Bruno Zuberbuhler1, Theo Signer, Richard Gale, Eduard Haefliger.
Abstract
BACKGROUND: To evaluate the rotational stability of the three types of AcrySof SA60TT toric intraocular lenses (Alcon, Switzerland) in cataract surgery after the first postoperative week.Entities:
Mesh:
Year: 2008 PMID: 18460196 PMCID: PMC2408563 DOI: 10.1186/1471-2415-8-8
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Patient demographics, best-corrected distance visual acuity, refractive astigmatism and corneal astigmatism before and three months after implantation of a AcrySof SA60TT toric intraocular lens (SA60T3, SA60T4 or SA60T5 type).
| Parameters | SA60T3 | SA60T4 | SA60T5 | Summarized |
| Eyes (n) | 7 | 17 | 20 | 44 |
| Patients (n) | 7 | 14 | 17 | 33 |
| Age (y) | ||||
| Mean ± SD | 66 ± 9 | 71 ± 14 | 72 ± 13 | 71 ± 13 |
| Range | 52 to 76 | 35 to 88 | 35 to 86 | 35 to 88 |
| Preoperative Parameters | ||||
| Mean BCDVA (LogMAR) ± SD | 0.28 ± 0.22 | 0.27 ± 0.15 | 0.30 ± 0.17 | 0.28 ± 0.17 |
| Mean SE (D) ± SD | -1.90 ± 3.27 | -1.60 ± 3.38 | -1.50 ± 3.15 | -1.60 ± 3.19 |
| Mean refractive astigmatism (D) ± SD* | 1.86 ± 0.83 | 2.24 ± 0.83 | 2.98 ± 1.07 | - |
| Mean corneal astigmatism (D) ± SD† | 1.40 ± 0.55 | 1.95 ± 0.56 | 3.06 ± 0.74 | - |
| Postoperative Parameters | ||||
| Mean BCDVA (LogMAR) ± SD | 0.00 ± 0.06 | -0.02 ± 0.04 | 0.05 ± 0.15 | 0.01 ± 0.11 |
| Mean SE (D) ± SD | -0.75 ± 0.95 | -0.80 ± 1.44 | -0.38 ± 1.26 | -0.61 ± 1.28 |
| Mean refractive astigmatism (D) ± SD* | 0.43 ± 0.59 | 0.22 ± 0.26 | 0.89 ± 1.29 | - |
| Mean corneal astigmatism (D) ± SD† | 1.54 ± 0.51 | 1.85 ± 0.50 | 3.19 ± 0.93 | - |
BCDVA = best-corrected distance visual acuity; SE = spherical equivalent
*The change in refractive astigmatism was statistically significant due to the lens implantation (SA60T3 p = 0.04; SA60T4 p < 0.001; SA60T5 p < 0.001).
†The corneal astigmatism did not change significantly due to the surgery (SA60T3 p = 0.17; SA60T4 p = 0.98; SA60T5 p = 0.26).
Figure 1Rotation of the AcrySof SA60TT toric intraocular lens between the one week and three months postoperative follow-up.
Absolute rotation and direction of rotation of the AcrySof SA60TT toric intraocular lenses between the one week and three months follow-up after implantation in standard cataract surgery.
| Parameters | SA60T3 | SA60T4 | SA60T5 | Summarized |
| Eyes (n) | 7 | 17 | 20 | 44 |
| Absolute rotation (degrees) | ||||
| Mean ± SD | 1.6 ± 1.0* | 3.3 ± 2.4* | 1.5 ± 2.0* | 2.2 ± 2.2 |
| Range | 0 to 3 | 0 to 9 | 0 to 7 | 0 to 9 |
| Direction of rotation, n (%) | ||||
| Clockwise | 4 (57%) | 7 (41%) | 5 (25%) | 16 (36%) |
| Anti-clockwise | 2 (29%) | 7 (41%) | 7 (35%) | 16 (36%) |
| No rotation | 1 (14%) | 3 (18%) | 8 (40%) | 12 (28%) |
*Statistically, no difference in absolute rotation was noticed between the three types (p > 0.18).