Accuracy of torque-limiting devices for mini-implant removal: an in vitro study.

A Pauls1, M Nienkemper, D Drescher.   

Abstract

INTRODUCTION: Mini-implants, due to their potential for osseointegration, are exposed to torque levels that may cause them to fracture during removal. Thus, it is advisable to control the torque levels applied during mini-implant removal.
MATERIALS AND METHODS: A torque sensor with strain gauges was used to analyze torque-limiting devices for their accuracy in reverse (counterclockwise) operation. Eight devices were tested in this manner, including a group of hand-operated drivers (n=3), a group of battery-operated drivers (n=4), and a mains-operated surgical unit (n=1). Each device was analyzed eight times at each of the various torque levels. Shapiro-Wilk, Kruskal-Wallis H-, and Mann-Whitney U-tests were used to analyze the results.
RESULTS: Most of the various devices revealed significant differences upon comparison. The accuracy of torque control offered by the three hand drivers was clinically acceptable. As two of the four battery-operated drivers did not feature torque limitation in reverse mode, they did not prevent high torque levels from occurring. Likewise, some of the maximum torque levels observed in conjunction with the other two battery-operated drivers and the mains-operated surgical unit exceeded considerably the clinically recommended range of 10-25 Ncm.
CONCLUSION: Although miniscrews can be removed successfully with hand-operated drivers while limiting torque, we advise against the use of battery-operated drivers or mains-operated surgical units not offering torque limitation in reverse mode.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23649278     DOI: 10.1007/s00056-013-0141-4

Source DB:  PubMed          Journal:  J Orofac Orthop        ISSN: 1434-5293            Impact factor:   1.938


  30 in total

1.  In vitro evaluation of insertion and removal torques of orthodontic mini-implants.

Authors:  M M Pithon; M G Nojima; L I Nojima
Journal:  Int J Oral Maxillofac Surg       Date:  2010-11-02       Impact factor: 2.789

2.  Physical properties of root cementum: part 8. Volumetric analysis of root resorption craters after application of controlled intrusive light and heavy orthodontic forces: a microcomputed tomography scan study.

Authors:  Debora Alvares Harris; Allan S Jones; M Ali Darendeliler
Journal:  Am J Orthod Dentofacial Orthop       Date:  2006-11       Impact factor: 2.650

3.  Load-related implant reaction of mini-implants used for orthodontic anchorage.

Authors:  André Büchter; Dirk Wiechmann; Stefan Koerdt; Hans Peter Wiesmann; Josef Piffko; Ulrich Meyer
Journal:  Clin Oral Implants Res       Date:  2005-08       Impact factor: 5.977

4.  Parameters affecting primary stability of orthodontic mini-implants.

Authors:  Benedict Wilmes; Carsten Rademacher; Gudrun Olthoff; Dieter Drescher
Journal:  J Orofac Orthop       Date:  2006-05       Impact factor: 1.938

5.  Removal torque values of surface-treated mini-implants after loading.

Authors:  Seong-Hun Kim; Jae-Hee Cho; Kyu-Rhim Chung; Yoon-Ah Kook; Gerald Nelson
Journal:  Am J Orthod Dentofacial Orthop       Date:  2008-07       Impact factor: 2.650

6.  Mechanical loading of orthodontic miniscrews - significance and problems: an experimental study.

Authors:  Claudia Reicheneder; Kurt Rottner; Ivan Bokan; Ronald Mai; Günter Lauer; Gerd Richter; Tomasz Gedrange; Peter Proff
Journal:  Biomed Tech (Berl)       Date:  2008-10       Impact factor: 1.411

7.  Assessment of masticatory performance, bite force, orthodontic treatment need and orofacial dysfunction in children and adolescents.

Authors:  Maria Carolina Salomé Marquezin; Fernanda Yukie Kobayashi; Ana Bheatriz Marangoni Montes; Maria Beatriz Duarte Gavião; Paula Midori Castelo
Journal:  Arch Oral Biol       Date:  2012-08-29       Impact factor: 2.633

8.  Miniscrews as orthodontic anchorage: a preliminary report.

Authors:  A Costa; M Raffainl; B Melsen
Journal:  Int J Adult Orthodon Orthognath Surg       Date:  1998

9.  Influence of loading frequency on implant failure under cyclic fatigue conditions.

Authors:  Matthias Karl; J Robert Kelly
Journal:  Dent Mater       Date:  2009-07-29       Impact factor: 5.304

10.  Effects of predrilling on the osseointegration potential of mini-implants.

Authors:  Yun-Seob Shin; Hyo-Won Ahn; Young-Guk Park; Seong-Hun Kim; Kyu-Rhim Chung; Il-Sik Cho; Gerald Nelson
Journal:  Angle Orthod       Date:  2012-04-12       Impact factor: 2.079

View more
  1 in total

Review 1.  Revisiting the Complications of Orthodontic Miniscrew.

Authors:  Van Mai Truong; Soyeon Kim; Jaeheon Kim; Joo Won Lee; Young-Seok Park
Journal:  Biomed Res Int       Date:  2022-08-01       Impact factor: 3.246

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.