64-MDCT coronary angiography: phantom study of effects of vascular attenuation on detection of coronary stenosis.

Xiaolu Fei1, Xiangying Du, Qi Yang, Yun Shen, Pengyu Li, Jingmin Liao, Kuncheng Li.   

Abstract

OBJECTIVE: The purpose of this study was to investigate the effects of vascular attenuation on the accuracy of stenosis evaluation with 64-MDCT coronary angiography.
MATERIALS AND METHODS: A pulsating cardiac phantom was used to simulate the beating heart and coronary arteries of 5 and 3 mm in diameter with three degrees of stenosis (25%, 50%, and 75%) at a heart rate of 55 beats per minute. Coronary vascular enhancement had four attenuation levels: low, 200 H; moderately low, 300 H; moderately high, 350 H; and high, 500 H. Cardiac scans were obtained with 64-MDCT. Percentage stenosis, plaque area, and plaque density were measured on axial images.
RESULTS: For 50% and 75% stenosis in 5-mm vessels, there were no significant differences among the four attenuation groups. For 50% and 75% stenosis in 3-mm vessels, significant underestimation of percentage stenosis occurred in the high-attenuation group compared with the moderate- and low-attenuation groups (p < 0.05). For 25% stenosis in 5-mm vessels, low attenuation led to significant overestimation of degree of stenosis compared with the moderate and high attenuation levels (p < 0.05). None of the instances of 25% stenosis in 3-mm vessels were detected in the high-attenuation group. Underestimation was found only for 3-mm vessels. For 75% stenosis, all plaques were detected irrespective of contrast attenuation and vessel size.
CONCLUSION: Use of higher attenuation leads to a significant underestimation of stenosis in smaller vessels. Lower attenuation leads to slight and clinically acceptable overestimation of stenosis. The optimal vascular attenuation for stenosis detection in coronary 64-MDCT angiography is approximately 350 H.

Entities:  

Mesh:

Year:  2008        PMID: 18562723     DOI: 10.2214/AJR.07.2653

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  33 in total

1.  An objective evaluation method designed for pulsating cardiac phantom with 64-row MDCT.

Authors:  Xiaolu Fei; Xiangying Du; Mei Bai; Yan Li; Pengyu Li; Lan Wei; Kuncheng Li
Journal:  J Med Syst       Date:  2010-06       Impact factor: 4.460

2.  Patient-related factors influencing detectability of coronary arteries in 320-row CT angiography in infants with complex congenital heart disease.

Authors:  Yuzo Yamasaki; Satoshi Kawanami; Takeshi Kamitani; Koji Sagiyama; Seitaro Shin; Takuya Hino; Hazumu Nagata; Hidetake Yabuuchi; Michinobu Nagao; Hiroshi Honda
Journal:  Int J Cardiovasc Imaging       Date:  2018-05-05       Impact factor: 2.357

3.  Low-dose contrast protocol using the test bolus technique for 64-detector computed tomography coronary angiography.

Authors:  Takeshi Nakaura; Kazuo Awai; Yumi Yanaga; Tomohiro Namimoto; Daisuke Utsunomiya; Toshinori Hirai; Seigo Sugiyama; Hisao Ogawa; Masahito Aoyama; Yasuyuki Yamashita
Journal:  Jpn J Radiol       Date:  2011-09-01       Impact factor: 2.374

4.  Relationship between diverse patient body size- and image acquisition-related factors, and quantitative and qualitative image quality in coronary computed tomography angiography: a multicenter observational study.

Authors:  Daisuke Utsunomiya; Ryoichi Tanaka; Kunihiro Yoshioka; Kazuo Awai; Teruhito Mochizuki; Naofumi Matsunaga; Tomoaki Ichikawa; Masayuki Kanematsu; Tonsok Kim; Yasuyuki Yamashita
Journal:  Jpn J Radiol       Date:  2016-06-07       Impact factor: 2.374

5.  A low-dose and an ultra-low-dose contrast agent protocol for coronary CT angiography in a clinical setting: quantitative and qualitative comparison to a standard dose protocol.

Authors:  Dominik C Benz; Christoph Gräni; Beatrice Hirt Moch; Fran Mikulicic; Jan Vontobel; Tobias A Fuchs; Julia Stehli; Olivier F Clerc; Mathias Possner; Aju P Pazhenkottil; Oliver Gaemperli; Ronny R Buechel; Philipp A Kaufmann
Journal:  Br J Radiol       Date:  2017-05-25       Impact factor: 3.039

6.  Non-calcified coronary atherosclerotic plaque visualization on CT: effects of contrast-enhancement and lipid-content fractions.

Authors:  Wisnumurti Kristanto; Peter M A van Ooijen; Marcel J W Greuter; Jaap M Groen; Rozemarijn Vliegenthart; Matthijs Oudkerk
Journal:  Int J Cardiovasc Imaging       Date:  2013-01-17       Impact factor: 2.357

7.  Coronary computed tomography angiography using ultra-low-dose contrast media: radiation dose and image quality.

Authors:  Sei Komatsu; Teruaki Kamata; Atsuko Imai; Tomoki Ohara; Mitsuhiko Takewa; Ryoko Ohe; Kazuaki Miyaji; Junichi Yoshida; Kazuhisa Kodama
Journal:  Int J Cardiovasc Imaging       Date:  2013-02-26       Impact factor: 2.357

8.  Contrast agent bolus tracking with a fixed threshold or a manual fast start for coronary CT angiography.

Authors:  Fabian Stenzel; Matthias Rief; Elke Zimmermann; Johannes Greupner; Felicitas Richter; Marc Dewey
Journal:  Eur Radiol       Date:  2014-03-26       Impact factor: 5.315

9.  Feasibility of prospectively ECG-triggered high-pitch coronary CT angiography with 30 mL iodinated contrast agent at 70 kVp: initial experience.

Authors:  Long Jiang Zhang; Li Qi; Jing Wang; Chun Xiang Tang; Chang Sheng Zhou; Xue Man Ji; James V Spearman; Carlo Nicola De Cecco; Felix G Meinel; U Joseph Schoepf; Guang Ming Lu
Journal:  Eur Radiol       Date:  2014-04-16       Impact factor: 5.315

10.  CT imaging before transcatheter aortic valve implantation (TAVI) using variable helical pitch scanning and its diagnostic performance for coronary artery disease.

Authors:  Shunsuke Matsumoto; Yoshitake Yamada; Masahiro Hashimoto; Teppei Okamura; Minoru Yamada; Fumiaki Yashima; Kentaro Hayashida; Keiichi Fukuda; Masahiro Jinzaki
Journal:  Eur Radiol       Date:  2016-08-25       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.