Radiation dose values for various coronary calcium scoring protocols in dual-source CT.

Paul Stolzmann1, Sebastian Leschka, Thomas Betschart, Lotus Desbiolles, Thomas G Flohr, Borut Marincek, Hatem Alkadhi.   

Abstract

PURPOSE: The purpose of this study was to assess the radiation dose and associated image noise of previously suggested calcium scoring protocols using dual-source CT.
METHODS: One hundred consecutive patients underwent coronary calcium scoring using dual-source CT. Patients were randomly assigned to five different protocols: retrospective ECG-gating and tube current reduction to 4% outside the pulsing window at 120 (protocol A) and 100 kV (B), prospective ECG-triggering at 120 (C) and 100 kV (D), and prospective ECG-triggering at 100 kV with attenuation-based tube current modulation (E). Radiation dose parameters and image noise were determined and compared.
RESULTS: Protocol A resulted in an effective dose of 1.3 +/- 0.2 mSv, protocol B in 0.8 +/- 0.2 mSv, protocol C in 1.0 +/- 0.2 mSv, protocol D in 0.6 +/- 0.1 mSv, and protocol E in 0.7 +/- 0.1 mSv. Effective doses were significantly lower (P < 0.001) with 100 kV when compared to 120 kV protocols, and were significantly lower (P < 0.001) for prospective versus retrospective ECG-gating. No significant difference was found between protocol D and E. Significant negative correlations were found between the CTDI(vol) and heart rate for both retrospective ECG-gating protocols (protocol A: r = -0.98, P < 0.001; protocol B: r = -0.83, P < 0.001). The mean image noise was 29.0 +/- 6.7 HU, with no significant differences between the five protocols. The image noise was significantly correlated with the body weight (r = 0.21, P < 0.05) and BMI (r = 0.31, P < 0.01).
CONCLUSIONS: Effective dose of calcium scoring using dual-source CT ranges from 0.6 to 1.3 mSv. Prospective triggering and lower tube voltage significantly reduces the radiation but yield similar image noise.

Entities:  

Mesh:

Year:  2008        PMID: 19082870     DOI: 10.1007/s10554-008-9397-y

Source DB:  PubMed          Journal:  Int J Cardiovasc Imaging        ISSN: 1569-5794            Impact factor:   2.357


  37 in total

1.  Reproducibility and accuracy of coronary calcium measurements with multi-detector row versus electron-beam CT.

Authors:  Andreas F Kopp; Bernd Ohnesorge; Christoph Becker; Stephen Schröder; Martin Heuschmid; Axel Küttner; Ronald Kuzo; Claus D Claussen
Journal:  Radiology       Date:  2002-10       Impact factor: 11.105

2.  Calcified coronary artery plaque measurement with cardiac CT in population-based studies: standardized protocol of Multi-Ethnic Study of Atherosclerosis (MESA) and Coronary Artery Risk Development in Young Adults (CARDIA) study.

Authors:  J Jeffrey Carr; Jennifer Clark Nelson; Nathan D Wong; Michael McNitt-Gray; Yadon Arad; David R Jacobs; Stephan Sidney; Diane E Bild; O Dale Williams; Robert C Detrano
Journal:  Radiology       Date:  2005-01       Impact factor: 11.105

3.  Variability of repeated coronary artery calcium measurements on low-dose ECG-gated 16-MDCT.

Authors:  Jun Horiguchi; Hideya Yamamoto; Nobuhiko Hirai; Yuji Akiyama; Chikako Fujioka; Kazushi Marukawa; Hiroshi Fukuda; Katsuhide Ito
Journal:  AJR Am J Roentgenol       Date:  2006-07       Impact factor: 3.959

4.  Image quality and reconstruction intervals of dual-source CT coronary angiography: recommendations for ECG-pulsing windowing.

Authors:  Sebastian Leschka; Hans Scheffel; Lotus Desbiolles; Andre Plass; Oliver Gaemperli; Ines Valenta; Lars Husmann; Thomas G Flohr; Michele Genoni; Borut Marincek; Philipp A Kaufmann; Hatem Alkadhi
Journal:  Invest Radiol       Date:  2007-08       Impact factor: 6.016

5.  Dual-source CT coronary angiography: image quality, mean heart rate, and heart rate variability.

Authors:  David Matt; Hans Scheffel; Sebastian Leschka; Thomas G Flohr; Borut Marincek; Philipp A Kaufmann; Hatem Alkadhi
Journal:  AJR Am J Roentgenol       Date:  2007-09       Impact factor: 3.959

6.  The influence of heart rate, slice thickness, and calcification density on calcium scores using 64-slice multidetector computed tomography: a systematic phantom study.

Authors:  Jaap M Groen; Marcel J Greuter; Bernhard Schmidt; Christoph Suess; Rozemarijn Vliegenthart; Matthijs Oudkerk
Journal:  Invest Radiol       Date:  2007-12       Impact factor: 6.016

7.  Detection of coronary artery stenoses with thin-slice multi-detector row spiral computed tomography and multiplanar reconstruction.

Authors:  Dieter Ropers; Ulrich Baum; Karsten Pohle; Katharina Anders; Stefan Ulzheimer; Bernd Ohnesorge; Christian Schlundt; Werner Bautz; Werner G Daniel; Stephan Achenbach
Journal:  Circulation       Date:  2003-02-11       Impact factor: 29.690

8.  Detection of coronary calcifications: feasibility of dose reduction with a body weight-adapted examination protocol.

Authors:  A H Mahnken; J E Wildberger; J Simon; R Koos; T G Flohr; S Schaller; R W Günther
Journal:  AJR Am J Roentgenol       Date:  2003-08       Impact factor: 3.959

9.  Ultra-low-dose coronary artery calcium screening using multislice CT with retrospective ECG gating.

Authors:  Tobias F Jakobs; Bernd J Wintersperger; Peter Herzog; Thomas Flohr; Christoph Suess; Andreas Knez; Maximilian F Reiser; Christoph R Becker
Journal:  Eur Radiol       Date:  2003-05-21       Impact factor: 5.315

10.  Non-invasive cardiac assessment in high risk patients (The GROUND study): rationale, objectives and design of a multi-center randomized controlled clinical trial.

Authors:  Alexander M de Vos; Annemarieke Rutten; Hester J van de Zaag-Loonen; Michiel L Bots; Riksta Dikkers; Robert A Buiskool; Willem P Mali; Daniel D Lubbers; Arend Mosterd; Mathias Prokop; Benno J Rensing; Maarten J Cramer; H Wouter van Es; Frans L Moll; Eric D van de Pavoordt; Pieter A Doevendans; Birgitta K Velthuis; Albert J Mackaay; Felix Zijlstra; Matthijs Oudkerk
Journal:  Trials       Date:  2008-08-01       Impact factor: 2.279

View more
  4 in total

1.  Absorbed radiation dose in radiosensitive organs during coronary CT angiography using 320-MDCT: effect of maximum tube voltage and heart rate variations.

Authors:  Boris Nikolic; Faisal Khosa; Pei-Jan Paul Lin; Atif N Khan; Sheryar Sarwar; Chun-Shan Yam; Laurence E Court; Vassilios Raptopoulos; Melvin E Clouse
Journal:  AJR Am J Roentgenol       Date:  2010-12       Impact factor: 3.959

2.  Reproducibility of coronary artery calcium measurements using 0.8-mm-thickness 256-slice coronary CT.

Authors:  Soe Hee Ann; Jae Hee Kim; Nam Du Ha; Seong Hoon Choi; Scot Garg; Gillian Balbir Singh; Eun-Seok Shin
Journal:  Jpn J Radiol       Date:  2014-10-03       Impact factor: 2.374

3.  Non-enhanced cardiac computed tomography-still an open book.

Authors:  Ronak Rajani; Damini Dey; Daniel S Berman
Journal:  J Nucl Cardiol       Date:  2011-02       Impact factor: 5.952

4.  Diagnostic efficacy of vessel specific coronary calcium score in detection of coronary artery stenosis.

Authors:  Marzieh Motevalli; Hossein Ghanaati; Kavous Firouznia; Jalal Kargar; Mounes Aliyari Ghasabeh; Mona Shahriari; Amir Hosein Jalali; Madjid Shakiba
Journal:  Iran Red Crescent Med J       Date:  2014-12-30       Impact factor: 0.611

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.