Closed bore XMR (CBXMR) systems for aortic valve replacement: investigation of rotating-anode x-ray tube heat loadability.

John A Bracken1, Prasheel V Lillaney, Rebecca Fahrig, J A Rowlands.   

Abstract

In order to improve the safety and efficacy of percutaneous aortic valve replacement procedures, a closed bore hybrid x-ray/MRI (CBXMR) system is proposed in which an x-ray C-arm will be positioned with its isocenter approximately =1 m from the entrance of a clinical MRI scanner. This system will harness the complementary strengths of both modalities to improve clinical outcome. A key component of the CBXMR system will be a rotating anode x-ray tube to produce high-quality x-ray images. There are challenges in positioning an x-ray tube in the magnetic fringe field of the MRI magnet. Here, the effects of an external magnetic field on x-ray tube induction motors of radiography x-ray tubes and the corresponding reduction of x-ray tube heat loadability are investigated. Anode rotation frequency f(aode) was unaffected when the external magnetic field Bb was parallel to the axis of rotation of the anode but decreased when Bb was perpendicular to the axis of rotation. The experimental f(anode) values agreed with predicted values to within +/-3% over a Bb range of 0-30 mT. The MRI fringe field at the proposed location of the x-ray tube mounted on the C-arm (approximately =4 mT) reduced f(anode) by only 1%, so x-ray tube heat loadability will not be compromised when using CBXMR systems for percutaneous aortic valve replacement procedures. Eddy current heating power in the rotor due to an MRI fringe field was found to be two orders of magnitude weaker than the heating power produced on the anode due to a fluoroscopic exposure, so eddy current heating had no effect on x-ray tube heat loadability.

Entities:  

Mesh:

Year:  2008        PMID: 18841857      PMCID: PMC2855716          DOI: 10.1118/1.2968337

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  23 in total

1.  A system for real-time XMR guided cardiovascular intervention.

Authors:  Kawal S Rhode; Maxime Sermesant; David Brogan; Sanjeet Hegde; John Hipwell; Pier Lambiase; Eric Rosenthal; Clifford Bucknall; Shakeel A Qureshi; Jaswinder S Gill; Reza Razavi; Derek L G Hill
Journal:  IEEE Trans Med Imaging       Date:  2005-11       Impact factor: 10.048

2.  Robust x-ray tubes for use within magnetic fields of MR scanners.

Authors:  Zhifei Wen; Rebecca Fahrig; Norbert J Pelc
Journal:  Med Phys       Date:  2005-07       Impact factor: 4.071

3.  Performance of a static-anode/flat-panel x-ray fluoroscopy system in a diagnostic strength magnetic field: a truly hybrid x-ray/MR imaging system.

Authors:  R Fahrig; Z Wen; A Ganguly; G DeCrescenzo; J A Rowlands; G M Stevens; R F Saunders; N J Pelc
Journal:  Med Phys       Date:  2005-06       Impact factor: 4.071

4.  Real-time interactive MRI-guided cardiac surgery: aortic valve replacement using a direct apical approach.

Authors:  Elliot R McVeigh; Michael A Guttman; Robert J Lederman; Ming Li; Ozgur Kocaturk; Timothy Hunt; Shawn Kozlov; Keith A Horvath
Journal:  Magn Reson Med       Date:  2006-11       Impact factor: 4.668

5.  Percutaneous aortic valve implantation retrograde from the femoral artery.

Authors:  John G Webb; Mann Chandavimol; Christopher R Thompson; Donald R Ricci; Ronald G Carere; Brad I Munt; Christopher E Buller; Sanjeevan Pasupati; Samuel Lichtenstein
Journal:  Circulation       Date:  2006-02-06       Impact factor: 29.690

Review 6.  Percutaneous valve procedures: what is the future?

Authors:  Alec Vahanian; Christophe Acar
Journal:  Curr Opin Cardiol       Date:  2005-03       Impact factor: 2.161

7.  MR-guided transjugular intrahepatic portosystemic shunt creation with use of a hybrid radiography/MR system.

Authors:  Stephen T Kee; Arundhuti Ganguly; Bruce L Daniel; Zhifei Wen; Kim Butts; Anne Shimikawa; Norbert J Pelc; Rebecca Fahrig; Michael D Dake
Journal:  J Vasc Interv Radiol       Date:  2005-02       Impact factor: 3.464

8.  Transapical transcatheter aortic valve implantation in humans: initial clinical experience.

Authors:  Samuel V Lichtenstein; Anson Cheung; Jian Ye; Christopher R Thompson; Ronald G Carere; Sanjeevan Pasupati; John G Webb
Journal:  Circulation       Date:  2006-07-31       Impact factor: 29.690

9.  Carotid stent delivery in an XMR suite: immediate assessment of the physiologic impact of extracranial revascularization.

Authors:  Alastair J Martin; David A Saloner; Timothy P L Roberts; Heidi Roberts; Oliver M Weber; William Dillon; Sean Cullen; Van Halbach; Christopher F Dowd; Randall T Higashida
Journal:  AJNR Am J Neuroradiol       Date:  2005-03       Impact factor: 3.825

10.  Percutaneous aortic valve replacement for severe aortic stenosis in high-risk patients using the second- and current third-generation self-expanding CoreValve prosthesis: device success and 30-day clinical outcome.

Authors:  Eberhard Grube; Gerhard Schuler; Lutz Buellesfeld; Ulrich Gerckens; Axel Linke; Peter Wenaweser; Barthel Sauren; Friedrich-Wilhelm Mohr; Thomas Walther; Bernfried Zickmann; Stein Iversen; Thomas Felderhoff; Raymond Cartier; Raoul Bonan
Journal:  J Am Coll Cardiol       Date:  2007-06-06       Impact factor: 24.094

View more
  2 in total

1.  Closed-bore XMR (CBXMR) systems for aortic valve replacement: x-ray tube imaging performance.

Authors:  John A Bracken; Philip Komljenovic; Prasheel V Lillaney; Rebecca Fahrig; J A Rowlands
Journal:  Med Phys       Date:  2009-04       Impact factor: 4.071

2.  Closed bore XMR (CBXMR) systems for aortic valve replacement: active magnetic shielding of x-ray tubes.

Authors:  John A Bracken; Giovanni DeCrescenzo; Philip Komljenovic; Prasheel V Lillaney; Rebecca Fahrig; J A Rowlands
Journal:  Med Phys       Date:  2009-05       Impact factor: 4.071

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.