Atrial sensing and AV synchrony in single lead VDD pacemakers: a prospective comparison to DDD devices with bipolar atrial leads.

U K Wiegand1, F Bode, R Schneider, G Taubert, A Brandes, W Peters, H A Katus, J Potratz.   

Abstract

INTRODUCTION: Single lead VDD pacing has offered an alternative to DDD systems in patients with isolated AV block. Up to now, however, the relative performance of these pacemaker systems was not systematically compared. METHODS AND
RESULTS: Three hundred sixty patients who received either a VDD pacemaker (n = 180) or a DDD device (n = 180) with a bipolar atrial lead were investigated prospectively for a mean period of 30 +/- 13 months. Pacemaker function was analyzed by telemetry, Holter monitoring, and exercise ECG. Time of implantation and fluoroscopy was significantly lower with VDD devices (44.3 +/- 5.1 min vs 74.4 +/- 13.5 min and 4.6 +/- 2.5 min vs 10.3 +/- 5.6 min in DDD pacemakers, respectively). Intermittent atrial undersensing occurred in 23.3% of patients with a VDD pacemaker and in 9.4% with DDD devices (NS). The incidence of atrial tachyarrhythmias did not differ between the VDD (6.7%) and the DDD group (6.1%). Sinus node dysfunction developed in 1.9% of patients, but the vast majority (85.7%) of patients were asymptomatic. There was a tendency for a higher rate of operative revisions in the DDD group (6.1% vs 3.3% in VDD pacemakers, P = 0.15). Cumulative maintenance of AV-synchronized pacing mode was 94.9% in patients with VDD pacemakers and 92.1% with DDD devices (NS).
CONCLUSION: With the benefit of a simpler implant procedure, long-term outcome of single lead VDD pacing is equivalent to DDD pacing in patients with AV block and preoperative normal sinus node function.

Entities:  

Mesh:

Year:  1999        PMID: 10355692     DOI: 10.1111/j.1540-8167.1999.tb00707.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  8 in total

1.  VDD pacemaker replacement is safe and reliable independently of the previously implanted lead: a prospective and controlled study.

Authors:  Alessandro Fabiani; Antonio Burali; Enrico Manfredini; Giorgio Corbucci; Leonardo Bolognese; S Serge Barold
Journal:  J Interv Card Electrophysiol       Date:  2006-03       Impact factor: 1.900

2.  [Not Available].

Authors:  U K Wiegand; F Bode; H Bonnemeier; R Tölg; W Peters; H A Katus
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2000-01

3.  [Not Available].

Authors:  A Schuchert; C Müller; K Malinowski; K H Seidl; T Hilbel; T Meinertz; J Brachmann; C Stellbrink; H U Altenburg; B D Gonska; M Zehender; F Goss; G Hoffmann; W Karman; R Kolb; H G Olbrich; M Rachor; W Sondern; H Roth; B Uhlmann
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2000-01

4.  Long term complications in single and dual chamber pacing are influenced by surgical experience and patient morbidity.

Authors:  F Eberhardt; F Bode; H Bonnemeier; F Boguschewski; M Schlei; W Peters; U K H Wiegand
Journal:  Heart       Date:  2005-04       Impact factor: 5.994

5.  Long-term outcome of atrial synchronous mode pacing in patients with atrioventricular block using a single lead.

Authors:  Miry Blich; Mahmoud Suleiman; Tawfiq Zeidan Shwiri; Ibrahim Marai; Monther Boulos; Shlomo Amikam
Journal:  Clin Cardiol       Date:  2010-01       Impact factor: 2.882

6.  The impact of the distance between the atrial electrode and the atrial wall on atrial undersensing in patients with VDD pacemakers: long-term follow-up.

Authors:  Timucin Altin; Muharrem Guldal; Basar Candemir; Cegerhun Polat; Aydan O Ozdemir; Cansin Tulunay; Cagdas Ozdol; Omer Akyurek; Remzi Karaoguz; Cetin Erol
Journal:  Ann Noninvasive Electrocardiol       Date:  2008-10       Impact factor: 1.468

7.  Intracardiac atrial signal amplitude in congenital and acquired complete heart block.

Authors:  Mohammad Reza Samiei Nasab; Mohammad Reza Dehghani; Mehrdad Taherioun; Alireza Rostamzadeh
Journal:  Indian Pacing Electrophysiol J       Date:  2010-10-31

8.  Long-term outcome in patients receiving permanent pacemaker implantation for atrioventricular block: Comparison of VDD and DDD pacing.

Authors:  Jo-Nan Liao; Tze-Fan Chao; Ta-Chuan Tuan; Chi-Woon Kong; Shih-Ann Chen
Journal:  Medicine (Baltimore)       Date:  2016-08       Impact factor: 1.889

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.