Evaluation of a low-dose/slow-rotating SPECT-CT system.

M Hamann1, M Aldridge, J Dickson, R Endozo, K Lozhkin, B F Hutton.   

Abstract

The 4-slice CT that forms part of the GE Infinia Hawkeye-4 SPECT-CT scanner (Hawkeye) is evaluated against the diagnostic 16-slice CT that is incorporated in the GE Discovery ST PET-CT system (DST). The x-ray tube of the slow-rotating Hawkeye system (23 s/rotation) operates at approximately a third of the dose of diagnostic systems as used for conventional diagnostic imaging. Image reconstruction is optimized for low noise. High-contrast spatial resolution significantly falls behind diagnostic figures: the average of MTF(50) and MTF(10) (resolution where the MTF has fallen to 50% and 10%) is 2.8 +/- 0.1 cm(-1) for Hawkeye and 5.3 +/- 0.1 cm(-1) for the DST (standard reconstruction filters). Resolution in the direction of the couch movement (z coordinate) is governed by the fixed Hawkeye slice width of 5 mm. Reconstruction accuracy is found to be increased by reducing the default z increment from 4.4 mm to 2.2 mm. Low-contrast object detectability is superior compared with diagnostic systems operating in the Hawkeye dose range. In the diagnostic dose regime, however, small low-contrast details remain visible in DST that are not detectable with Hawkeye. Although not of diagnostic quality, the low-dose Hawkeye provides appropriate data for SPECT attenuation correction and anatomical localization capability.

Entities:  

Mesh:

Year:  2008        PMID: 18424877     DOI: 10.1088/0031-9155/53/10/003

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  9 in total

Review 1.  The origins of SPECT and SPECT/CT.

Authors:  Brian F Hutton
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-11-12       Impact factor: 9.236

2.  Influence of CT-based attenuation correction in assessment of left and right ventricular functions with count-based gated blood-pool SPECT.

Authors:  Louis Sibille; Fayçal Ben Bouallegue; Aurélie Bourdon; Denis Mariano-Goulart
Journal:  J Nucl Cardiol       Date:  2011-06-07       Impact factor: 5.952

3.  Improving perfusion defect detection with respiratory motion correction in cardiac SPECT at standard and reduced doses.

Authors:  Chao Song; Yongyi Yang; Albert Juan Ramon; Miles N Wernick; P Hendrik Pretorius; Karen L Johnson; Piotr J Slomka; Michael A King
Journal:  J Nucl Cardiol       Date:  2018-07-30       Impact factor: 5.952

4.  4D non-local means post-filtering for cardiac gated SPECT.

Authors:  Chao Song; Yongyi Yang; P Hendrik Pretorius; Michael A King
Journal:  Phys Med Biol       Date:  2018-01-31       Impact factor: 3.609

5.  Agreement of visual estimation of coronary artery calcium from low-dose CT attenuation correction scans in hybrid PET/CT and SPECT/CT with standard Agatston score.

Authors:  Andrew J Einstein; Lynne L Johnson; Sabahat Bokhari; Jessica Son; Randall C Thompson; Timothy M Bateman; Sean W Hayes; Daniel S Berman
Journal:  J Am Coll Cardiol       Date:  2010-11-30       Impact factor: 24.094

6.  Evaluation of diagnosis techniques used for spinal injury related back pain.

Authors:  Meaghan Janssen; Aliaa Nabih; Walied Moussa; Gregory N Kawchuk; Jason P Carey
Journal:  Pain Res Treat       Date:  2011-06-13

7.  A study to improve the image quality in low-dose computed tomography (SPECT) using filtration.

Authors:  Sc Kheruka; Uc Naithani; Ak Maurya; Nk Painuly; Lm Aggarwal; S Gambhir
Journal:  Indian J Nucl Med       Date:  2011-01

8.  Evaluation of single-photon emission computed tomography images obtained with and without copper filter by segmentation.

Authors:  Subhash Chand Kheruka; Lalit Mohan Aggarwal; Neeraj Sharma; Umesh Chand Naithani; Anil Kumar Maurya; Sanjay Gambhir
Journal:  Indian J Nucl Med       Date:  2016 Apr-Jun

9.  Characterization of atherosclerotic plaque: a contrast-detail study using multidetector and cone-beam computed tomography.

Authors:  Nima Kasraie; Peter Mah; Carl R Keener; Geoffrey D Clarke
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.