| Literature DB >> 18534457 |
Chen-Kun Liaw1, Rong-Sen Yang, Sheng-Mou Hou, Tai-Yin Wu, Chiou-Shann Fuh.
Abstract
Widmer (J Arthroplasty 2004;19:387) reported a protractor for measuring the anteversion of acetabular cups on radiographs but with limited precision. We intended to improve its precision by trigonometric mathematics. We measured the anteversion of the acetabular cups on 336 simulated radiographs using aforementioned 2 methods. The anteversion measured by Widmer's protractor ranged from 7 degrees to 41 degrees (mean +/- SD = 28.0 degrees +/- 9.8 degrees), and our methods, 5 degrees to 51 degrees (27.7 degrees +/- 13.2 degrees). The mean +/- SD of error by Widmer's protractor was 5.2 +/- 2.5 degrees, and our protractor, 0.8 degrees +/- 0.8 degrees (Student t test, P b .0001). The interobserver study showed the difference between measurements less than 2 degrees for each method. Therefore, the smaller error of our method than that of Widmer implicated a potentially precise measurement of the anteversion (level of evidence: diagnostic study, level II).Mesh:
Year: 2008 PMID: 18534457 DOI: 10.1016/j.arth.2007.10.029
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757