Abstract
This ex vivo study evaluated the quality of digital radiographic images obtained with the photostimulable phosphor plate system (Digora) according to the processing delay and maintenance of optical plates in either opaque (supplied with the system) or transparent protective plastic cases during this period. Five radiographs were obtained from the mandibular molar region of a dry human mandible using optical plates. These plates were placed in the protective plastic cases before obtaining the radiographs and were processed immediately or after processing delays of 5, 60 and 120 min, when the case was removed. The results revealed a reduction in image quality when processing was delay 120 min compared to the other times. The opaque case provided better protection to the sensor than the transparent case. In conclusion, a 120-min processing delay for the Digora system caused a reduction in image quality, yet without interfering with the quality of diagnosis. The opaque case supplied by the system's manufacturer provided better protection to the optical plate than the transparent case.Entities:
Mesh:
Substances:
Year: 2008 PMID: 19089233 PMCID: PMC4327602 DOI: 10.1590/s1678-77572008000500009
Source DB: PubMed Journal: J Appl Oral Sci ISSN: 1678-7757 Impact factor: 2.698
FIGURE 1Mandible on acrylic base for standardization of x-ray incidence angle and position of the optical plate
FIGURE 2Transparent and opaque plastic cases for the optical plates
Scores (0 to 2) assigned to radiographic images (1 to 5) by examiners (I, II and III), according to the type of plastic case (A-opaque and B-transparent) and processing delay (min.)
| Processing delay (min) | |||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Immediate Images | 5 Images | 60 Images | 120 Images | ||||||||||||||||||
| 1 | 2 | 3 | 4 | 5 | 1 | 2 | 3 | 4 | 5 | 1 | 2 | 3 | 4 | 5 | 1 | 2 | 3 | 4 | 5 | ||
| A | |||||||||||||||||||||
| Opaque | Examiner I | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 1 | 2 | 2 | 1 | 1 | 1 | 2 | 2 |
| case | Examiner II | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 1 | 2 | 1 | 2 | 2 | 1 | 1 | 1 | 1 | 2 | 2 |
| Examiner III | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 0 | 2 | 2 | 2 | |
| B | |||||||||||||||||||||
| Transparent | Examiner I | 2 | 2 | 2 | 2 | 2 | 2 | 1 | 2 | 2 | 2 | 2 | 2 | 2 | 1 | 2 | 0 | 1 | 0 | 1 | 2 |
| case | Examiner II | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 1 | 2 | 2 | 2 | 2 | 2 | 1 | 2 | 2 | 0 | 1 | 0 | 2 |
| Examiner III | 2 | 2 | 2 | 2 | 2 | 2 | 1 | 2 | 2 | 2 | 2 | 1 | 1 | 2 | 2 | 2 | 1 | 0 | 1 | 1 | |
Kruskal Wallis test for analysis of the processing delay of optical plates protected with opaque cases
| Processing delay | Median | Sum of ranks | Mean ranks | Number of values |
|---|---|---|---|---|
| Immediate | 2.0 | 60.0 | 12.0 | 5 |
| 5 min | 2.0 | 60.0 | 12.0 | 5 |
| 60 min | 2.0 | 60.0 | 12.0 | 5 |
| 120 min | 1.0 | 30.0 | 6.0 | 5 |
10.05882; Chi-square with 3 degrees of freedom; Probability = 0.018072 (significant)
Mann Whitney test for analysis of the processing delay of optical plates protected with opaque cases
| Processing delay | Mean | Median | Sum of ranks | Mean ranks | Number |
|---|---|---|---|---|---|
| 60 min | 2.0 | 2 | 35.0 | 7.0 | 5 |
| 120 min | 1.4 | 1 | 20.0 | 4.0 | 5 |
U=5.00000000;Exact probability = 0.150794; Normal approximation: Z=1.96396101; Probability = 0.04953461
Kruskal Wallis test for analysis of the processing delay of optical plates protected with transparent cases
| Processing delay | Median | Sum of ranks | Mean ranks | Number of values |
|---|---|---|---|---|
| Immediate | 2.0 | 65.0 | 13.0 | 5 |
| 5 min | 2.0 | 55.5 | 11.1 | 5 |
| 60 min | 2.0 | 55.5 | 11.1 | 5 |
| 120 min | 1.0 | 34.0 | 6.8 | 5 |
Hc= 5.164999; Chi-square with 3 degrees of freedom; Probability = 0. 160106 (non significant)
Mann Whitney test for comparison between opaque and transparent cases used for protection of optical plates
| Case | Mean | Median | Sum of ranks | Mean ranks | Number |
|---|---|---|---|---|---|
| Opaque | 1.85 | 2.0 | 431.5 | 21.6 | 20 |
| Transparent | 1.70 | 2.0 | 388.5 | 19.4 | 20 |
U=178.5. Normal approximation: Z=0.83688636; Probability = 0.40265645 (non significant)