Towards optimized acquisition scheme for multiprojection correlation imaging of breast cancer.

Amarpreet S Chawla1, Robert S Saunders, Swatee Singh, Joseph Y Lo, Ehsan Samei.   

Abstract

RATIONALE AND
OBJECTIVES: Correlation imaging (CI) is a form of multiprojection imaging in which multiple images of a patient are acquired from slightly different angles. Information from these images is combined to make the final diagnosis. A critical factor affecting the performance of CI is its data acquisition scheme, because nonoptimized acquisition may distort pathologic indicators. The authors describe a computer-aided detection (CADe) methodology to optimize the acquisition scheme of CI for superior diagnostic accuracy.
MATERIALS AND METHODS: Images from 106 subjects were used. For each subject, 25 angular projections of a single breast were acquired. Projection images were supplemented with a simulated 3-mm three-dimensional lesion. Each projection was then processed using a traditional CADe algorithm at high sensitivity, followed by the reduction of false-positives by combining the geometric correlation information available from the multiple images. The performance of the CI system was determined in terms of free-response receiver-operating characteristic curves and the areas under receiver-operating characteristic curves. For optimization, the components of acquisition, such as the number of projections and their angular span, were systematically changed to investigate which of the many possible combinations maximized the obtainable CADe sensitivity and specificity.
RESULTS: The performance of the CI system was improved by increasing the angular span. Increasing the number of angular projections beyond a certain number did not improve performance. Maximum performance was obtained between 7 and 10 projections spanning a maximum angular arc of 45 degrees .
CONCLUSION: The findings suggest the existence of an optimum acquisition scheme for CI of the breast. CADe results confirmed earlier predictions on the basis of observer models. An optimized CI system may be an important diagnostic tool for improved breast cancer detection.

Entities:  

Mesh:

Year:  2009        PMID: 19268858      PMCID: PMC3789538          DOI: 10.1016/j.acra.2008.09.013

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  9 in total

Review 1.  Digital x-ray tomosynthesis: current state of the art and clinical potential.

Authors:  James T Dobbins; Devon J Godfrey
Journal:  Phys Med Biol       Date:  2003-10-07       Impact factor: 3.609

2.  Effects of quantum noise and binocular summation on dose requirements in stereoradiography.

Authors:  Andrew D A Maidment; Predrag R Bakic; Michael Albert
Journal:  Med Phys       Date:  2003-12       Impact factor: 4.071

3.  Impact of resolution and noise characteristics of digital radiographic detectors on the detectability of lung nodules.

Authors:  Robert S Saunders; Ehsan Samei; Christoph Hoeschen
Journal:  Med Phys       Date:  2004-06       Impact factor: 4.071

4.  Voting strategy for artifact reduction in digital breast tomosynthesis.

Authors:  Tao Wu; Richard H Moore; Daniel B Kopans
Journal:  Med Phys       Date:  2006-07       Impact factor: 4.071

5.  Characterization of difference of Gaussian filters in the detection of mammographic regions.

Authors:  David M Catarious; Alan H Baydush; Carey E Floyd
Journal:  Med Phys       Date:  2006-11       Impact factor: 4.071

6.  Optimized approach to decision fusion of heterogeneous data for breast cancer diagnosis.

Authors:  Jonathan L Jesneck; Loren W Nolte; Jay A Baker; Carey E Floyd; Joseph Y Lo
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

7.  Multiprojection correlation imaging for improved detection of pulmonary nodules.

Authors:  Ehsan Samei; Stanton A Stebbins; James T Dobbins; H Page McAdams; Joseph Y Lo
Journal:  AJR Am J Roentgenol       Date:  2007-05       Impact factor: 3.959

8.  Computerized mass detection for digital breast tomosynthesis directly from the projection images.

Authors:  I Reiser; R M Nishikawa; M L Giger; T Wu; E A Rafferty; R Moore; D B Kopans
Journal:  Med Phys       Date:  2006-02       Impact factor: 4.071

9.  A mathematical model platform for optimizing a multiprojection breast imaging system.

Authors:  Amarpreet S Chawla; Ehsan Samei; Robert S Saunders; Joseph Y Lo; Jay A Baker
Journal:  Med Phys       Date:  2008-04       Impact factor: 4.071

  9 in total
  2 in total

1.  Optimized image acquisition for breast tomosynthesis in projection and reconstruction space.

Authors:  Amarpreet S Chawla; Joseph Y Lo; Jay A Baker; Ehsan Samei
Journal:  Med Phys       Date:  2009-11       Impact factor: 4.071

2.  Effect of slice thickness on detectability in breast CT using a prewhitened matched filter and simulated mass lesions.

Authors:  Nathan J Packard; Craig K Abbey; Kai Yang; John M Boone
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.