Comparative bench study of triggering, pressurization, and cycling between the home ventilator VPAP II and three ICU ventilators.

Didier Tassaux1, Susannah Strasser, Silvana Fonseca, Eric Dalmas, Philippe Jolliet.   

Abstract

OBJECTIVE: To compare triggering, pressurization, and cycling of the home ventilator VPAP II with those of three ICU ventilators (Evita 4, Galileo, and Servo 300). DESIGN AND
SETTING: Two-compartment lung model study in a research laboratory, university hospital.
METHODS: One compartment was driven by an ICU ventilator to mimic "patient" inspiratory effort, while the other was connected to the tested ventilator. Pressure support of 10, 15, 20, and 25 cmH2O, and inspiratory efforts of 5, 10, 15, 20, and 25 cmH2O (inspiratory time 1 s) were used in normal, obstructive, and restrictive conditions. Triggering delay (Td), triggering workload, pressurization at 300 and 500 ms, and difference between the "patient's" inspiratory time and that of the ventilator were analyzed.
RESULTS: No difference was noted in triggering workload between VPAP II, Evita 4, and Galileo while Servo 300 had a lower value. Pressurization at 300 ms on Evita 4 and Servo 300 reached 75% of the ideal value, on Galileo 35%, and on VPAP II 45%. Pressurization at 500 ms on Evita 4 and Servo 300 reached 85% of the ideal value, on Galileo 50%, and on VPAP II 55%. Cycling was delayed in obstructive conditions and premature in restrictive conditions with each of the devices.
CONCLUSIONS: The VPAP II performed as well as one ICU ventilator and less well than two. Home devices for noninvasive ventilation in acute respiratory failure outside the ICU could prove attractive as they are smaller, less costly, and easier to use than ICU machines.

Entities:  

Mesh:

Year:  2002        PMID: 12209273     DOI: 10.1007/s00134-002-1421-4

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  18 in total

1.  Bench studies evaluating devices for non-invasive ventilation: critical analysis and future perspectives.

Authors:  Carlo Olivieri; Roberta Costa; Giorgio Conti; Paolo Navalesi
Journal:  Intensive Care Med       Date:  2011-11-29       Impact factor: 17.440

2.  Performance of noninvasive ventilation modes on ICU ventilators during pressure support: a bench model study.

Authors:  Laurence Vignaux; Didier Tassaux; Philippe Jolliet
Journal:  Intensive Care Med       Date:  2007-06-12       Impact factor: 17.440

3.  A bench study of intensive-care-unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators.

Authors:  Arnaud W Thille; Aissam Lyazidi; Jean-Christophe M Richard; Fabrice Galia; Laurent Brochard
Journal:  Intensive Care Med       Date:  2009-04-08       Impact factor: 17.440

4.  Trigger performance of mid-level ICU mechanical ventilators during assisted ventilation: a bench study.

Authors:  Juliana C Ferreira; Daniel W Chipman; Robert M Kacmarek
Journal:  Intensive Care Med       Date:  2008-04-30       Impact factor: 17.440

5.  Noninvasive positive pressure ventilation in infants with upper airway obstruction: comparison of continuous and bilevel positive pressure.

Authors:  Sandrine Essouri; Frédéric Nicot; Annick Clément; Erea-Noel Garabedian; Gilles Roger; Frédéric Lofaso; Brigitte Fauroux
Journal:  Intensive Care Med       Date:  2005-02-15       Impact factor: 17.440

6.  Non-invasive ventilation in acute respiratory distress syndrome: helmet use saves lives?

Authors:  Inderpaul Singh Sehgal; Ritesh Agarwal
Journal:  Ann Transl Med       Date:  2016-09

7.  A bench evaluation of fraction of oxygen in air delivery and tidal volume accuracy in home care ventilators available for hospital use.

Authors:  Loredana Baboi; Fabien Subtil; Claude Guérin
Journal:  J Thorac Dis       Date:  2016-12       Impact factor: 2.895

8.  Bench comparative evaluation of a new generation and standard helmet for delivering non-invasive ventilation.

Authors:  C Olivieri; R Costa; G Spinazzola; G Ferrone; F Longhini; G Cammarota; G Conti; P Navalesi
Journal:  Intensive Care Med       Date:  2012-12-06       Impact factor: 17.440

9.  Patient-ventilator asynchrony during non-invasive ventilation for acute respiratory failure: a multicenter study.

Authors:  Laurence Vignaux; Frédéric Vargas; Jean Roeseler; Didier Tassaux; Arnaud W Thille; Michel P Kossowsky; Laurent Brochard; Philippe Jolliet
Journal:  Intensive Care Med       Date:  2009-01-29       Impact factor: 17.440

10.  Monitoring Patient/Ventilator Interactions: Manufacturer's Perspective.

Authors:  Gerard Evers; Carl Van Loey
Journal:  Open Respir Med J       Date:  2009-03-12
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.