Literature DB >> 19544389

Validation of a new method for determination of cup orientation in THA.

Simon D Steppacher1, Moritz Tannast, Guoyan Zheng, Xuan Zhang, Jens Kowal, Suzanne E Anderson, Klaus A Siebenrock, Stephen B Murphy.   

Abstract

Our goal was to validate accuracy, consistency, and reproducibility/reliability of a new method for determining cup orientation in total hip arthroplasty (THA). This method allows matching the 3D-model from CT images or slices with the projected pelvis on an anteroposterior pelvic radiograph using a fully automated registration procedure. Cup orientation (inclination and anteversion) is calculated relative to the anterior pelvic plane, corrected for individual malposition of the pelvis during radiograph acquisition. Measurements on blinded and randomized radiographs of 80 cadaver and 327 patient hips were investigated. The method showed a mean accuracy of 0.7 +/- 1.7 degrees (-3.7 degrees to 4.0 degrees) for inclination and 1.2 +/- 2.4 degrees (-5.3 degrees to 5.6 degrees) for anteversion in the cadaver trials and 1.7 +/- 1.7 degrees (-4.6 degrees to 5.5 degrees) for inclination and 0.9 +/- 2.8 degrees (-5.2 degrees to 5.7 degrees) for anteversion in the clinical data when compared to CT-based measurements. No systematic errors in accuracy were detected with the Bland-Altman analysis. The software consistency and the reproducibility/reliability were very good. This software is an accurate, consistent, reliable, and reproducible method to measure cup orientation in THA using a sophisticated 2D/3D-matching technique. Its robust and accurate matching algorithm can be expanded to statistical models.

Entities:  

Mesh:

Year:  2009        PMID: 19544389     DOI: 10.1002/jor.20929

Source DB:  PubMed          Journal:  J Orthop Res        ISSN: 0736-0266            Impact factor:   3.494


  12 in total

1.  Improving cup positioning using a mechanical navigation instrument.

Authors:  Simon D Steppacher; Jens H Kowal; Stephen Barry Murphy
Journal:  Clin Orthop Relat Res       Date:  2011-02       Impact factor: 4.176

2.  Anteroposterior Radiographs Are More Accurate than Cross-Table Lateral Radiographs for Acetabular Anteversion Assessment: a Retrospective Cohort Study.

Authors:  Peter K Sculco; Alexander S McLawhorn; Kaitlin M Carroll; Benjamin A McArthur; David J Mayman
Journal:  HSS J       Date:  2015-10-19

3.  Assessing the accuracy factors in the determination of postoperative acetabular cup orientation using hybrid 2D-3D registration.

Authors:  Guoyan Zheng
Journal:  J Digit Imaging       Date:  2009-07-21       Impact factor: 4.056

4.  Computer assisted determination of acetabular cup orientation using 2D-3D image registration.

Authors:  Guoyan Zheng; Xuan Zhang
Journal:  Int J Comput Assist Radiol Surg       Date:  2010-02-24       Impact factor: 2.924

5.  A cadaver study validating CT assessment of acetabular component orientation: the Perth CT hip protocol.

Authors:  Varun Arora; Richard Hannan; Richard Beaver; Timothy Fletcher; Paul Harvie
Journal:  Skeletal Radiol       Date:  2016-11-08       Impact factor: 2.199

6.  Reliability and validity of measuring acetabular component orientation by plain anteroposterior radiographs.

Authors:  Ming Lu; Yi-Xin Zhou; Hui Du; Ji Zhang; Jian Liu
Journal:  Clin Orthop Relat Res       Date:  2013-05-04       Impact factor: 4.176

7.  Is the acetabular cup orientation after total hip arthroplasty on a two dimension or three dimension model accurate?

Authors:  Benjamin Craiovan; Tobias Renkawitz; Markus Weber; Joachim Grifka; Lutz Nolte; Guoyan Zheng
Journal:  Int Orthop       Date:  2014-04-16       Impact factor: 3.075

Review 8.  Radiographic assessment of the cup orientation after total hip arthroplasty: a literature review.

Authors:  Jing-Xin Zhao; Xiu-Yun Su; Zhe Zhao; Ruo-Xiu Xiao; Li-Cheng Zhang; Pei-Fu Tang
Journal:  Ann Transl Med       Date:  2020-02

9.  Validation of a statistical shape model-based 2D/3D reconstruction method for determination of cup orientation after THA.

Authors:  G Zheng; J von Recum; L-P Nolte; P A Grützner; S D Steppacher; J Franke
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-07-27       Impact factor: 2.924

10.  Pelvic tilt is minimally changed by total hip arthroplasty.

Authors:  William S Murphy; Greg Klingenstein; Stephen B Murphy; Guoyan Zheng
Journal:  Clin Orthop Relat Res       Date:  2013-02       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.