Literature DB >> 19138616

Clinical accuracy of 3 different types of computed tomography-derived stereolithographic surgical guides in implant placement.

Oguz Ozan1, Ilser Turkyilmaz, Ahmet Ersan Ersoy, Edwin A McGlumphy, Stephen F Rosenstiel.   

Abstract

PURPOSE: Presurgical planning is essential to achieve esthetic and functional implants. The goal of this clinical study was to determine the angular and linear deviations at the implant neck and apex between planned and placed implants using stereolithographic (SLA) surgical guides. PATIENTS AND METHODS: A total of 110 implants were placed using SLA surgical guides generated from computed tomography (CT). All patients used the radiographic templates during CT scanning. After obtaining 3-dimensional CT scans, each implant insertion was simulated on the CT images. SLA surgical guides using a rapid prototyping method including a laser beam were used during implant insertion. A new CT scan was made for each patient after implant insertion. Special software was used to match images of the planned and placed implants, and their positions and axes were compared.
RESULTS: The mean angular deviation of all placed implants was 4.1 degrees+/-2.3 degrees, whereas mean linear deviation was 1.11+/-0.7 mm at the implant neck and 1.41+/-0.9 mm at the implant apex compared with the planned implants. The angular deviations of the placed implants compared with the planned implants were 2.91 degrees+/-1.3 degrees, 4.63 degrees+/-2.6 degrees, and 4.51 degrees+/-2.1 degrees for the tooth-supported, bone-supported, and mucosa-supported SLA surgical guides, respectively.
CONCLUSION: The results of this study suggested that stereolithographic surgical guides using CT data may be reliable in implant placement, and tooth-supported SLA surgical guides were more accurate than bone- or mucosa-supported SLA surgical guides.

Entities:  

Mesh:

Year:  2009        PMID: 19138616     DOI: 10.1016/j.joms.2008.09.033

Source DB:  PubMed          Journal:  J Oral Maxillofac Surg        ISSN: 0278-2391            Impact factor:   1.895


  36 in total

1.  The influence of exposure parameters on jawbone model accuracy using cone beam CT and multislice CT.

Authors:  B Vandenberghe; S Luchsinger; J Hostens; E Dhoore; R Jacobs
Journal:  Dentomaxillofac Radiol       Date:  2012-01-26       Impact factor: 2.419

2.  Accuracy of different types of computer-aided design/computer-aided manufacturing surgical guides for dental implant placement.

Authors:  Wei Geng; Changying Liu; Yucheng Su; Jun Li; Yanmin Zhou
Journal:  Int J Clin Exp Med       Date:  2015-06-15

3.  Stereolithographic surgical template: a review.

Authors:  Shilpa Sudesh Dandekeri; M K Sowmya; Shruthi Bhandary
Journal:  J Clin Diagn Res       Date:  2013-09-10

4.  Implant Bed Preparation with an Erbium, Chromium Doped Yttrium Scandium Gallium Garnet (Er,Cr: YSGG) Laser Using Stereolithographic Surgical Guide.

Authors:  Gülin Seymen; Zeynep Turgut; Gizem Berk; Ayşen Bodur
Journal:  J Lasers Med Sci       Date:  2013

5.  Survey of UK dentists regarding the use of CAD/CAM technology.

Authors:  D Tran; M Nesbit; H Petridis
Journal:  Br Dent J       Date:  2016-11-18       Impact factor: 1.626

6.  Digital implant planning and guided implant surgery - workflow and reliability.

Authors:  O Schubert; J Schweiger; M Stimmelmayr; E Nold; J-F Güth
Journal:  Br Dent J       Date:  2019-01-25       Impact factor: 1.626

Review 7.  3D imaging, 3D printing and 3D virtual planning in endodontics.

Authors:  Pratik Shah; B S Chong
Journal:  Clin Oral Investig       Date:  2018-01-12       Impact factor: 3.573

8.  [Research advances in the use of digital surgical guides in implantology].

Authors:  Xiao-Hua Wang; Ai-Peng Liu; Wen-Zheng Deng
Journal:  Hua Xi Kou Qiang Yi Xue Za Zhi       Date:  2020-02-01

9.  [Precise investigation of digital guide plates applied to implant surgery of anterior teeth].

Authors:  Fang-Zhi Lou; Mao-Rui Zhang; Peng-Cheng Rao; Shi-Hong Luo; Meng-Ying Tang; Jin-Gang Xiao
Journal:  Hua Xi Kou Qiang Yi Xue Za Zhi       Date:  2020-04-01

10.  Transcrestal guided sinus lift without grafting materials: a 36 months clinical prospective study.

Authors:  D Spinelli; G DE Vico; R Condò; L Ottria; C Arcuri
Journal:  Oral Implantol (Rome)       Date:  2016-07-25