Literature DB >> 19182503

Slot-scan digital radiography of the lower extremities: a comparison to computed radiography with respect to image quality and radiation dose.

Kwang Hwi Lee1, Jong Won Kwon, Young Cheol Yoon, Sang Hee Choi, Jee Young Jung, Ji Hye Kim, Sang Jun Lee.   

Abstract

OBJECTIVE: To compare the slot-scan digital radiography (SSDR) of the lower extremity region and the computed radiography (CR) method with respect to the image quality and radiation exposure.
MATERIALS AND METHODS: We enrolled 54 patients who underwent both the SSDR and CR of the lower extremities. The study evaluated and statistically compared the image quality of four features (outer cortex, inner cortex, trabeculae and intermuscular fat) at six different levels (pelvis, hip, femur, knee, tibia and ankle) between each method. The image quality was evaluated using a visibility scale, and the entrance skin dose was measured using a dosimeter at three different levels of a phantom (hip, knee, and ankle).
RESULTS: The mean image visibility scale values for the SSDR method were significantly higher than for the CR method. The entrance skin dose for the SSDR method was 278 microGy at each level, compared to the entrance skin doses of the CR method, which were 3,410 microGy for the hip, 1,152 microGy for the knee, and 580 microGy for the ankle.
CONCLUSION: Both the image quality and patient entrance skin dose data suggest that the SSDR method is superior to the CR method for the lower extremity musculoskeletal examination.

Entities:  

Mesh:

Year:  2009        PMID: 19182503      PMCID: PMC2647173          DOI: 10.3348/kjr.2009.10.1.51

Source DB:  PubMed          Journal:  Korean J Radiol        ISSN: 1229-6929            Impact factor:   3.500


  15 in total

1.  A slot-scanned photodiode-array/CCD hybrid detector for digital mammography.

Authors:  James G Mainprize; Nancy L Ford; Shi Yin; Türmay Tümer; Martin J Yaffe
Journal:  Med Phys       Date:  2002-02       Impact factor: 4.071

2.  Lumbar spine radiography: digital flat-panel detector versus screen-film and storage-phosphor systems in monkeys as a pediatric model.

Authors:  Karl Ludwig; Kathrin Ahlers; Dag Wormanns; Michael Freund; Thomas M Bernhardt; Stefan Diederich; Walter Heindel
Journal:  Radiology       Date:  2003-08-18       Impact factor: 11.105

3.  Selenium-based digital radiography versus high-resolution storage phosphor radiography in the detection of solitary pulmonary nodules without calcification: receiver operating characteristic curve analysis.

Authors:  K Awai; M Komi; S Hori
Journal:  AJR Am J Roentgenol       Date:  2001-11       Impact factor: 3.959

4.  Detection of porcine bone lesions and fissures: comparing digital selenium, digital luminescence, and analog film-screen radiography.

Authors:  M Zähringer; B Krug; K F Kamm; G Wassmer; M Hellmich; G Winnekendonk; J Andermahr; A Gossmann; K J Lackner
Journal:  AJR Am J Roentgenol       Date:  2001-12       Impact factor: 3.959

5.  Reliability analysis for radiographic measurement of limb length discrepancy: full-length standing anteroposterior radiograph versus scanogram.

Authors:  Sanjeev Sabharwal; Caixia Zhao; John McKeon; Todd Melaghari; Marcia Blacksin; Cornelia Wenekor
Journal:  J Pediatr Orthop       Date:  2007 Jan-Feb       Impact factor: 2.324

6.  Direct digital mammography using a scanned-slot CCD imaging system.

Authors:  M J Yaffe
Journal:  Med Prog Technol       Date:  1993

7.  Selenium-based digital radiography in the detection of bone lesions: preliminary experience with experimentally created defects.

Authors:  K Ludwig; T M Link; M Fiebich; B Renger; S Diederich; M Oelerich; H Lenzen; W Heindel
Journal:  Radiology       Date:  2000-07       Impact factor: 11.105

8.  Selenium-based digital radiography versus conventional film-screen radiography of the hands and feet: a subjective comparison.

Authors:  D W Piraino; W J Davros; M Lieber; B J Richmond; J P Schils; M P Recht; P N Grooff; G H Belhobek
Journal:  AJR Am J Roentgenol       Date:  1999-01       Impact factor: 3.959

9.  Digital slot-scan charge-coupled device radiography versus AMBER and Bucky screen-film radiography for detection of simulated nodules and interstitial disease in a chest phantom.

Authors:  Lucia J M Kroft; Jacob Geleijns; Bart J A Mertens; Wouter J H Veldkamp; Harmine M Zonderland; Albert de Roos
Journal:  Radiology       Date:  2004-04       Impact factor: 11.105

10.  Photostimulable phosphor digital radiography of the extremities: diagnostic accuracy compared with conventional radiography.

Authors:  A J Wilson; F A Mann; W A Murphy; B S Monsees; M R Linn
Journal:  AJR Am J Roentgenol       Date:  1991-09       Impact factor: 3.959

View more
  1 in total

1.  Abdominal Digital Radiography with a Novel Post-Processing Technique: Phantom and Patient Studies.

Authors:  Hyein Kang; Eun Sun Lee; Hyun Jeong Park; Byung Kwan Park; Jae Yong Park; Suk-Won Suh
Journal:  Taehan Yongsang Uihakhoe Chi       Date:  2020-05-08
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.