A design methodology using signal-to-noise ratio for plastic scintillation detectors design and performance optimization.

Frédéric Lacroix1, A Sam Beddar, Mathieu Guillot, Luc Beaulieu, Luc Gingras.   

Abstract

PURPOSE: The design of novel plastic scintillation detectors (PSDs) is impeded by the lack of a suitable framework to simulate and predict their performance. The authors propose to use the signal-to-noise ratio (SNR) to model the performance of PSDs that use charge-coupled devices (CCDs) as photodetectors.
METHODS: In PSDs using CCDs, the SNR is inversely related to the normalized standard deviation of the dose measurement. Thus, optimizing the SNR directly optimizes the system's precision. In this work, a model of SNR as a function of the system parameters is derived for optical fiber-based PSD systems. Furthermore, this proposed model is validated using experimental results. A formula for the efficiency of fiber coupling to CCDs is derived and used to simulate the performance of a PSD under varying magnifications.
RESULTS: The proposed model is shown to simulate the experimental performance of an actual PSD to a suitable degree of accuracy under various conditions.
CONCLUSIONS: The SNR constitutes a useful tool to simulate the dosimetric precision of PSDs. Using the SNR model, recommendations for the design and optimization of PSDs are provided. Using the same framework, recommendations for non-fiber-based PSDs are also provided.

Mesh:

Substances:

Year:  2009        PMID: 19994531      PMCID: PMC2774352          DOI: 10.1118/1.3231947

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  18 in total

1.  New water equivalent liquid scintillation solutions for 3D dosimetry.

Authors:  A S Kirov; S Shrinivas; C Hurlbut; J F Dempsey; W R Binns; J L Poblete
Journal:  Med Phys       Date:  2000-05       Impact factor: 4.071

2.  Plastic scintillation dosimetry: optimization of light collection efficiency.

Authors:  A Sam Beddar; Susan Law; Natalka Suchowerska; T Rockwell Mackie
Journal:  Phys Med Biol       Date:  2003-05-07       Impact factor: 3.609

3.  The DOSIMAP, a high spatial resolution tissue equivalent 2D dosimeter for LINAC QA and IMRT verification.

Authors:  V Collomb-Patton; P Boher; T Leroux; J M Fontbonne; A Vela; A Batalla
Journal:  Med Phys       Date:  2009-02       Impact factor: 4.071

4.  Transient noise characterization and filtration in CCD cameras exposed to stray radiation from a medical linear accelerator.

Authors:  Louis Archambault; Tina Marie Briere; Sam Beddar
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

5.  Direct reading measurement of absorbed dose with plastic scintillators--the general concept and applications to ophthalmic plaque dosimetry.

Authors:  D Flühs; M Heintz; F Indenkämpen; C Wieczorek
Journal:  Med Phys       Date:  1996-03       Impact factor: 4.071

6.  Characteristics of a scintillator-based daily quality assurance device for radiation oncology beams.

Authors:  I J Das; M J Gazda; A S Beddar
Journal:  Med Phys       Date:  1996-12       Impact factor: 4.071

7.  Lens coupling efficiency: derivation and application under differing geometrical assumptions.

Authors:  T Yu; J M Boone
Journal:  Med Phys       Date:  1997-04       Impact factor: 4.071

8.  Plastic scintillator response to low-energy photons.

Authors:  J F Williamson; J F Dempsey; A S Kirov; J I Monroe; W R Binns; H Hedtjärn
Journal:  Phys Med Biol       Date:  1999-04       Impact factor: 3.609

9.  Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurements.

Authors:  A S Beddar; T R Mackie; F H Attix
Journal:  Phys Med Biol       Date:  1992-10       Impact factor: 3.609

10.  Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical consideration.

Authors:  A S Beddar; T R Mackie; F H Attix
Journal:  Phys Med Biol       Date:  1992-10       Impact factor: 3.609

View more
  3 in total

1.  Extraction of depth-dependent perturbation factors for parallel-plate chambers in electron beams using a plastic scintillation detector.

Authors:  Frédéric Lacroix; Mathieu Guillot; Malcolm McEwen; Claudiu Cojocaru; Luc Gingras; A Sam Beddar; Luc Beaulieu
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

2.  Simulation of the precision limits of plastic scintillation detectors using optimal component selection.

Authors:  Frédéric Lacroix; Luc Beaulieu; Louis Archambault; A Sam Beddar
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

3.  Quality assurance in proton beam therapy using a plastic scintillator and a commercially available digital camera.

Authors:  Mansour Almurayshid; Yusuf Helo; Andrzej Kacperek; Jennifer Griffiths; Jem Hebden; Adam Gibson
Journal:  J Appl Clin Med Phys       Date:  2017-07-29       Impact factor: 2.102