Accuracy in the localization of thoracic and abdominal tumors using respiratory displacement, velocity, and phase.

U W Langner1, P J Keall.   

Abstract

UNLABELLED: Current four dimensional (4D) computed tomography (CT) reconstruction techniques are retrospectively created based on either the phase or displacement of the respiratory signal. Both techniques have known limitations which cause clinically significant motion artifacts in 4D CT images. These artifacts, which appear as undefined or irregular boundaries in the 4D CT images, cause systematic errors in patient contouring and dose calculations. The aim of this work was to evaluate the reproducibility of tumor position as a function of displacement, phase, and velocity of the respiratory signal, respectively, in order to determine the optimum parameter or combination of parameters to use in order to minimize artifacts in 4D CT images or to accurately deliver radiation to relevant structures during treatment. METHOD AND MATERIALS: Estimated tumor centroid position and respiratory signal data were acquired with the Cyberknife Synchrony system for 26 thoracic radiotherapy patients (52 fractions). A reference respiratory cycle was calculated for each patient. Displacement, phase, and velocity of ten data points were calculated from this reference respiratory cycle, where each point represents an image bin. The corresponding tumor position was then sorted into these image bins if the phase, displacement, simultaneous displacement and phase, or simultaneous displacement and velocity of the respiratory signal were within tolerances of 0.5 mm for displacement and 0.5 mm/s for velocity, respectively, from the corresponding data of the reference cycle for each image bin.
RESULTS: The mean of the standard deviations of tumor positions over all bins and all fractions for the superior-inferior direction were 2.13 +/- 1.01 mm for phase sorting, 1.20 +/- 0.76 mm for displacement sorting, 1.20 +/- 0.71 mm for simultaneous displacement and phase sorting, and 1.10 +/- 0.71 mm for simultaneous displacement and velocity sorting, with maximum deviations of 43.0, 16.1, 15.5, and 14.1 mm for each scenario, respectively. The same trend was observed for the anterior-posterior and left-right directions. A linear dependence was observed between the mean of the standard deviations of tumor positions over all fractions as a function of the velocity of the respiratory signal at each bin for all the sorting scenarios. A substantially larger gradient for the phase sorting scenario, compared to the other scenarios, suggests that tumor localization will become increasingly less accurate as the velocity of the tumor increases during a breathing cycle, e.g., if the amplitude of motion increases while the period of the respiratory cycle stays constant or during mid inhale or exhale phases of the respiratory cycle.
CONCLUSION: This study illustrates that position of a tumor can be determined more accurately if displacement and velocity are used simultaneously as sorting parameters for 4D CT images or during treatment. A real-time displacement and velocity based 4D CT image sorting method may therefore produce fewer and smaller artifacts in 4D CT images than current retrospective sorting methods.

Entities:  

Mesh:

Year:  2009        PMID: 19291977      PMCID: PMC2736730          DOI: 10.1118/1.3049595

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  19 in total

1.  Robotics and radiosurgery--the cyberknife.

Authors:  S D Chang; J R Adler
Journal:  Stereotact Funct Neurosurg       Date:  2001       Impact factor: 1.875

2.  The impact of temporal inaccuracies on 4DCT image quality.

Authors:  Y D Mutaf; J A Antolak; D H Brinkmann
Journal:  Med Phys       Date:  2007-05       Impact factor: 4.071

3.  Improving retrospective sorting of 4D computed tomography data.

Authors:  Eike Rietzel; George T Y Chen
Journal:  Med Phys       Date:  2006-02       Impact factor: 4.071

4.  Tracking lung tissue motion and expansion/compression with inverse consistent image registration and spirometry.

Authors:  Gary E Christensen; Joo Hyun Song; Wei Lu; Issam El Naqa; Daniel A Low
Journal:  Med Phys       Date:  2007-06       Impact factor: 4.071

5.  Novel breathing motion model for radiotherapy.

Authors:  Daniel A Low; Parag J Parikh; Wei Lu; James F Dempsey; Sasha H Wahab; James P Hubenschmidt; Michelle M Nystrom; Maureen Handoko; Jeffrey D Bradley
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-11-01       Impact factor: 7.038

6.  A comparison between amplitude sorting and phase-angle sorting using external respiratory measurement for 4D CT.

Authors:  Wei Lu; Parag J Parikh; James P Hubenschmidt; Jeffrey D Bradley; Daniel A Low
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

7.  An optical flow based method for improved reconstruction of 4D CT data sets acquired during free breathing.

Authors:  Jan Ehrhardt; René Werner; Dennis Säring; Thorsten Frenzel; Wei Lu; Daniel Low; Heinz Handels
Journal:  Med Phys       Date:  2007-02       Impact factor: 4.071

8.  Geometric uncertainty of 2D projection imaging in monitoring 3D tumor motion.

Authors:  Yelin Suh; Sonja Dieterich; Paul J Keall
Journal:  Phys Med Biol       Date:  2007-05-18       Impact factor: 3.609

9.  Phase and amplitude binning for 4D-CT imaging.

Authors:  A F Abdelnour; S A Nehmeh; T Pan; J L Humm; P Vernon; H Schöder; K E Rosenzweig; G S Mageras; E Yorke; S M Larson; Y E Erdi
Journal:  Phys Med Biol       Date:  2007-05-18       Impact factor: 3.609

10.  Prospective displacement and velocity-based cine 4D CT.

Authors:  U W Langner; P J Keall
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

View more
  13 in total

1.  Four-dimensional intensity-modulated radiation therapy planning for dynamic tracking using a direct aperture deformation (DAD) method.

Authors:  Minzhi Gui; Yuanming Feng; Byongyong Yi; Anil Arvind Dhople; Cedric Yu
Journal:  Med Phys       Date:  2010-05       Impact factor: 4.071

2.  On correlated sources of uncertainty in four dimensional computed tomography data sets.

Authors:  Eric D Ehler; Wolfgang A Tome
Journal:  Technol Cancer Res Treat       Date:  2010-06

3.  Technical Note: Simulation of 4DCT tumor motion measurement errors.

Authors:  Tai H Dou; David H Thomas; Dylan O'Connell; Jeffrey D Bradley; James M Lamb; Daniel A Low
Journal:  Med Phys       Date:  2015-10       Impact factor: 4.071

4.  Correlation between external and internal respiratory motion: a validation study.

Authors:  Floris Ernst; Ralf Bruder; Alexander Schlaefer; Achim Schweikard
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-08-19       Impact factor: 2.924

5.  Rapid estimation of 4DCT motion-artifact severity based on 1D breathing-surrogate periodicity.

Authors:  Guang Li; Marshall Caraveo; Jie Wei; Andreas Rimner; Abraham J Wu; Karyn A Goodman; Ellen Yorke
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

Review 6.  Advances in 4D radiation therapy for managing respiration: part I - 4D imaging.

Authors:  Geoffrey D Hugo; Mihaela Rosu
Journal:  Z Med Phys       Date:  2012-07-10       Impact factor: 4.820

7.  Patient specific respiratory motion modeling using a 3D patient's external surface.

Authors:  Hadi Fayad; Tinsu Pan; Olivier Pradier; Dimitris Visvikis
Journal:  Med Phys       Date:  2012-06       Impact factor: 4.071

8.  A Method for Assessing Ground-Truth Accuracy of the 5DCT Technique.

Authors:  Tai H Dou; David H Thomas; Dylan P O'Connell; James M Lamb; Percy Lee; Daniel A Low
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-08-07       Impact factor: 7.038

9.  A new respiratory monitor system for four-dimensional computed tomography by measuring the pressure change on the back of body.

Authors:  Xianwen Zhang; Jintian Tang; Gregory C Sharp; Lei Xiao; Shouping Xu; Hsiao-Ming Lu
Journal:  Br J Radiol       Date:  2020-01-20       Impact factor: 3.039

10.  Reproducibility of image quality for moving objects using respiratory-gated computed tomography: a study using a phantom model.

Authors:  Nobuyoshi Fukumitsu; Masaya Ishida; Toshiyuki Terunuma; Masashi Mizumoto; Takayuki Hashimoto; Takashi Moritake; Toshiyuki Okumura; Takeji Sakae; Koji Tsuboi; Hideyuki Sakurai
Journal:  J Radiat Res       Date:  2012-09-10       Impact factor: 2.724

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.