Beamwidth measurement of individual lithotripter shock waves.

Wayne Kreider1, Michael R Bailey, Jeffrey A Ketterling.   

Abstract

New lithotripters with narrower foci and higher peak pressures than the original Dornier HM3 electrohydraulic lithotripter have proven to be less effective and less safe. Hence, accurate measurements of the focal characteristics of lithotripter shock waves are important. The current technique for measuring beamwidth requires a collection of single-point measurements over multiple shock waves, thereby introducing error as a result of any shock-to-shock variability. This work reports on the construction of a hydrophone array sensor and on array measurements of individual lithotripter shock waves. Beamwidths for an electrohydraulic lithotripter with a broad-focus HM3-style reflector and a narrow-focus modified reflector were measured using both new and worn electrodes as well as two different electrical charging potentials. The array measured the waveform, beamwidth, and focal location of individual shock waves. The HM3-style reflector produced repeatable focal waveforms and beam profiles at an 18 kV charging potential with new and worn electrodes. Corresponding measurements suggest a narrower beamwidth than reported previously from averaged point measurements acquired under the same conditions. In addition, a lack of consistency in the measured beam profiles at 23 kV underscores the value of measuring individual shock waves.

Entities:  

Mesh:

Year:  2009        PMID: 19206897      PMCID: PMC2677365          DOI: 10.1121/1.3050272

Source DB:  PubMed          Journal:  J Acoust Soc Am        ISSN: 0001-4966            Impact factor:   1.840


  13 in total

Review 1.  The mechanisms of stone fragmentation in ESWL.

Authors:  W Eisenmenger
Journal:  Ultrasound Med Biol       Date:  2001-05       Impact factor: 2.998

2.  Current management of urolithiasis: progress or regress?

Authors:  Kurt Kerbl; Jamil Rehman; Jaime Landman; David Lee; Chandru Sundaram; Ralph V Clayman
Journal:  J Endourol       Date:  2002-06       Impact factor: 2.942

3.  [The Dornier lithotripter in a comparison. Measuring shockwave fields and fragmentation effects].

Authors:  M Müller
Journal:  Biomed Tech (Berl)       Date:  1990-11       Impact factor: 1.411

Review 4.  Strategies for improved shock wave lithotripsy.

Authors:  J A McAteer; M R Bailey; J C Williams; R O Cleveland; A P Evan
Journal:  Minerva Urol Nefrol       Date:  2005-12       Impact factor: 3.720

5.  Modeling elastic wave propagation in kidney stones with application to shock wave lithotripsy.

Authors:  Robin O Cleveland; Oleg A Sapozhnikov
Journal:  J Acoust Soc Am       Date:  2005-10       Impact factor: 1.840

6.  Urologic diseases in America project: urolithiasis.

Authors:  Margaret S Pearle; Elizabeth A Calhoun; Gary C Curhan
Journal:  J Urol       Date:  2005-03       Impact factor: 7.450

7.  Air pockets trapped during routine coupling in dry head lithotripsy can significantly decrease the delivery of shock wave energy.

Authors:  Yuri A Pishchalnikov; Joshua S Neucks; R Jason VonDerHaar; Irina V Pishchalnikova; James C Williams; James A McAteer
Journal:  J Urol       Date:  2006-12       Impact factor: 7.450

8.  Cavitation selectively reduces the negative-pressure phase of lithotripter shock pulses.

Authors:  Yuri A Pishchalnikov; Oleg A Sapozhnikov; Michael R Bailey; Irina V Pishchalnikova; James C Williams; James A McAteer
Journal:  Acoust Res Lett Online       Date:  2005-11-03

9.  In vivo pressure measurements of lithotripsy shock waves in pigs.

Authors:  R O Cleveland; D A Lifshitz; B A Connors; A P Evan; L R Willis; L A Crum
Journal:  Ultrasound Med Biol       Date:  1998-02       Impact factor: 2.998

10.  A mechanistic analysis of stone fracture in lithotripsy.

Authors:  Oleg A Sapozhnikov; Adam D Maxwell; Brian MacConaghy; Michael R Bailey
Journal:  J Acoust Soc Am       Date:  2007-02       Impact factor: 1.840

View more
  1 in total

1.  Single-shot measurements of the acoustic field of an electrohydraulic lithotripter using a hydrophone array.

Authors:  Mohammad A Alibakhshi; Jonathan M Kracht; Robin O Cleveland; Erwan Filoux; Jeffrey A Ketterling
Journal:  J Acoust Soc Am       Date:  2013-05       Impact factor: 1.840

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.