Fixed gantry tomosynthesis system for radiation therapy image guidance based on a multiple source x-ray tube with carbon nanotube cathodes.

Jonathan S Maltz1, Frank Sprenger, Jens Fuerst, Ajay Paidi, Franz Fadler, Ali R Bani-Hashemi.   

Abstract

The authors present the design and simulation of an imaging system that employs a compact multiple source x-ray tube to produce a tomosynthesis image from a set of projections obtained at a single tube position. The electron sources within the tube are realized using cold cathode carbon nanotube technology. The primary intended application is tomosynthesis-based 3D image guidance during external beam radiation therapy. The tube, which is attached to the gantry of a medical linear accelerator (linac) immediately below the multileaf collimator, operates within the voltage range of 80-160 kVp and contains a total of 52 sources that are arranged in a rectilinear array. This configuration allows for the acquisition of tomographic projections from multiple angles without any need to rotate the linac gantry. The x-ray images are captured by the same amorphous silicon flat panel detector employed for portal imaging on contemporary linacs. The field of view (FOV) of the system corresponds to that part of the volume that is sampled by rays from all sources. The present tube and detector configuration provides an 8 x 8 cm2 FOV in the plane of the linac isocenter when the 40.96 x 40.96 cm2 imaging detector is placed 40 cm from the isocenter. Since this tomosynthesis application utilizes the extremities of the detector to record image detail relating to structures near the isocenter, simultaneous treatment and imaging is possible for most clinical cases, where the treated target is a small region close to the linac isocenter. The tomosynthesis images are reconstructed using the simultaneous iterative reconstruction technique, which is accelerated using a graphic processing unit. The authors present details of the system design as well as simulated performance of the imaging system based on reprojections of patient CT images.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19544779     DOI: 10.1118/1.3110067

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  18 in total

1.  Multibeam field emission x-ray system with half-scan reconstruction algorithm.

Authors:  Yang Lu; Hengyong Yu; Guohua Cao; Jun Zhao; Ge Wang; Otto Zhou
Journal:  Med Phys       Date:  2010-07       Impact factor: 4.071

2.  Accuracy estimation for projection-to-volume targeting during rotational therapy: a feasibility study.

Authors:  Yong Long; Jeffrey A Fessler; James M Balter
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

3.  A novel solid-angle tomosynthesis (SAT) scanning scheme.

Authors:  Jin Zhang; Cedric Yu
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

4.  Distributed source x-ray tube technology for tomosynthesis imaging.

Authors:  F Sprenger; X Calderon-Colon; Y Cheng; K Englestad; J Lu; J Maltz; A Paidi; X Qian; D Spronk; S Sultana; G Yang; O Zhou
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2010-01-01

5.  Stationary intraoral digital tomosynthesis using a carbon nanotube X-ray source array.

Authors:  J Shan; A W Tucker; L R Gaalaas; G Wu; E Platin; A Mol; J Lu; O Zhou
Journal:  Dentomaxillofac Radiol       Date:  2015-06-19       Impact factor: 2.419

6.  Pilot study for compact microbeam radiation therapy using a carbon nanotube field emission micro-CT scanner.

Authors:  Mike Hadsell; Guohua Cao; Jian Zhang; Laurel Burk; Torsten Schreiber; Eric Schreiber; Sha Chang; Jianping Lu; Otto Zhou
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

7.  Multisource inverse-geometry CT. Part II. X-ray source design and prototype.

Authors:  V Bogdan Neculaes; Antonio Caiafa; Yang Cao; Bruno De Man; Peter M Edic; Kristopher Frutschy; Satish Gunturi; Lou Inzinna; Joseph Reynolds; Mark Vermilyea; David Wagner; Xi Zhang; Yun Zou; Norbert J Pelc; Brian Lounsberry
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

8.  2D/3D Image Registration using Regression Learning.

Authors:  Chen-Rui Chou; Brandon Frederick; Gig Mageras; Sha Chang; Stephen Pizer
Journal:  Comput Vis Image Underst       Date:  2013-09-01       Impact factor: 3.876

9.  Image-guided microbeam irradiation to brain tumour bearing mice using a carbon nanotube x-ray source array.

Authors:  Lei Zhang; Hong Yuan; Laurel M Burk; Christy R Inscoe; Michael J Hadsell; Pavel Chtcheprov; Yueh Z Lee; Jianping Lu; Sha Chang; Otto Zhou
Journal:  Phys Med Biol       Date:  2014-02-20       Impact factor: 3.609

10.  Design and characterization of a spatially distributed multibeam field emission x-ray source for stationary digital breast tomosynthesis.

Authors:  Xin Qian; Ramya Rajaram; Xiomara Calderon-Colon; Guang Yang; Tuyen Phan; David S Lalush; Jianping Lu; Otto Zhou
Journal:  Med Phys       Date:  2009-10       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.