Data processing for noninvasive continuous glucose monitoring with a multisensor device.

Martin Mueller1, Mark S Talary, Lisa Falco, Oscar De Feo, Werner A Stahel, Andreas Caduff.   

Abstract

BACKGROUND: Impedance spectroscopy has been shown to be a candidate for noninvasive continuous glucose monitoring in humans. However, in addition to glucose, other factors also have effects on impedance characteristics of the skin and underlying tissue.
METHOD: Impedance spectra were summarized through a principal component analysis and relevant variables were identified with Akaike's information criterion. In order to model blood glucose, a linear least-squares model was used. A Monte Carlo simulation was applied to examine the effects of personalizing models.
RESULTS: The principal component analysis was able to identify two major effects in the impedance spectra: a blood glucose-related process and an equilibration process related to moisturization of the skin and underlying tissue. With a global linear least-squares model, a coefficient of determination (R²) of 0.60 was achieved, whereas the personalized model reached an R² of 0.71. The Monte Carlo simulation proved a significant advantage of personalized models over global models.
CONCLUSION: A principal component analysis is useful for extracting glucose-related effects in the impedance spectra of human skin. A linear global model based on Solianis Multisensor data yields a good predictive power for blood glucose estimation. However, a personalized linear model still has greater predictive power.
© 2011 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21722585      PMCID: PMC3192636          DOI: 10.1177/193229681100500324

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  14 in total

1.  Impact of posture and fixation technique on impedance spectroscopy used for continuous and noninvasive glucose monitoring.

Authors:  Andreas Pfützner; Andreas Caduff; Martin Larbig; Thomas Schrepfer; Thomas Forst
Journal:  Diabetes Technol Ther       Date:  2004-08       Impact factor: 6.118

Review 2.  Dielectric properties of tissues and biological materials: a critical review.

Authors:  K R Foster; H P Schwan
Journal:  Crit Rev Biomed Eng       Date:  1989

3.  Noninvasive glucose sensing.

Authors:  Mark A Arnold; Gary W Small
Journal:  Anal Chem       Date:  2005-09-01       Impact factor: 6.986

4.  Characteristics of a multisensor system for non invasive glucose monitoring with external validation and prospective evaluation.

Authors:  Andreas Caduff; Martin Mueller; Alexander Megej; Francois Dewarrat; Roland E Suri; Jelena Klisic; Marc Donath; Pavel Zakharov; Dominik Schaub; Werner A Stahel; Mark S Talary
Journal:  Biosens Bioelectron       Date:  2011-04-13       Impact factor: 10.618

5.  The Monte Carlo method.

Authors:  N METROPOLIS; S ULAM
Journal:  J Am Stat Assoc       Date:  1949-09       Impact factor: 5.033

6.  The dynamics of transepidermal water loss (TEWL) from hydrated skin.

Authors:  Francesco Gioia; Leonardo Celleno
Journal:  Skin Res Technol       Date:  2002-08       Impact factor: 2.365

7.  Full-field optical coherence tomography for the rapid estimation of epidermal thickness: study of patients with diabetes mellitus type 1.

Authors:  P Zakharov; M S Talary; I Kolm; A Caduff
Journal:  Physiol Meas       Date:  2009-12-16       Impact factor: 2.833

8.  Non-invasive glucose monitoring in patients with Type 1 diabetes: a Multisensor system combining sensors for dielectric and optical characterisation of skin.

Authors:  Andreas Caduff; Mark S Talary; Martin Mueller; Francois Dewarrat; Jelena Klisic; Marc Donath; Lutz Heinemann; Werner A Stahel
Journal:  Biosens Bioelectron       Date:  2009-02-20       Impact factor: 10.618

9.  Impact of environmental temperature on skin thickness and microvascular blood flow in subjects with and without diabetes.

Authors:  T Forst; A Caduff; M Talary; M Weder; M Brändle; P Kann; F Flacke; Ch Friedrich; A Pfützner
Journal:  Diabetes Technol Ther       Date:  2006-02       Impact factor: 6.118

10.  Non-invasive glucose monitoring in patients with diabetes: a novel system based on impedance spectroscopy.

Authors:  A Caduff; F Dewarrat; M Talary; G Stalder; L Heinemann; Yu Feldman
Journal:  Biosens Bioelectron       Date:  2006-03-09       Impact factor: 10.618

View more
  5 in total

1.  Non-invasive continuous glucose monitoring: improved accuracy of point and trend estimates of the Multisensor system.

Authors:  Mattia Zanon; Giovanni Sparacino; Andrea Facchinetti; Michela Riz; Mark S Talary; Roland E Suri; Andreas Caduff; Claudio Cobelli
Journal:  Med Biol Eng Comput       Date:  2012-06-22       Impact factor: 2.602

2.  Selection of Noninvasive Features in Wrist-Based Wearable Sensors to Predict Blood Glucose Concentrations Using Machine Learning Algorithms.

Authors:  Brian Bogue-Jimenez; Xiaolei Huang; Douglas Powell; Ana Doblas
Journal:  Sensors (Basel)       Date:  2022-05-06       Impact factor: 3.847

3.  Electromagnetic Differential Measuring Method: Application in Microstrip Sensors Developing.

Authors:  Francisco Javier Ferrández-Pastor; Juan Manuel García-Chamizo; Mario Nieto-Hidalgo
Journal:  Sensors (Basel)       Date:  2017-07-18       Impact factor: 3.576

Review 4.  Review of Non-invasive Glucose Sensing Techniques: Optical, Electrical and Breath Acetone.

Authors:  Maryamsadat Shokrekhodaei; Stella Quinones
Journal:  Sensors (Basel)       Date:  2020-02-25       Impact factor: 3.576

Review 5.  Italian contributions to the development of continuous glucose monitoring sensors for diabetes management.

Authors:  Giovanni Sparacino; Mattia Zanon; Andrea Facchinetti; Chiara Zecchin; Alberto Maran; Claudio Cobelli
Journal:  Sensors (Basel)       Date:  2012-10-12       Impact factor: 3.576

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.