Improvement in the assessment of wear of total knee replacements using coordinate-measuring machine techniques.

L A Blunt1, P J Bills, X-Q Jiang, G Chakrabarty.   

Abstract

Total joint replacement is one of the most common elective surgical procedures performed worldwide, with an estimate of 1.5x 10(6) operations performed annually. Currently joint replacements are expected to function for 10-15 years; however, with an increase in life expectancy, and a greater call for knee replacement due to increased activity levels, there is a requirement to improve their function to offer longer-term improved quality of life for patients. Wear analysis of total joint replacements has long been an important means in determining failure mechanisms and improving longevity of these devices. The effectiveness of the coordinate-measuring machine (CMM) technique for assessing volumetric material loss during simulated life testing of a replacement knee joint has been proved previously by the present authors. The purpose of the current work is to present an improvement to this method for situations where no pre-wear data are available. To validate the method, simulator tests were run and gravimetric measurements taken throughout the test, such that the components measured had a known wear value. The implications of the results are then discussed in terms of assessment of joint functionality and development of standardized CMM-based product standards. The method was then expanded to allow assessment of clinically retrieved bearings so as to ascertain a measure of true clinical wear.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18491700     DOI: 10.1243/09544119JEIM289

Source DB:  PubMed          Journal:  Proc Inst Mech Eng H        ISSN: 0954-4119            Impact factor:   1.617


  7 in total

1.  In vitro quantification of wear in tibial inserts using microcomputed tomography.

Authors:  Matthew G Teeter; Douglas D R Naudie; David D McErlain; Jan-M Brandt; Xunhua Yuan; Steven J Macdonald; David W Holdsworth
Journal:  Clin Orthop Relat Res       Date:  2011-01       Impact factor: 4.176

2.  How do CAD models compare with reverse engineered manufactured components for use in wear analysis?

Authors:  Matthew G Teeter; Douglas D R Naudie; Robert B Bourne; David W Holdsworth
Journal:  Clin Orthop Relat Res       Date:  2012-07       Impact factor: 4.176

3.  A pictographic atlas for classifying damage modes on polyethylene bearings.

Authors:  Melinda Harman; Luca Cristofolini; Paolo Erani; Susanna Stea; Marco Viceconti
Journal:  J Mater Sci Mater Med       Date:  2011-04-02       Impact factor: 3.896

4.  Relationship of surface damage appearance and volumetric wear in retrieved TKR polyethylene liners.

Authors:  Christopher B Knowlton; Priyanka Bhutani; Markus A Wimmer
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2016-07-12       Impact factor: 3.368

5.  Long-term wear of dual mobility total hip replacement cups: explant study.

Authors:  Bertrand Boyer; Thomas Neri; Jean Geringer; Alexandre Di Iorio; Remi Philippot; Frederic Farizon
Journal:  Int Orthop       Date:  2017-06-03       Impact factor: 3.075

6.  An autonomous mathematical reconstruction to effectively measure volume loss on retrieved polyethylene tibial inserts.

Authors:  Christopher B Knowlton; Markus A Wimmer
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2012-08-22       Impact factor: 3.368

7.  CMM-Based Volumetric Assessment Methodology for Polyethylene Tibial Knee Inserts in Total Knee Replacement.

Authors:  Wei Jiang; Cuicui Ji; Zhongmin Jin; Yuntian Dai
Journal:  Appl Bionics Biomech       Date:  2018-04-10       Impact factor: 1.781

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.