FLOW-i ventilator performance in the presence of a circle system leak.

Umberto Lucangelo1,2, Miloš Ajčević3, Agostino Accardo3, Massimo Borelli4, Alberto Peratoner5, Lucia Comuzzi5, Walter A Zin6.   

Abstract

Recently, the FLOW-i anaesthesia ventilator was developed based on the SERVO-i intensive care ventilator. The aim of this study was to test the FLOW-i's tidal volume delivery in the presence of a leak in the breathing circuit. We ventilated a test lung model in volume-, pressure-, and pressure-regulated volume-controlled modes (VC, PC, and PRVC, respectively) with a FLOW-i. First, the circuit remained airtight and the ventilator was tested with fresh gas flows of 6, 1, and 0.3 L/min in VC, PC, and PRVC modes and facing 4 combinations of different resistive and elastic loads. Second, a fixed leak in the breathing circuit was introduced and the measurements repeated. In the airtight system, FLOW-i maintained tidal volume (VT) and circuit pressure at approximately the set values, independently of respiratory mode, load, or fresh gas flow. In the leaking circuit, set VT = 500 mL, FLOW-i delivered higher VTs in PC (about 460 mL) than in VC and PRVC, where VTs were substantially less than 500 mL. Interestingly, VT did not differ appreciably from 6 to 0.3 L/min of fresh air flow among the 3 ventilatory modes. In the absence of leakage, peak inspiratory pressures were similar, while they were 35-45 % smaller in PRVC and VC than in PC mode in the presence of leaks. In conclusion, FLOW-i maintained VT (down to fresh gas flows of 0.3 L/min) to 90 % of its preset value in PC mode, which was 4-5 times greater than in VC or PRVC modes.

Entities:  

Keywords:  Air leaks; Low flow anaesthesia; Mechanical loads; Ventilatory modality; Volume reflector

Mesh:

Substances:

Year:  2016        PMID: 27062381     DOI: 10.1007/s10877-016-9867-z

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  18 in total

1.  Continuous positive airway pressure in new-generation mechanical ventilators: a lung model study.

Authors:  Muneyuki Takeuchi; Purris Williams; Dean Hess; Robert M Kacmarek
Journal:  Anesthesiology       Date:  2002-01       Impact factor: 7.892

2.  Performance characteristics of five new anesthesia ventilators and four intensive care ventilators in pressure-support mode: a comparative bench study.

Authors:  Samir Jaber; Didier Tassaux; Mustapha Sebbane; Yvan Pouzeratte; Anne Battisti; Xavier Capdevila; Jean-Jacques Eledjam; Philippe Jolliet
Journal:  Anesthesiology       Date:  2006-11       Impact factor: 7.892

3.  Bilevel vs ICU ventilators providing noninvasive ventilation: effect of system leaks: a COPD lung model comparison.

Authors:  Juliana C Ferreira; Daniel W Chipman; Nicholas S Hill; Robert M Kacmarek
Journal:  Chest       Date:  2009-05-08       Impact factor: 9.410

Review 4.  Simultaneous inference in general parametric models.

Authors:  Torsten Hothorn; Frank Bretz; Peter Westfall
Journal:  Biom J       Date:  2008-06       Impact factor: 2.207

5.  Critical incident reports concerning anaesthetic equipment: analysis of the UK National Reporting and Learning System (NRLS) data from 2006-2008*.

Authors:  C J Cassidy; A Smith; J Arnot-Smith
Journal:  Anaesthesia       Date:  2011-07-25       Impact factor: 6.955

Review 6.  Lung mechanics at the bedside: make it simple.

Authors:  Umberto Lucangelo; Francesca Bernabè; Lluis Blanch
Journal:  Curr Opin Crit Care       Date:  2007-02       Impact factor: 3.687

7.  Respiratory mechanics during halothane anesthesia and anesthesia-paralysis in humans.

Authors:  P K Behrakis; B D Higgs; A Baydur; W A Zin; J Milic-Emili
Journal:  J Appl Physiol Respir Environ Exerc Physiol       Date:  1983-10

8.  Do new anesthesia ventilators deliver small tidal volumes accurately during volume-controlled ventilation?

Authors:  Patricia R Bachiller; Joseph M McDonough; Jeffrey M Feldman
Journal:  Anesth Analg       Date:  2008-05       Impact factor: 5.108

9.  Accuracy of inhaled agent usage displays of automated target control anesthesia machines.

Authors:  Alexander Dehouwer; Rik Carette; Simon De Ridder; Andre M De Wolf; Jan F A Hendrickx
Journal:  J Clin Monit Comput       Date:  2015-08-07       Impact factor: 2.502

10.  A comparison of volume control and pressure-regulated volume control ventilation in acute respiratory failure.

Authors: 
Journal:  Crit Care       Date:  1997       Impact factor: 9.097

View more
  2 in total

Review 1.  Journal of clinical monitoring and computing 2017 end of year summary: anesthesia.

Authors:  Jan F A Hendrickx; Andre M De Wolf; Stanley Skinner
Journal:  J Clin Monit Comput       Date:  2018-02-24       Impact factor: 2.502

2.  Wash-in and wash-out of sevoflurane in a test-lung model: A comparison between Aisys and FLOW-i.

Authors:  Petter Jakobsson; Madleine Lindgren; Jan G Jakobsson
Journal:  F1000Res       Date:  2017-03-29
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.