Effect of decreased implant healing time on bone (re)modeling adjacent to plateaued implants under functional loading in a dog model.

Jing Dai1, Ning Cheng, Richard J Miron, Bin Shi, Xiangrong Cheng, Yufeng Zhang.   

Abstract

OBJECTIVES: The purpose of this study was to evaluate the effect of early functional loading to plateaued implants on bone formation and implant stability in a dog model.
MATERIALS AND METHODS: Early loading (EL), nonloading control, and delayed loading (DL) groups were compared using six beagle dogs under functional loading. The Periotest® values were measured dynamically for 6 weeks. Peri-implant bone architecture was evaluated qualitatively by microcomputed tomography (μCT) and analyzed quantitatively by mineral apposition rates (MAR), bone-to-implant contact (BIC), and bone volumes (BV/TV) after the euthanasia at 3 and 6 weeks after loading.
RESULTS: The EL implants showed poor stability at 1 week, but greater stability at 2 and 4 weeks after loading compared to DL implants. There was no significant difference between MAR of EL and unloaded implants at both time intervals. The EL implants displayed a significantly higher MAR when compared to DL implants at 3-5 weeks. A significantly higher BIC for the DL group was observed when compared to the EL group at 3 weeks following loading, however at 6 weeks; no significant difference between these groups was observed. The EL group gained a higher BIC than the no-treatment control group at 6 weeks.
CONCLUSIONS: For plateaued implant, the decreased healing time (1 week) displays a positive effect on peri-implant bone (re)modeling under functional loading during the early phase. CLINICAL RELEVANCE: The early application of functional loading on plateaued implants can be used clinically to shorten the course of treatment and improve esthetics.

Entities:  

Mesh:

Year:  2013        PMID: 23377840      PMCID: PMC5995475          DOI: 10.1007/s00784-013-0929-z

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.573


  38 in total

1.  Qualitative and quantitative observations of bone tissue reactions to anodised implants.

Authors:  Young-Taeg Sul; Carina B Johansson; Kerstin Röser; Tomas Albrektsson
Journal:  Biomaterials       Date:  2002-04       Impact factor: 12.479

2.  Preliminary 3-dimensional surface texture measurement and early loading results with a microtextured implant surface.

Authors:  Ziv Mazor; Donald K Cohen
Journal:  Int J Oral Maxillofac Implants       Date:  2003 Sep-Oct       Impact factor: 2.804

3.  Effects of implant healing time on crestal bone loss of a controlled-load dental implant.

Authors:  C C Ko; W H Douglas; R DeLong; M D Rohrer; J Q Swift; J S Hodges; K-N An; E L Ritman
Journal:  J Dent Res       Date:  2003-08       Impact factor: 6.116

4.  A biomechanical assessment of the relation between the oral implant stability at insertion and subjective bone quality assessment.

Authors:  Ghada Alsaadi; Marc Quirynen; Katleen Michiels; Reinhilde Jacobs; Daniel van Steenberghe
Journal:  J Clin Periodontol       Date:  2007-04       Impact factor: 8.728

5.  Roles of bone scintigraphy and resonance frequency analysis in evaluating osseointegration of endosseous implant.

Authors:  Yi Zhou; Tao Jiang; Mingbo Qian; Xiangyuan Zhang; Jiawei Wang; Bin Shi; Haibin Xia; Xiangrong Cheng; Yining Wang
Journal:  Biomaterials       Date:  2007-11-05       Impact factor: 12.479

6.  A comparative study of two noninvasive techniques to evaluate implant stability: Periotest and Osstell Mentor.

Authors:  Ji-Su Oh; Su-Gwan Kim; Sung-Chul Lim; Joo L Ong
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2008-12-17

7.  Bone reactions to early occlusal loading of two-stage titanium plasma-sprayed implants: a pilot study in monkeys.

Authors:  A Piattelli; M Corigliano; A Scarano; M Quaranta
Journal:  Int J Periodontics Restorative Dent       Date:  1997-04       Impact factor: 1.840

8.  Damping characteristics of bone-to-implant interfaces. A clinical study with the Periotest device.

Authors:  D van Steenberghe; J Tricio; I Naert; M Nys
Journal:  Clin Oral Implants Res       Date:  1995-03       Impact factor: 5.977

9.  Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man.

Authors:  T Albrektsson; P I Brånemark; H A Hansson; J Lindström
Journal:  Acta Orthop Scand       Date:  1981

10.  Modelling the early phases of bone regeneration around an endosseous oral implant.

Authors:  N Amor; L Geris; J Vander Sloten; H Van Oosterwyck
Journal:  Comput Methods Biomech Biomed Engin       Date:  2009-08       Impact factor: 1.763

View more
  3 in total

1.  Mechanical regulation of bone formation and resorption around implants in a mouse model of osteopenic bone.

Authors:  Zihui Li; Duncan Betts; Gisela Kuhn; Michael Schirmer; Ralph Müller; Davide Ruffoni
Journal:  J R Soc Interface       Date:  2019-03-29       Impact factor: 4.118

2.  Histochemical examination on the peri-implant bone with early occlusal loading after the immediate placement into extraction sockets.

Authors:  Yoshiki Ikeda; Tomoka Hasegawa; Tomomaya Yamamoto; Paulo Henrique Luiz de Freitas; Kimimitsu Oda; Akiko Yamauchi; Atsuro Yokoyama
Journal:  Histochem Cell Biol       Date:  2018-02-12       Impact factor: 4.304

3.  Osseointegration behavior of novel Ti-Nb-Zr-Ta-Si alloy for dental implants: an in vivo study.

Authors:  Xiaona Wang; Xing Meng; Shunli Chu; Xingchen Xiang; Zhenzhen Liu; Jinghui Zhao; Yanmin Zhou
Journal:  J Mater Sci Mater Med       Date:  2016-08-17       Impact factor: 3.896

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.