Non-invasive glucose monitoring in patients with Type 1 diabetes: a Multisensor system combining sensors for dielectric and optical characterisation of skin.

Andreas Caduff1, Mark S Talary, Martin Mueller, Francois Dewarrat, Jelena Klisic, Marc Donath, Lutz Heinemann, Werner A Stahel.   

Abstract

In vivo variations of blood glucose (BG) are affecting the biophysical characteristics (e.g. dielectric and optical) of skin and underlying tissue (SAUT) at various frequencies. However, the skin impedance spectra for instance can also be affected by other factors, perturbing the glucose related information, factors such as temperature, skin moisture and sweat, blood perfusion as well as body movements affecting the sensor-skin contact. In order to be able to correct for such perturbing factors, a Multisensor system was developed including sensors to measure the identified factors. To evaluate the quality of glucose monitoring, the Multisensor was applied in 10 patients with Type 1 diabetes. Glucose was administered orally to induce hyperglycaemic excursions at two different study visits. For analysis of the sensor signals, a global multiple linear regression model was derived. The respective coefficients of the variables were determined from the sensor signals of this first study visit (R(2)=0.74, MARD=18.0%--mean absolute relative difference). The identical set of modelling coefficients of the first study visit was re-applied to the test data of the second study visit to evaluate the predictive power of the model (R(2)=0.68, MARD=27.3%). It appears as if the Multisensor together with the global linear regression model applied, allows for tracking glucose changes non-invasively in patients with diabetes without requiring new model coefficients for each visit. Confirmation of these findings in a larger study group and under less experimentally controlled conditions is required for understanding whether a global parameterisation routine is feasible.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19286364     DOI: 10.1016/j.bios.2009.02.001

Source DB:  PubMed          Journal:  Biosens Bioelectron        ISSN: 0956-5663            Impact factor:   10.618


  18 in total

Review 1.  Technologies for continuous glucose monitoring: current problems and future promises.

Authors:  Santhisagar Vaddiraju; Diane J Burgess; Ioannis Tomazos; Faquir C Jain; Fotios Papadimitrakopoulos
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

2.  Non-invasive continuous glucose monitoring: improved accuracy of point and trend estimates of the Multisensor system.

Authors:  Mattia Zanon; Giovanni Sparacino; Andrea Facchinetti; Michela Riz; Mark S Talary; Roland E Suri; Andreas Caduff; Claudio Cobelli
Journal:  Med Biol Eng Comput       Date:  2012-06-22       Impact factor: 2.602

3.  Data processing for noninvasive continuous glucose monitoring with a multisensor device.

Authors:  Martin Mueller; Mark S Talary; Lisa Falco; Oscar De Feo; Werner A Stahel; Andreas Caduff
Journal:  J Diabetes Sci Technol       Date:  2011-05-01

4.  Dynamics of blood electrolytes in repeated hyper- and/or hypoglycaemic events in patients with type 1 diabetes.

Authors:  A Caduff; H U Lutz; L Heinemann; G Di Benedetto; M S Talary; S Theander
Journal:  Diabetologia       Date:  2011-06-15       Impact factor: 10.122

5.  First Experiences With a Wearable Multisensor Device in a Noninvasive Continuous Glucose Monitoring Study at Home, Part II: The Investigators' View.

Authors:  Mattia Zanon; Martin Mueller; Pavel Zakharov; Mark S Talary; Marc Donath; Werner A Stahel; Andreas Caduff
Journal:  J Diabetes Sci Technol       Date:  2017-11-16

6.  The Effect of a Global, Subject, and Device-Specific Model on a Noninvasive Glucose Monitoring Multisensor System.

Authors:  Andreas Caduff; Mattia Zanon; Martin Mueller; Pavel Zakharov; Yuri Feldman; Oscar De Feo; Marc Donath; Werner A Stahel; Mark S Talary
Journal:  J Diabetes Sci Technol       Date:  2015-04-24

7.  Accuracy of a Novel Noninvasive Multisensor Technology to Estimate Glucose in Diabetic Subjects During Dynamic Conditions.

Authors:  Sandra I Sobel; Peter J Chomentowski; Nisarg Vyas; David Andre; Frederico G S Toledo
Journal:  J Diabetes Sci Technol       Date:  2014-01-01

8.  First Experiences With a Wearable Multisensor in an Outpatient Glucose Monitoring Study, Part I: The Users' View.

Authors:  Andreas Caduff; Mattia Zanon; Pavel Zakharov; Martin Mueller; Mark Talary; Achim Krebs; Werner A Stahel; Marc Donath
Journal:  J Diabetes Sci Technol       Date:  2018-01-14

9.  Enhancing the Accuracy of Non-Invasive Glucose Sensing in Aqueous Solutions Using Combined Millimeter Wave and Near Infrared Transmission.

Authors:  Helena Cano-Garcia; Rohit Kshirsagar; Roberto Pricci; Ahmed Teyeb; Fergus O'Brien; Shimul Saha; Panagiotis Kosmas; Efthymios Kallos
Journal:  Sensors (Basel)       Date:  2021-05-10       Impact factor: 3.576

Review 10.  Italian contributions to the development of continuous glucose monitoring sensors for diabetes management.

Authors:  Giovanni Sparacino; Mattia Zanon; Andrea Facchinetti; Chiara Zecchin; Alberto Maran; Claudio Cobelli
Journal:  Sensors (Basel)       Date:  2012-10-12       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.