A bench study of intensive-care-unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators.

Arnaud W Thille1, Aissam Lyazidi, Jean-Christophe M Richard, Fabrice Galia, Laurent Brochard.   

Abstract

OBJECTIVE: To compare 13 commercially available, new-generation, intensive-care-unit (ICU) ventilators in terms of trigger function, pressurization capacity during pressure-support ventilation (PSV), accuracy of pressure measurements, and expiratory resistance. DESIGN AND
SETTING: Bench study at a research laboratory in a university hospital.
METHODS: Four turbine-based ventilators and nine conventional servo-valve compressed-gas ventilators were tested using a two-compartment lung model. Three levels of effort were simulated. Each ventilator was evaluated at four PSV levels (5, 10, 15, and 20 cm H2O), with and without positive end-expiratory pressure (5 cm H2O). Trigger function was assessed as the time from effort onset to detectable pressurization. Pressurization capacity was evaluated using the airway pressure-time product computed as the net area under the pressure-time curve over the first 0.3 s after inspiratory effort onset. Expiratory resistance was evaluated by measuring trapped volume in controlled ventilation.
RESULTS: Significant differences were found across the ventilators, with a range of triggering delays from 42 to 88 ms for all conditions averaged (P < 0.001). Under difficult conditions, the triggering delay was longer than 100 ms and the pressurization was poor for five ventilators at PSV5 and three at PSV10, suggesting an inability to unload patient's effort. On average, turbine-based ventilators performed better than conventional ventilators, which showed no improvement compared to a bench comparison in 2000.
CONCLUSION: Technical performance of trigger function, pressurization capacity, and expiratory resistance differs considerably across new-generation ICU ventilators. ICU ventilators seem to have reached a technical ceiling in recent years, and some ventilators still perform inadequately.

Entities:  

Mesh:

Year:  2009        PMID: 19352622      PMCID: PMC2873304          DOI: 10.1007/s00134-009-1467-7

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  30 in total

1.  Electrical activity of the diaphragm during pressure support ventilation in acute respiratory failure.

Authors:  J Beck; S B Gottfried; P Navalesi; Y Skrobik; N Comtois; M Rossini; C Sinderby
Journal:  Am J Respir Crit Care Med       Date:  2001-08-01       Impact factor: 21.405

2.  Evaluation of ventilators used during transport of ICU patients -- a bench study.

Authors:  G Zanetta; D Robert; C Guérin
Journal:  Intensive Care Med       Date:  2002-03-12       Impact factor: 17.440

Review 3.  Patient-ventilator interaction.

Authors:  M J Tobin; A Jubran; F Laghi
Journal:  Am J Respir Crit Care Med       Date:  2001-04       Impact factor: 21.405

4.  Bench testing of pressure support ventilation with three different generations of ventilators.

Authors:  J-C Richard; A Carlucci; L Breton; N Langlais; S Jaber; S Maggiore; S Fougère; A Harf; L Brochard
Journal:  Intensive Care Med       Date:  2002-05-30       Impact factor: 17.440

5.  Trigger performance of mid-level ICU mechanical ventilators during assisted ventilation: a bench study.

Authors:  Juliana C Ferreira; Daniel W Chipman; Robert M Kacmarek
Journal:  Intensive Care Med       Date:  2008-04-30       Impact factor: 17.440

6.  Work of breathing and airway occlusion pressure during assist-mode mechanical ventilation.

Authors:  C S Sassoon; C K Mahutte; T T Te; D H Simmons; R W Light
Journal:  Chest       Date:  1988-03       Impact factor: 9.410

7.  The effects of pressurization rate on breathing pattern, work of breathing, gas exchange and patient comfort in pressure support ventilation.

Authors:  D Chiumello; P Pelosi; M Croci; L M Bigatello; L Gattinoni
Journal:  Eur Respir J       Date:  2001-07       Impact factor: 16.671

8.  Performance of transport ventilator with patient-triggered ventilation.

Authors:  E Miyoshi; Y Fujino; T Mashimo; M Nishimura
Journal:  Chest       Date:  2000-10       Impact factor: 9.410

9.  Comparative bench study of triggering, pressurization, and cycling between the home ventilator VPAP II and three ICU ventilators.

Authors:  Didier Tassaux; Susannah Strasser; Silvana Fonseca; Eric Dalmas; Philippe Jolliet
Journal:  Intensive Care Med       Date:  2002-08-01       Impact factor: 17.440

10.  Influence of pressure- and flow-triggered synchronous intermittent mandatory ventilation on inspiratory muscle work.

Authors:  C S Sassoon; N Del Rosario; R Fei; C H Rheeman; S E Gruer; C K Mahutte
Journal:  Crit Care Med       Date:  1994-12       Impact factor: 7.598

View more
  24 in total

1.  Bench studies evaluating devices for non-invasive ventilation: critical analysis and future perspectives.

Authors:  Carlo Olivieri; Roberta Costa; Giorgio Conti; Paolo Navalesi
Journal:  Intensive Care Med       Date:  2011-11-29       Impact factor: 17.440

2.  Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilation.

Authors:  Aissam Lyazidi; Arnaud W Thille; Guillaume Carteaux; Fabrice Galia; Laurent Brochard; Jean-Christophe M Richard
Journal:  Intensive Care Med       Date:  2010-09-23       Impact factor: 17.440

3.  Bench testing of a new hyperbaric chamber ventilator at different atmospheric pressures.

Authors:  Jean-Claude Lefebvre; Aissam Lyazidi; Miguel Parceiro; Giuseppe F Sferrazza Papa; Evangelia Akoumianaki; Deborah Pugin; Didier Tassaux; Laurent Brochard; Jean-Christophe M Richard
Journal:  Intensive Care Med       Date:  2012-05-17       Impact factor: 17.440

4.  ICU mechanical ventilators, technological advances vs. user friendliness: the right picture is worth a thousand numbers.

Authors:  Jean Christophe Marie Richard; Robert M Kacmarek
Journal:  Intensive Care Med       Date:  2009-07-29       Impact factor: 17.440

5.  Mechanical influences on fluid leakage past the tracheal tube cuff in a benchtop model.

Authors:  Islem Ouanes; Aissam Lyazidi; Pierre Eric Danin; Nerlep Rana; Annalisa Di Bari; Fekri Abroug; Bruno Louis; Laurent Brochard
Journal:  Intensive Care Med       Date:  2011-02-12       Impact factor: 17.440

6.  FLOW-i ventilator performance in the presence of a circle system leak.

Authors:  Umberto Lucangelo; Miloš Ajčević; Agostino Accardo; Massimo Borelli; Alberto Peratoner; Lucia Comuzzi; Walter A Zin
Journal:  J Clin Monit Comput       Date:  2016-04-09       Impact factor: 2.502

7.  Accuracy of tidal volume delivered by home mechanical ventilation during mouthpiece ventilation: A bench evaluation.

Authors:  Adam Ogna; Helene Prigent; Line Falaize; Karl Leroux; Dante Santos; Isabelle Vaugier; David Orlikowski; Frederic Lofaso
Journal:  Chron Respir Dis       Date:  2016-07-08       Impact factor: 2.444

Review 8.  Year in review in Intensive Care Medicine 2009. Part III: mechanical ventilation, acute lung injury and respiratory distress syndrome, pediatrics, ethics, and miscellanea.

Authors:  Massimo Antonelli; Elie Azoulay; Marc Bonten; Jean Chastre; Giuseppe Citerio; Giorgio Conti; Daniel De Backer; François Lemaire; Herwig Gerlach; Goran Hedenstierna; Michael Joannidis; Duncan Macrae; Jordi Mancebo; Salvatore M Maggiore; Alexandre Mebazaa; Jean-Charles Preiser; Jerôme Pugin; Jan Wernerman; Haibo Zhang
Journal:  Intensive Care Med       Date:  2010-02-23       Impact factor: 17.440

9.  Potentially harmful effects of inspiratory synchronization during pressure preset ventilation.

Authors:  J C M Richard; A Lyazidi; E Akoumianaki; S Mortaza; R L Cordioli; J C Lefebvre; N Rey; L Piquilloud; G F Sferrazza Papa; G F Sferrazza-Papa; A Mercat; L Brochard
Journal:  Intensive Care Med       Date:  2013-08-09       Impact factor: 17.440

10.  A Novel Low-Cost Ventilator for Use in a Worldwide Pandemic: The Portsmouth Ventilator.

Authors:  Jacob H Cole; Scott B Hughey; Christopher H Rector; Gregory J Booth
Journal:  Crit Care Explor       Date:  2020-12-02
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.