Literature DB >> 19092005

The low contact stress mobile-bearing total knee replacement: a prospective study with a minimum follow-up of ten years.

G Hooper1, A Rothwell, C Frampton.   

Abstract

We have examined the outcome of 400 consecutive patients who underwent total knee replacement with the Low Contact Stress mobile-bearing system between 1993 and 1994 and were followed up for a minimum of ten years. All operations were performed by surgeons in Christchurch, New Zealand, who used no other knee prosthesis during the study time. At ten years after operation 238 patients (244 knees) remained for independent clinical and radiological assessment. There was a significant improvement (p < 0.001) in the postoperative knee scores at one, three, seven and ten years, although a slight deterioration in the score occurred between seven and ten years which did not reach statistical significance. The survival for polyethylene wear or loosening was 97% (95% CI 96 to 98) and survival using reoperation for any cause was 92% (95% CI 90 to 94) at 12 years. Polyethylene wear was more common in the meniscal-bearing component, with five knees requiring revision and a further eight demonstrating early wear. Osteolysis was not seen in the rotating platform component, but was present in three of the meniscal-bearing knees. There was no association between the radiological alignment at one year and those knees that subsequently required revision. However, there was an association between the overall limb alignment and the Western Ontario McMasters University score (p < 0.001). The Low Contact Stress mobile-bearing total knee replacement has proved to be a reliable implant at ten years when used in primary knee replacement irrespective of the deformity and diagnosis.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19092005     DOI: 10.1302/0301-620X.91B1.20484

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  15 in total

1.  Durability of a cruciate-retaining TKA with modular tibial trays at 20 years.

Authors:  John J Callaghan; Mitchell W Beckert; David W Hennessy; Devon D Goetz; Scott S Kelley
Journal:  Clin Orthop Relat Res       Date:  2013-01       Impact factor: 4.176

2.  Do "premium" joint implants add value?: analysis of high cost joint implants in a community registry.

Authors:  Terence J Gioe; Amit Sharma; Penny Tatman; Susan Mehle
Journal:  Clin Orthop Relat Res       Date:  2011-01       Impact factor: 4.176

3.  Mid-term results with a highly congruous mobile-bearing knee prosthesis.

Authors:  Roger G Lemaire
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2009-08-22       Impact factor: 4.342

4.  Revision total knee arthroplasty with porous-coated metaphyseal sleeves provides radiographic ingrowth and stable fixation.

Authors:  Catherine J Fedorka; Antonia F Chen; Michael R Pagnotto; Lawrence S Crossett; Brian A Klatt
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2017-03-17       Impact factor: 4.342

5.  Tibial component coverage and rotational alignment accuracy after mobile-bearing total knee arthroplasty.

Authors:  Yoshinori Ishii; Hideo Noguchi; Junko Sato; Hana Ishii; Koji Todoroki; Shin-Ichi Toyabe
Journal:  Eur J Orthop Surg Traumatol       Date:  2018-02-10

6.  Survivorship at minimum 10-year follow-up of a rotating-platform, mobile-bearing, posterior-stabilised total knee arthroplasty.

Authors:  Michele Ulivi; Luca Orlandini; Valentina Meroni; Olmo Consonni; Valerio Sansone
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-06-18       Impact factor: 4.342

7.  Preoperative flexion does not influence postoperative flexion after rotating-platform total knee arthroplasty.

Authors:  Robert D Russell; Michael H Huo; Leslie de Jong; Richard E Jones
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-01-12       Impact factor: 4.342

8.  Comparison of long-term clinical outcomes after bilateral mobile-bearing total knee arthroplasties using PCL-retaining and PCL-substituting implants in the same patients.

Authors:  Yoshinori Ishii; Hideo Noguchi; Junko Sato; Tetsuya Sakurai; Shin-Ichi Toyabe
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-04-30       Impact factor: 4.342

9.  Minimum twelve-year follow-up of fixed- vs mobile-bearing total knee arthroplasty: Double blinded randomized trial.

Authors:  Cameron J Killen; Michael P Murphy; William J Hopkinson; Melvyn A Harrington; William H Adams; Harold W Rees
Journal:  J Clin Orthop Trauma       Date:  2019-03-29

10.  Time-dependent clinical results of rotating-platform total knee arthroplasty according to mechanical axis deviation.

Authors:  Han-Jun Lee; Ho-Joong Jung; Young-Bok Jung; Young-Bong Ko; Min-Ku Song; Seong-Hwan Kim
Journal:  Knee Surg Relat Res       Date:  2014-08-29
View more