Literature DB >> 19885712

Delta ceramic-on-alumina ceramic articulation in primary THA: prospective, randomized FDA-IDE study and retrieval analysis.

Adolph V Lombardi1, Keith R Berend, Brian E Seng, Ian C Clarke, Joanne B Adams.   

Abstract

UNLABELLED: Wear and osteolysis continue to be major reasons for revision surgery in THA. Ceramic-on-ceramic bearings eliminate polyethylene wear debris. The newest generation of these bearings incorporate nanosized, yttria-stabilized tetragonal zirconia particles producing an alumina matrix composite. We asked whether this new material would perform as well as a conventional bearing in terms of functional hip scores, radiographic migration and osteolysis, complications and survival. As part of a US FDA investigational device exemption study (G000075), we conducted an initial prospective safety study of 21 alumina matrix composite femoral heads articulating on alumina liners followed by a prospective, randomized study with 44 more of these articulations and 45 zirconia femoral heads on polyethylene liners. The minimum followup for all patients was 26 months (mean, 73 months; range, 26-108 months). Harris hip scores and radiographic findings were similar in the two groups as was survivorship (trial 95% versus control 93%). There were three reoperations in the trial group and three in the control group. A fractured head retrieval showed a 33% monoclinic transformation with an increase in surface roughness from 3 to 5 nm at the main wear zone. While our numbers were insufficient to compare device-related complications, the trial device performed as well as the control device in terms of reoperation, and clinical and radiographic outcome. The alumina matrix composite femoral head on an alumina liner provided high survivorship. LEVEL OF EVIDENCE: Level II, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 19885712      PMCID: PMC2807003          DOI: 10.1007/s11999-009-1143-9

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  52 in total

1.  Phase transformation of a zirconia ceramic head after total hip arthroplasty.

Authors:  K Haraguchi; N Sugano; T Nishii; H Miki; K Oka; H Yoshikawa
Journal:  J Bone Joint Surg Br       Date:  2001-09

2.  Long-term wear of HIPed alumina on alumina bearings for THR under microseparation conditions.

Authors:  T Stewart; J Tipper; R Streicher; E Ingham; J Fisher
Journal:  J Mater Sci Mater Med       Date:  2001 Oct-Dec       Impact factor: 3.896

3.  Cobalt-chromium head wear following revision hip arthroplasty performed after ceramic fracture--a case report.

Authors:  Masahiro Hasegawa; Akihiro Sudo; Atsumasa Uchida
Journal:  Acta Orthop       Date:  2006-10       Impact factor: 3.717

4.  Analysis of prosthetic femoral heads retrieved at autopsy.

Authors:  C J Sychterz; C A Engh; S W Swope; D E McNulty; C A Engh
Journal:  Clin Orthop Relat Res       Date:  1999-01       Impact factor: 4.176

5.  "Modes of failure" of cemented stem-type femoral components: a radiographic analysis of loosening.

Authors:  T A Gruen; G M McNeice; H C Amstutz
Journal:  Clin Orthop Relat Res       Date:  1979-06       Impact factor: 4.176

Review 6.  Ceramic femoral head retrieval data.

Authors:  G Willmann
Journal:  Clin Orthop Relat Res       Date:  2000-10       Impact factor: 4.176

Review 7.  Review on ceramic-on-ceramic total hip arthroplasty.

Authors:  S Mehmood; Riyaz H Jinnah; Hemant Pandit
Journal:  J Surg Orthop Adv       Date:  2008

8.  The squeaking hip: a cause for concern-disagrees.

Authors:  Charlie C Yang; Raymond H Kim; Douglas A Dennis
Journal:  Orthopedics       Date:  2007-09       Impact factor: 1.390

9.  Squeaking in ceramic-on-ceramic hips: the importance of acetabular component orientation.

Authors:  William L Walter; Gary C O'toole; William K Walter; Andrew Ellis; Bernard A Zicat
Journal:  J Arthroplasty       Date:  2007-03-28       Impact factor: 4.757

10.  Third-generation alumina-on-alumina ceramic bearings in cementless total hip arthroplasty.

Authors:  P J Lusty; C C Tai; R P Sew-Hoy; W L Walter; W K Walter; B A Zicat
Journal:  J Bone Joint Surg Am       Date:  2007-12       Impact factor: 5.284

View more
  19 in total

1.  Causes of failure of ceramic-on-ceramic and metal-on-metal hip arthroplasties.

Authors:  Manny Porat; Javad Parvizi; Peter F Sharkey; Keith R Berend; Adolph V Lombardi; Robert L Barrack
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

Review 2.  Survival of hard-on-hard bearings in total hip arthroplasty: a systematic review.

Authors:  Michael G Zywiel; Siraj A Sayeed; Aaron J Johnson; Thomas P Schmalzried; Michael A Mont
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

3.  Hard-on-Hard Bearings Are Associated With Increased Noise Generation in Young Patients Undergoing Hip Arthroplasty.

Authors:  Denis Nam; Toby Barrack; Staci R Johnson; Ryan M Nunley; Robert L Barrack
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

Review 4.  Hip replacement in the athlete: is there a role?

Authors:  M J Wilson; R N Villar
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-04-12       Impact factor: 4.342

Review 5.  [Evidence-based update in hip arthroplasty].

Authors:  H Gollwitzer; L Gerdesmeyer; R Gradinger; R von Eisenhart-Rothe
Journal:  Orthopade       Date:  2011-06       Impact factor: 1.087

6.  Comparative effectiveness of ceramic-on-ceramic implants in stemmed hip replacement: a multinational study of six national and regional registries.

Authors:  Art Sedrakyan; Stephen Graves; Barbara Bordini; Miquel Pons; Leif Havelin; Susan Mehle; Elizabeth Paxton; Thomas Barber; Guy Cafri
Journal:  J Bone Joint Surg Am       Date:  2014-12-17       Impact factor: 5.284

7.  Squeaking and microcracks in a delta-delta ceramic coupling: pin-on-disc study.

Authors:  Kiyokazu Fukui; Ayumi Kaneuji; Tadami Matsumoto; Kazuhiro Shintani
Journal:  Eur J Orthop Surg Traumatol       Date:  2016-03-12

Review 8.  Advances in zirconia toughened alumina biomaterials for total joint replacement.

Authors:  Steven M Kurtz; Sevi Kocagöz; Christina Arnholt; Roland Huet; Masaru Ueno; William L Walter
Journal:  J Mech Behav Biomed Mater       Date:  2013-04-10

9.  Does acetabular inclination angle affect survivorship of alumina-ceramic articulations?

Authors:  Henning R Johansson; Aaron J Johnson; Michael G Zywiel; Marybeth Naughton; Michael A Mont; Peter M Bonutti
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

10.  Is there any difference in survivorship of total hip arthroplasty with different bearing surfaces? A systematic review and network meta-analysis.

Authors:  Si Yin; Dangfeng Zhang; Hui Du; Heng Du; Zhanhai Yin; Yusheng Qiu
Journal:  Int J Clin Exp Med       Date:  2015-11-15
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.