Evaluation of the effects of processing delays and protective plastic cases on image quality of a photostimulable phosphor plate system.

Clóvis Monteiro Bramante1, Alexandre Silva Bramante, Rogério Emílio de Souza, Ivaldo Gomes Moraes, Norberti Bernardineli, Roberto Brandão Garcia.   

Abstract

This ex vivo study evaluated the quality of digital radiographic images obtained with the photostimulable phosphor plate system (Digora) according to the processing delay and maintenance of optical plates in either opaque (supplied with the system) or transparent protective plastic cases during this period. Five radiographs were obtained from the mandibular molar region of a dry human mandible using optical plates. These plates were placed in the protective plastic cases before obtaining the radiographs and were processed immediately or after processing delays of 5, 60 and 120 min, when the case was removed. The results revealed a reduction in image quality when processing was delay 120 min compared to the other times. The opaque case provided better protection to the sensor than the transparent case. In conclusion, a 120-min processing delay for the Digora system caused a reduction in image quality, yet without interfering with the quality of diagnosis. The opaque case supplied by the system's manufacturer provided better protection to the optical plate than the transparent case.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19089233      PMCID: PMC4327602          DOI: 10.1590/s1678-77572008000500009

Source DB:  PubMed          Journal:  J Appl Oral Sci        ISSN: 1678-7757            Impact factor:   2.698


INTRODUCTION

Digital radiograph or radiovisiography was first described in 1987 and was commercially introduced by Trophy Radiologie in collaboration with Mouyen12 in 1993. Three systems are employed for radiovisiography : the CCD sensor (Charge-Coupled Device), CMOS sensor (Complementary Metallic Oxide Sensor), and PSPs (PSPs – Phosphor Storage Plate – optical plate); the latter includes the Digora, Digident, Denoptix and Den Ortix systems11. Some studies have found better outcomes of digitized images for endodontic diagnosis and treatment5–9,13,15–18. However, some doubts remain in the use of radiovisiography based on the optical plate system as to the durability of the plate3, image quality according to the processing delay1,2,14, and storage conditions10. No study has yet investigated the influence of utilization of opaque cases on the quality of digital image. This ex vivo study evaluated the quality of digital radiographic images obtained with the Digora system according to the processing delay and maintenance of optical plates in either opaque (supplied with the system) or transparent protective plastic cases during this period.

MATERIAL AND METHODS

The study was conducted on a dry human mandible fixed on an acrylic resin base. A plastic tube was fixed frontally to the mandible to standardize the position of the X-ray source, thus standardizing the vertical and horizontal angles (Figure 1). A groove was prepared on this resin base to standardize the position of the optical plate.
FIGURE 1

Mandible on acrylic base for standardization of x-ray incidence angle and position of the optical plate

Radiographs were obtained from the mandibular molar region by utilization of optical plates with an x-ray unit (Gnatus model XR 6010; Gnatus, Ribeirão Preto, SP, Brazil) set at 60 kVp and 10 mA, with open tube, total filtration of 1.5 mm Al, exposure time of 0.3 s, and optical plate of the Digora system (Soredex, Orion Corporation, Finland). Two groups were constituted for x-ray exposure: in group A, the optical plate was inserted and sealed in the opaque plastic case supplied by the manufacturer and kept as such until processing; in group B, the optical plate was stored in a transparent plastic case (Figure 2). Five optical plates were exposed for each period of processing delay.
FIGURE 2

Transparent and opaque plastic cases for the optical plates

The optical plates were processed in the Digora scanner immediately and 5, 60 and 120 min after X-ray exposure, and were transferred to the computer for analysis of digitized images. Analysis was performed directly on the computer screen by three examiners using the Digora for Windows software. The images were analyzed as to brightness, contrast and resolution of tooth structures (enamel, dentin, root and pulp cavity) and periapical structures (periodontal space, cortical bone, alveolar bone), which received scores 0 to 2, as follows: 0- Poor image quality; 1- Good image quality; 2- Excellent image quality. Data were submitted to the Kruskal-Wallis non-parametric test, and comparison of results between opaque and transparent cases were submitted by the Mann-Whitney test. Significance level was set at 5%.

RESULTS

Table 1 presents the scores assigned by examiners to the images, according to processing delays and utilization of opaque or transparent plastic cases.
TABLE 1

Scores (0 to 2) assigned to radiographic images (1 to 5) by examiners (I, II and III), according to the type of plastic case (A-opaque and B-transparent) and processing delay (min.)

Processing delay (min)
Immediate Images5 Images60 Images120 Images
12345123451234512345
A
OpaqueExaminer I22222222222212211122
caseExaminer II22222222212122111122
Examiner III22222222222222220222
B
TransparentExaminer I22222212222221201012
caseExaminer II22222221222221220102
Examiner III22222212222112221011
There was statistically significant difference (p<0.05) regarding the processing delays, namely immediate, 5 and 60 min with utilization of opaque case (Table 2). This difference was observed between the 120-min delay and the other periods (Table 3). No statistically significant difference (p>0.05) was observed among processing delays with utilization of transparent case (Table 4). Comparison of results between opaque and transparent cases did not reveal significant differences (p>0.05) (Table 5).
TABLE 2

Kruskal Wallis test for analysis of the processing delay of optical plates protected with opaque cases

Processing delayMedianSum of ranksMean ranksNumber of values
Immediate2.060.012.05
5 min2.060.012.05
60 min2.060.012.05
120 min1.030.06.05

10.05882; Chi-square with 3 degrees of freedom; Probability = 0.018072 (significant)

TABLE 3

Mann Whitney test for analysis of the processing delay of optical plates protected with opaque cases

Processing delayMeanMedianSum of ranksMean ranksNumber
60 min2.0235.07.05
120 min1.4120.04.05

U=5.00000000;Exact probability = 0.150794; Normal approximation: Z=1.96396101; Probability = 0.04953461

TABLE 4

Kruskal Wallis test for analysis of the processing delay of optical plates protected with transparent cases

Processing delayMedianSum of ranksMean ranksNumber of values
Immediate2.065.013.05
5 min2.055.511.15
60 min2.055.511.15
120 min1.034.06.85

Hc= 5.164999; Chi-square with 3 degrees of freedom; Probability = 0. 160106 (non significant)

TABLE 5

Mann Whitney test for comparison between opaque and transparent cases used for protection of optical plates

CaseMeanMedianSum of ranksMean ranksNumber
Opaque1.852.0431.521.620
Transparent1.702.0388.519.420

U=178.5. Normal approximation: Z=0.83688636; Probability = 0.40265645 (non significant)

10.05882; Chi-square with 3 degrees of freedom; Probability = 0.018072 (significant) U=5.00000000;Exact probability = 0.150794; Normal approximation: Z=1.96396101; Probability = 0.04953461 Hc= 5.164999; Chi-square with 3 degrees of freedom; Probability = 0. 160106 (non significant) U=178.5. Normal approximation: Z=0.83688636; Probability = 0.40265645 (non significant)

DISCUSSION

Digital radiography was introduced to dental practice with a view to replace the conventional radiography due to the shorter exposure to radiation, achievement of high-quality images, possibility of adjustments with aid of software, and comparable reliability to conventional radiography, for both endodontic diagnosis and treatment3,5–7,19 The main disadvantage of the optical plate system is the need of additional time for processing and possible loss of quality when the sensitized optical plate is exposed to light. Some authors have emphasized that, after exposure to x-ray, the optical plates should be processed within 1 h in order to avoid loss of image quality1,2. Other authors believe that the plates can be processed within 6 h if stored in appropriate cases10. In the present study, a processing delay of 120 min caused a mild reduction in image quality, yet without compromising the quality of diagnosis (Table 1). The three examiners were able to interpret radiographic details in all study periods. It should be highlighted that the images were not altered as to the brightness and contrast provided by the system, being analyzed as produced on the computer screen. Together with the Digora system, the manufacturer supplies opaque plastic cases that offer protection against contamination and the deleterious effect of light. Images of optical plates maintained in the original opaque cases after exposure to x-ray maintained their image quality up to 120 min (Table 2), with significant difference between the 120- min processing delay and the other periods (Table 3). When the optical plates were maintained in transparent cases, there was also a progressive reduction of image quality (Table 1), which was more accentuated at 120 min (Table 4), though without statistical difference among periods. This is due to the fact that the optical plate protected with transparent cases is subjected to the continuous action of light, with progressive loss of image quality. However, images obtained with optical plates protected with both opaque and transparent cases and with processing delays of up to 120 min provided good quality for diagnosis.

CONCLUSIONS

The findings of the present study revealed that the processing delay of 120 min for the Digora system caused a reduction in image quality, yet without interfering with the quality of diagnosis. The opaque case supplied by the system's manufacturer provided better protection to the optical plate than the transparent case.
  17 in total

1.  Comparative study of two digital radiographic storage phosphor systems.

Authors:  A E Oliveira; S M de Almeida; G A Paganini; F Haiter Neto; F N Bóscolo
Journal:  Braz Dent J       Date:  2000

2.  Conventional and digital radiographic methods in the detection of simulated external root resorptions: a comparative study.

Authors:  V P D Westphalen; I Gomes de Moraes; F H Westphalen; W D Martins; P H Couto Souza
Journal:  Dentomaxillofac Radiol       Date:  2004-07       Impact factor: 2.419

3.  Conventional radiographs vs digitized radiographs: image quality assessment.

Authors:  N Parissis; A Kondylidou-Sidira; A Tsirlis; P Patias
Journal:  Dentomaxillofac Radiol       Date:  2005-11       Impact factor: 2.419

4.  Effect of delayed scanning of storage phosphor plates.

Authors:  B Güniz Akdeniz; Hans-Göran Gröndahl; Timur Kose
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2005-05

Review 5.  Digital and advanced imaging in endodontics: a review.

Authors:  Madhu K Nair; Umadevi P Nair
Journal:  J Endod       Date:  2007-01       Impact factor: 4.171

6.  The imaging performance of a storage phosphor system for dental radiography.

Authors:  D S Brettle; A Workman; R P Ellwood; J H Launders; K Horner; R M Davies
Journal:  Br J Radiol       Date:  1996-03       Impact factor: 3.039

7.  Presentation and physical evaluation of RadioVisioGraphy.

Authors:  F Mouyen; C Benz; E Sonnabend; J P Lodter
Journal:  Oral Surg Oral Med Oral Pathol       Date:  1989-08

8.  Radiographic determination of canal length direct digital radiography versus conventional radiography.

Authors:  R T Hedrick; S B Dove; D D Peters; W D McDavid
Journal:  J Endod       Date:  1994-07       Impact factor: 4.171

9.  Intra-oral storage phosphor and conventional radiography in the assessment of alveolar bone structures.

Authors:  G Kaeppler; A Vogel; D Axmann-Krcmar
Journal:  Dentomaxillofac Radiol       Date:  2000-11       Impact factor: 2.419

10.  Assessment of image quality in dental radiography, part 2: optimum exposure conditions for detection of small mass changes in 6 intraoral radiography systems.

Authors:  K Yoshiura; T Kawazu; T Chikui; M Tatsumi; K Tokumori; T Tanaka; S Kanda
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  1999-01
View more
  5 in total

1.  The effect of delayed scanning of storage phosphor plates on occlusal caries detection.

Authors:  E Sogur; B G Baksi; A Mert
Journal:  Dentomaxillofac Radiol       Date:  2012-01-26       Impact factor: 2.419

2.  Effect of Different Tube Potential Settings on Caries Detection using PSP Plate and Conventional Film.

Authors:  Daniela Pita De Melo; Adriana Dibo Cruz; Saulo Leonardo Sousa Melo; Julyanna Filgueiras GonçAlves De Farias; Francisco Haiter-Neto; Solange Maria De Almeida
Journal:  J Clin Diagn Res       Date:  2015-04-01

3.  Evaluation of the Effect of Light and Scanning Time Delay on The Image Quality of Intra Oral Photostimulable Phosphor Plates.

Authors:  Amir Eskandarloo; Arman Yousefi; Setareh Soheili; Karim Ghazikhanloo; Payam Amini; Haniyeh Mohammadpoor
Journal:  Open Dent J       Date:  2017-12-29

4.  Diagnosis of Approximal Caries after Delayed Scanning of Photostimulable Phosphor Plates.

Authors:  Zahra Ghoncheh; Yasaman Kheirandish; Hanieh Kaviani; Mohammad Javad Kharazi Fard; Kurosh Sadeghi
Journal:  J Dent (Shiraz)       Date:  2017-09

5.  Comparison of the delay in processing time and protective plastic cases in two phosphor plate systems.

Authors:  Ali Murat Aktan; Mehmet Ertuğrul Ciftçi; Faruk Akgünlü
Journal:  ScientificWorldJournal       Date:  2012-11-07
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.