Understanding the relationship between image quality and motion velocity in gated computed tomography: preliminary work for 4-dimensional musculoskeletal imaging.

Shian Chao Tay1, Andrew N Primak, Joel G Fletcher, Bernhard Schmidt, Kai-Nan An, Cynthia H McCollough.   

Abstract

OBJECTIVES: To study the effect of motion velocity on image quality to determine the requirements for 4-dimensional (4D; ie, 3D + time) musculoskeletal computed tomographic (CT) imaging.
MATERIALS AND METHODS: A phantom with resolution targets in both axial (x-y) and coronal (x-z) planes was attached to a motion device and scanned with 64-slice CT using a retrospectively gated CT protocol with pitch values of 0.1 and 0.2. Data were acquired with the phantom at rest and while moving periodically along the x axis at several velocities. Spatial resolution and motion artifacts were assessed both for the axial and coronal targets.
RESULTS: A linear relationship was found between motion artifact severity and phantom velocity. Spatial resolution was better preserved in the coronal target. However, coronal images displayed banding artifacts, with band displacements being linearly related to motion velocity.
CONCLUSIONS: The 4D CT imaging of periodically moving objects with velocities up to 20 mm/s is feasible using a pitch value of 0.1 and a motion frequency of 30 cycles per minute.

Entities:  

Mesh:

Year:  2008        PMID: 18664854      PMCID: PMC2744860          DOI: 10.1097/RCT.0b013e31815c5abc

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  10 in total

1.  Cardiac imaging by means of electrocardiographically gated multisection spiral CT: initial experience.

Authors:  B Ohnesorge; T Flohr; C Becker; A F Kopp; U J Schoepf; U Baum; A Knez; K Klingenbeck-Regn; M F Reiser
Journal:  Radiology       Date:  2000-11       Impact factor: 11.105

2.  4D rotational x-ray imaging of wrist joint dynamic motion.

Authors:  Bart Carelsen; Niels H Bakker; Simon D Strackee; Sjirk N Boon; Mario Maas; Jörg Sabczynski; Cornelis A Grimbergen; Geert J Streekstra
Journal:  Med Phys       Date:  2005-09       Impact factor: 4.071

3.  Kinematics of the midcarpal and radiocarpal joints in radioulnar deviation: an in vitro study.

Authors:  Robert Kaufmann; Jamie Pfaeffle; Brad Blankenhorn; Kathryne Stabile; Doug Robertson; Robert Goitz
Journal:  J Hand Surg Am       Date:  2005-09       Impact factor: 2.230

4.  Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot.

Authors:  T G Flohr; K Stierstorfer; S Ulzheimer; H Bruder; A N Primak; C H McCollough
Journal:  Med Phys       Date:  2005-08       Impact factor: 4.071

5.  The hysteresis effect in carpal kinematics.

Authors:  Sunjay Berdia; Walter H Short; Frederick W Werner; Jason K Green; Manohar Panjabi
Journal:  J Hand Surg Am       Date:  2006-04       Impact factor: 2.230

6.  Impact of the z-flying focal spot on resolution and artifact behavior for a 64-slice spiral CT scanner.

Authors:  Yiannis Kyriakou; Marc Kachelriess; Michael Knaup; Jens U Krause; Willi A Kalender
Journal:  Eur Radiol       Date:  2006-04-06       Impact factor: 5.315

Review 7.  The natural progression of scaphoid instability.

Authors:  H K Watson; J Weinzweig; J Zeppieri
Journal:  Hand Clin       Date:  1997-02       Impact factor: 1.907

8.  Four-dimensional computed tomographic imaging in the wrist: proof of feasibility in a cadaveric model.

Authors:  Shian-Chao Tay; Andrew N Primak; Joel G Fletcher; Bernhard Schmidt; Kimberly K Amrami; Richard A Berger; Cynthia H McCollough
Journal:  Skeletal Radiol       Date:  2007-09-06       Impact factor: 2.199

9.  Direct cone-beam cardiac reconstruction algorithm with cardiac banding artifact correction.

Authors:  Katsuyuki Taguchi; Beshan S Chiang; Ilmar A Hein
Journal:  Med Phys       Date:  2006-02       Impact factor: 4.071

Review 10.  CT of coronary artery disease.

Authors:  U Joseph Schoepf; Christoph R Becker; Bernd M Ohnesorge; E Kent Yucel
Journal:  Radiology       Date:  2004-07       Impact factor: 11.105

  10 in total
  5 in total

1.  Dynamic CT technique for assessment of wrist joint instabilities.

Authors:  Shuai Leng; Kristin Zhao; Mingliang Qu; Kai-Nan An; Richard Berger; Cynthia H McCollough
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

2.  Four-dimensional computed tomographic imaging in the wrist: proof of feasibility in a cadaveric model.

Authors:  Shian-Chao Tay; Andrew N Primak; Joel G Fletcher; Bernhard Schmidt; Kimberly K Amrami; Richard A Berger; Cynthia H McCollough
Journal:  Skeletal Radiol       Date:  2007-09-06       Impact factor: 2.199

3.  Improving small animal cone beam CT resolution by mitigating x-ray focal spot induced blurring via deconvolution.

Authors:  Xiaoyu Hu; Yuncheng Zhong; Yanqi Huang; Chenyang Shen; Xun Jia
Journal:  Phys Med Biol       Date:  2022-06-08       Impact factor: 4.174

Review 4.  Four-dimensional computed tomography: musculoskeletal applications.

Authors:  Murray T Wong; Charmaine Wiens; Michael Kuczynski; Sarah Manske; Prism S Schneider
Journal:  Can J Surg       Date:  2022-06-14       Impact factor: 2.840

5.  Evidence-based recommendations for musculoskeletal kinematic 4D-CT studies using wide area-detector scanners: a phantom study with cadaveric correlation.

Authors:  Pedro Augusto Gondim Teixeira; Anne-Sophie Formery; Gabriela Hossu; Daniel Winninger; Toufik Batch; Alban Gervaise; Alain Blum
Journal:  Eur Radiol       Date:  2016-04-19       Impact factor: 5.315

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.