amount of radiation to patients and operating surgeon depends on the position of the patient and the type of protection used during the surgery. A retrospective study to assess the influence of the radiation exposure of the operating surgeon during fluoroscopically assisted fixation of fractures of neck of femur (dynamic hip screw) and ankle (Weber B) was performed at a district general hospital in the United Kingdom.
MATERIALS AND METHODS: Sixty patients with undisplaced intertrochanteric fracture were included in the hip group, and 60 patients with isolated fracture of lateral malleolus without communition were included in the ankle group. The hip and ankle groups were further divided into subgroups of 20 patients each depending on the operative experience of the operating surgeon. All patients had fluoroscopically assisted fixation of fracture by the same approach and technique. The radiation dose and screening time of each group were recorded and analyzed.
RESULTS: The radiation dose and screening time during fluoroscopically assisted fixation of fracture neck of femur were significantly high with surgeons and trainees with less than 3 years of surgical experience in comparison with surgeons with more than 10 years of experience. The radiation dose and screening time during fluoroscopically assisted fixation of Weber B fracture of ankle were relatively independent of operating surgeon's surgical experience.
CONCLUSION: The experience of operating surgeon is one of the important factors affecting screening time and radiation dose during fluoroscopically assisted fixation of fracture neck of femur. The use of snapshot pulsed fluoroscopy and involvement of senior surgeons could significantly reduce the radiation dose and screening time.

Entities:  

Keywords:  Experience; fixation; fracture; radiation; surgeon

Year:  2008        PMID: 19753238      PMCID: PMC2740341          DOI: 10.4103/0019-5413.43398

Source DB:  PubMed          Journal:  Indian J Orthop        ISSN: 0019-5413            Impact factor:   1.251


INTRODUCTION

The use of fluoroscopy has increased tremendously in field of orthopedics.12 The risk of an orthopedic surgeon contracting cancer is significantly higher than that of a nonorthopedic professional and eight times more than that of an unexposed worker.3 The hip and spine surgeries account for most of the total radiation exposure.1 The amount of radiation exposed to individual surgeons is influenced by many factors.45 These include the type and difficulty of the surgical procedure, patient's position, and radiation protection measures used. The role of experience of the surgeon involved in the radiation exposure has not been studied well.346 The aim of the study was to assess the effect of surgical experience on the radiation exposure and screening time during fluoroscopically assisted fixation of undisplaced fractures of neck of femur (dynamic hip screw) and ankle (Weber B).

MATERIALS AND METHODS

A retrospective study was conducted at a district general hospital in the United Kingdom from 2003 to 2005. The surgeons were divided into three different groups based on their years of operative experience [Table 1]. Surgeons and trainees with operative experience of less than 3 years were grouped together in Group 1; surgeons and trainees with surgical experience between 3 and 10 years were put in Group 2, and surgeons with more than 10 years were in Group 3.
Table 1

Surgical experience

Group I< 3 years
Group II3–10 years
Group III> 10 years
Surgical experience A total of 700 patients with fixation of fracture of neck of femur underwent fluoroscopically assisted fixation of hip (dynamic hip screws), and 225 patients with ankle fracture underwent fluoroscopically assisted fixation during 2003–2005. Only patients with undisplaced intertrochanteric fracture were included in the hip group. Patients with multiple fractures, communited fracture, compound fracture, and significant morbidity were excluded. Patients with a fracture of lateral malleolus without communition was included in the ankle group. Patients with polytrauma, compound fracture, medial malleolar fracture, and communition were excluded. In each subgroup of the hip group, 20 consecutive patients who satisfied the inclusion criteria and were operated by the appropriate surgeon were included (20 patients operated by Group 1, 20 patients operated by Group 2, and 20 patients operated by Group 3). The age and sex of the patients were comparable in the hip and ankle group. Similarly in the ankle group, in each subgroup, 20 consecutive patients who satisfied the inclusion criteria and were operated by the appropriate surgeon were included (20 patients operated by Group 1, 20 patients operated by Group 2, and 20 patients operated by Group 3). All screening was performed using a mobile C arm fluoroscopy unit (Philips Libra, Philips Medical Systems, Best, The Netherlands). The total filtration of the x-ray tube was 6.35 mm; the focus to detector distance was 99.5 cm, and the diameter of the routinely used input field of view was 17 cm. The dose of the patient was routinely monitored and recorded in accordance with the requirements of Regulation 7 of The Ionising Radiation (Medical Exposure) Regulations [IR(ME)R] 2000 by means of a dose area product (DAP) meter permanently built into the x-ray tube housing.7 DAP is the currently accepted method of assessing the radiation exposure in complex diagnostic x-ray procedures and is measured in Gy cm2. The DAP meter was calibrated following the procedure described in the National Protocol for Patient Dose Measurement in Diagnostic Radiology.8 In addition to the DAP value, the overall fluoroscopic screening time (minutes) was recorded for each patient. All the patients with fracture neck of femur underwent dynamic hip screw fixation by lateral approach. The patients with Weber B fracture of ankle underwent fixation of lateral malleolus using a one-third tubular plate using a lateral approach.

RESULTS

The radiation dose and screening times of the various groups were compared using Student t test, and P values were calculated.

Fixation of fracture of neck of femur

The radiation dose (DAP) and screening time during fixation of a hip fracture were almost three times more with surgeons and trainees with less than 3 years of operative experience (Group I) when compared with that of the surgeons with more than 10 years of surgical experience (Group III) [Table 2] (P = 0.0005).
Table 2

Radiation dose and screening time for fixation of Tronzo 2 fracture neck of femur

Screening time (min)Dose area product (Gy cm2)


AverageMedianAverageMedian
Group I1.18821.231.62151.4
Group II0.66570.571.1111.18
Group III0.4920.360.59320.56
Radiation dose and screening time for fixation of Tronzo 2 fracture neck of femur

Fixation of fracture of ankle

The radiation dose and screening time during fixation of ankle were not significantly affected by surgical experience of the operating surgeon, and they were comparable among all the three groups [Table 3] (P = 0.1015).
Table 3

Radiation dose and screening time for fixation of Weber B fracture of ankle

Screening time (min)Dose area product (Gy cm2)


AverageMedianAverageMedian
Group I0.210.210.19330.19
Group II0.20950.230.15180.144
Group III0.15180.1450.12370.119
Radiation dose and screening time for fixation of Weber B fracture of ankle

DISCUSSION

The image intensifier can be considered as the backbone for intraoperative fracture fixation.25 Mastrangelo et al. had shown a significant increase in the incidence of cancers in orthopedic workers.6 The incidence was four times higher in an orthopedic surgeon than in a nonorthopedic professional and eight times more than that in an unexposed worker. He also emphasized the significance of radiation protection measures and warned against complacency. Radiation exposure can be decreased significantly by radiation protection measures and by increasing the distance between the source of radiation and the subject.12 Mehlman and DiPasquale had demonstrated that working more than 36 inches from the beam significantly decreases the radiation dose.9 Giannoudis et al. had demonstrated a significant correlation between the radiation dose administered and experience of the radiographer.3 The radiation dose of all the subjects included in the study conducted by Smith et al. was below the recommended limits.10 Crawley and Rogers in their study found the median radiation dose for dynamic hip screw fixation and ORIF ankle to be 2.58 Gy cm2 and 0.39 Gy cm2, respectively, whereas the median screening time was 0.90 and 0.55 min.1 In this study, overall patient doses as monitored by DAP were reassuringly well below these published values. The radiation exposure during fixation of fractures of hip was significantly higher when they were performed by surgeons with fewer than 3 years of operative experience, and this was statistically significant (P = 0.0005 ) [Figure 1]. In patients who underwent fixation for fractures of ankle, the radiation exposure and screening time was not significantly dependent on the surgical experience of the operating surgeon (P = 0.1015) [Figure 2].
Figure 1

Screening times and radiation dose in fixation of fracture neck of femur

Figure 2

Screening times and radiation dose in fixation of ankle fracture (Weber B)

Screening times and radiation dose in fixation of fracture neck of femur Screening times and radiation dose in fixation of ankle fracture (Weber B)

CONCLUSIONS

This study shows that the experience and training of the operating surgeon is one of the most important factors determining the radiation exposure to patients in fixation of hip fracture. The involvement of senior surgeons in hip fracture fixation is likely to reduce the radiation dose to patients and surgeons. We endorse the practice of using snap shot pulsed fluoroscopy, last image hold, and good set up geometry as a means of dose optimization. The practice of continuous screening during fixation of fractures of neck of femur and ankle must be discouraged.
  8 in total

1.  Occupational exposure from common fluoroscopic projections used in orthopaedic surgery.

Authors:  Nicholas Theocharopoulos; Kostas Perisinakis; John Damilakis; George Papadokostakis; Alexander Hadjipavlou; Nicholas Gourtsoyiannis
Journal:  J Bone Joint Surg Am       Date:  2003-09       Impact factor: 5.284

2.  Ionising radiation: are orthopaedic surgeons at risk?

Authors:  G L Smith; T W Briggs; C B Lavy; H Nordeen
Journal:  Ann R Coll Surg Engl       Date:  1992-09       Impact factor: 1.891

3.  Risks of radiation exposure to orthopaedic surgeons.

Authors:  P J O'Rourke; S Crerand; P Harrington; M Casey; W Quinlan
Journal:  J R Coll Surg Edinb       Date:  1996-02

4.  Radiation exposure to the orthopaedic surgical team during fluoroscopy: "how far away is far enough?".

Authors:  C T Mehlman; T G DiPasquale
Journal:  J Orthop Trauma       Date:  1997-08       Impact factor: 2.512

5.  Increased cancer risk among surgeons in an orthopaedic hospital.

Authors:  Giuseppe Mastrangelo; Ugo Fedeli; Emanuela Fadda; Angelo Giovanazzi; Luca Scoizzato; Bruno Saia
Journal:  Occup Med (Lond)       Date:  2005-09       Impact factor: 1.611

6.  Radiation exposure to an orthopedic surgeon.

Authors:  T P Barry
Journal:  Clin Orthop Relat Res       Date:  1984 Jan-Feb       Impact factor: 4.176

7.  Dose-area product measurements in a range of common orthopaedic procedures and their possible use in establishing local diagnostic reference levels.

Authors:  M T Crawley; A T Rogers
Journal:  Br J Radiol       Date:  2000-07       Impact factor: 3.039

8.  Ionising radiation during internal fixation of extracapsular neck of femur fractures.

Authors:  P V Giannoudis; J McGuigan; D L Shaw
Journal:  Injury       Date:  1998-07       Impact factor: 2.586

  8 in total
  13 in total

1.  Occupational radiation exposure from C arm fluoroscopy during common orthopaedic surgical procedures and its prevention.

Authors:  Anupam Mahajan; Sumant Samuel; Atul K Saran; M K Mahajan; M K Mam
Journal:  J Clin Diagn Res       Date:  2015-03-01

2.  Fluoroscopy time during uncomplicated unilateral ureteroscopy for urolithiasis decreases with urology resident experience.

Authors:  Lancaster R Weld; Uzoamaka O Nwoye; Richard B Knight; Timothy S Baumgartner; James S Ebertowski; Matthew T Stringer; Matthew C Kasprenski; Kyle J Weld
Journal:  World J Urol       Date:  2014-02-23       Impact factor: 4.226

3.  Underestimation of fluoroscopic exposure among orthopedic residents.

Authors:  Kelly D Carmichael; Lattisha L Bilbrew
Journal:  Skeletal Radiol       Date:  2019-08-05       Impact factor: 2.199

4.  Ionising radiation exposure in paediatric trauma.

Authors:  N G Patel; A M Mohamed; G Cooper; I McFadyen
Journal:  Ann R Coll Surg Engl       Date:  2014-04       Impact factor: 1.891

5.  The effect of surgical experience on the amount of radiation exposure from fluoroscopy during dynamic hip screw fixation.

Authors:  C Quah; R Mehta; F S Shivji; S Hassan; J Chandrasenan; C G Moran; D P Forward
Journal:  Ann R Coll Surg Engl       Date:  2016-08-23       Impact factor: 1.891

6.  Factors influencing radiation exposure during internal fixation of hip fractures.

Authors:  G A Kelly; F E Rowan; C Hurson
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-04-10

7.  Radiation exposure from fluoroscopy during tibia fracture intramedullary nailing - The effect of surgical experience.

Authors:  Jessica Jobson; Ahmed Saad; Ibrahim Jaly; Ravneet Singh; Khalid Baloch; Rajesh Botchu
Journal:  J Clin Orthop Trauma       Date:  2022-09-28

8.  Influence of surgeon's experience on fluoroscopy time during endourological interventions.

Authors:  M Ritter; F Siegel; P Krombach; A Martinschek; C Weiss; A Häcker; A E Pelzer
Journal:  World J Urol       Date:  2012-08-12       Impact factor: 4.226

9.  A New Preoperative Planning Technique Can Reduce Radiation Exposure During the Performance of Medial Opening-Wedge High Tibial Osteotomy.

Authors:  Elad Spitzer; Joseph J Ruzbarsky; John B Doyle; Kaitlyn L Yin; Robert G Marx
Journal:  HSS J       Date:  2017-12-26

10.  Analysis of radiation risk to patients from intra-operative use of the mobile X-ray system (C-arm).

Authors:  Yang-Sub Lee; Hae-Kag Lee; Jae-Hwan Cho; Ham-Gyum Kim
Journal:  J Res Med Sci       Date:  2015-01       Impact factor: 1.852

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.