Literature DB >> 18655949

The noisy ceramic hip: is component malpositioning the cause?

Camilo Restrepo1, Javad Parvizi, Steve M Kurtz, Peter F Sharkey, William J Hozack, Richard H Rothman.   

Abstract

Noisy ceramics bearing surfaces are a recently recognized problem in total hip arthroplasty. Component malposition as a potential cause has been proposed. Squeaking occurred in 28 (2.7%) of 999 patients undergoing ceramic on ceramic total hip arthroplasty at our institution. Patients were matched, in a 1:2 ratio by anthropometric and demographic variables and also prosthesis size and type. The acetabular position was measured using radiographs and computed tomography. There was no statistically significant difference in cup inclination (P = .25) or version (P = .38) between groups. Four hips that have been revised were available for retrieval analysis. Stripe wear and metal transfer to ceramic components were observed. Etiology of squeaking ceramic total hip arthroplasty remains elusive. Although malposition could be an important contributing factor, the latter cannot be the sole reason based on our findings. Further investigation to elucidate the etiology is warranted.

Entities:  

Mesh:

Year:  2008        PMID: 18655949     DOI: 10.1016/j.arth.2008.04.001

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  41 in total

1.  A squeaky reputation: the problem may be design-dependent.

Authors:  Javad Parvizi; Bahar Adeli; Justin C Wong; Camilo Restrepo; Richard H Rothman
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

2.  Is metal-on-metal squeaking related to acetabular angle of inclination?

Authors:  Thomas Bernasek; David Fisher; David Dalury; Melissa Levering; Kirk Dimitris
Journal:  Clin Orthop Relat Res       Date:  2011-04-22       Impact factor: 4.176

Review 3.  Surgical management of osteoarthritis.

Authors:  Jeffrey N Katz; Brandon E Earp; Andreas H Gomoll
Journal:  Arthritis Care Res (Hoboken)       Date:  2010-09       Impact factor: 4.794

4.  Squeaking: Current knowledge and how to avoid it.

Authors:  Arjuna M Imbuldeniya; Simon J Pearce; William L Walter; Bernard A Zicat; William K Walter
Journal:  Curr Rev Musculoskelet Med       Date:  2013-12

5.  The 2009 Frank Stinchfield Award: "Hip squeaking": a biomechanical study of ceramic-on-ceramic bearing surfaces.

Authors:  Christophe Chevillotte; Robert T Trousdale; Qingshan Chen; Olivier Guyen; Kai-Nan An
Journal:  Clin Orthop Relat Res       Date:  2009-06-19       Impact factor: 4.176

6.  Delta ceramic-on-alumina ceramic articulation in primary THA: prospective, randomized FDA-IDE study and retrieval analysis.

Authors:  Adolph V Lombardi; Keith R Berend; Brian E Seng; Ian C Clarke; Joanne B Adams
Journal:  Clin Orthop Relat Res       Date:  2010-02       Impact factor: 4.176

7.  In vivo wear of a squeaky alumina-on-alumina hip prosthesis: a case report.

Authors:  Michel P Laurent; Robin Pourzal; Alfons Fischer; Kim C Bertin; Joshua J Jacobs; Markus A Wimmer
Journal:  J Bone Joint Surg Am       Date:  2011-04-06       Impact factor: 5.284

8.  Distinctive damage patterns on THA metal bearing surfaces: case studies.

Authors:  Anneliese D Heiner; Nishant M Tikekar; Karen M Kruger; John J Lannutti; Thomas D Brown
Journal:  Iowa Orthop J       Date:  2014

9.  Incidence of 'squeaking' after ceramic-on-ceramic total hip arthroplasty.

Authors:  Kenny Mai; Christopher Verioti; Kace A Ezzet; Steven N Copp; Richard H Walker; Clifford W Colwell
Journal:  Clin Orthop Relat Res       Date:  2010-02       Impact factor: 4.176

10.  THA with Delta ceramic on ceramic: results of a multicenter investigational device exemption trial.

Authors:  William G Hamilton; James P McAuley; Douglas A Dennis; Jeffrey A Murphy; Thomas J Blumenfeld; Joel Politi
Journal:  Clin Orthop Relat Res       Date:  2009-09-19       Impact factor: 4.176