The impact of the distance between the atrial electrode and the atrial wall on atrial undersensing in patients with VDD pacemakers: long-term follow-up.

Timucin Altin1, Muharrem Guldal, Basar Candemir, Cegerhun Polat, Aydan O Ozdemir, Cansin Tulunay, Cagdas Ozdol, Omer Akyurek, Remzi Karaoguz, Cetin Erol.   

Abstract

AIM: Atrial undersensing (AUS) in single-lead VDD pacemakers may be due to diminished P-wave amplitude secondary to local inflammation beneath the electrodes closer to atrial wall. The aim of this study was to assess the potential effect of distance between atrial electrode and atrial wall on immediate and long-term atrial sensing stability in VDD systems.
METHODS: A total of 275 patients with normal sinus node function who received VDD pacemakers were enrolled into the study and were followed up for a median duration of 33 months. During each control visit, a standard 12-lead electrocardiogram (ECG) was obtained and standard pacemaker function assessment was performed including testing for pacing threshold and atrioventricular synchrony. The distance between atrial electrode and atrial wall was measured from chest X-ray.
RESULTS: Of the 275 patients, AUS was detected in 59 patients. Univariate predictors of AUS were use of closely spaced bipolar ring atrial electrode (CSBR) (P = 0.01), wider atrial ring-spacing (P = 0.03), and atrial sensitivity programmed to a higher level (P = 0.001). Use of CSBR (P = 0.04) and atrial sensitivity > or =0.3 mV (P = 0.02) were observed to be the independent predictors for AUS. When the distance between atrial electrode and atrial wall was <7 mm, AUS was less with diagonally arranged bipolar ring electrodes (DABR) than it was with CSBRs (P = 0.02).
CONCLUSIONS: The distance between atrial electrode and atrial wall does not appear to affect AUS incidence in VDD pacemakers. For VDD electrodes closer to atrial wall, AUS was significantly less likely in DABR-type electrodes.

Entities:  

Mesh:

Year:  2008        PMID: 18973490      PMCID: PMC6932008          DOI: 10.1111/j.1542-474X.2008.00243.x

Source DB:  PubMed          Journal:  Ann Noninvasive Electrocardiol        ISSN: 1082-720X            Impact factor:   1.468


  16 in total

1.  Can chest roentgenogram predict the posture dependent changes of atrial sensing performance in patients with a VDD pacemaker?

Authors:  Kee-Joon Choi; Gi-Byoung Nam; Jun Kim; Kyoung-Suk Rhee; You-Ho Kim
Journal:  Pacing Clin Electrophysiol       Date:  2004-10       Impact factor: 1.976

2.  Dependence of atrial sensing function on posture in a single-lead atrial triggered ventricular (VDD) pacemaker.

Authors:  L Toivonen; J Lommi
Journal:  Pacing Clin Electrophysiol       Date:  1996-03       Impact factor: 1.976

3.  Toward optimizing the detection of atrial depolarization with floating bipolar electrodes.

Authors:  R R Brownlee
Journal:  Pacing Clin Electrophysiol       Date:  1989-03       Impact factor: 1.976

4.  Atrial sensing and AV synchrony in single lead VDD pacemakers: a prospective comparison to DDD devices with bipolar atrial leads.

Authors:  U K Wiegand; F Bode; R Schneider; G Taubert; A Brandes; W Peters; H A Katus; J Potratz
Journal:  J Cardiovasc Electrophysiol       Date:  1999-04

Review 5.  Single lead atrial synchronous ventricular pacing: a dream come true.

Authors:  G E Antonioli
Journal:  Pacing Clin Electrophysiol       Date:  1994-09       Impact factor: 1.976

6.  Immediate and long-term atrial sensing stability in single-lead VDD pacing depends on right atrial dimensions.

Authors:  M Santini; R Ricci; C Pignalberi; A Auriti; M Pepe; R Assale; D Caporicci
Journal:  Europace       Date:  2001-10       Impact factor: 5.214

7.  Clinical relevance of loss of atrial sensing in patients with single lead VDD pacemakers.

Authors:  C M Van Campen; C C De Cock; J Huijgens; C A Visser
Journal:  Pacing Clin Electrophysiol       Date:  2001-05       Impact factor: 1.976

8.  Predictors of loss of atrioventricular synchrony in single lead VDD pacing.

Authors:  P Hunziker; P Buser; M Pfisterer; F Burkart; S Osswald
Journal:  Heart       Date:  1998-10       Impact factor: 5.994

Review 9.  The current status of single lead dual chamber sensing and pacing.

Authors:  H F Tse; C P Lau
Journal:  J Interv Card Electrophysiol       Date:  1998-09       Impact factor: 1.900

10.  Long-term stability of P wave sensing in single lead VDDR pacing: clinical versus subclinical atrial undersensing.

Authors:  C P Lau; Y T Tai; S K Leung; W H Leung; F L Chung; I S Lee
Journal:  Pacing Clin Electrophysiol       Date:  1994-11       Impact factor: 1.976

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.