Literature DB >> 18410306

Penile implantation in Europe: successes and complications with 253 implants in Italy and Germany.

Alessandro Natali1, Roberto Olianas, Margit Fisch.   

Abstract

INTRODUCTION: Results for prosthesis implantation from everyday clinical practice within Europe are few. This report provides data on the most commonly used penile prostheses (the American Medical Systems [AMS] series). AIM: The study aimed to assess, retrospectively, complications and patient satisfaction with AMS penile implants in 253 consecutive patients with erectile dysfunction from three European centers.
METHODS: Pre, intra- and postoperative data were obtained from chart review, with a mean follow-up of 60 months; 200 patients were available for evaluation. Patient satisfaction data were collected using the Erectile Dysfunction Inventory of Treatment Satisfaction (EDITS) questionnaire. MAIN OUTCOME MEASURE: Complications and patient satisfaction were assessed. Patient satisfaction was evaluated using a standardized assessment tool (the modified EDITS questionnaire).
RESULTS: Major postoperative complications occurred in 40 (20%) patients, including 9 (22.5%) prosthesis infections, 18 (45%) mechanical failures, and 13 (32.5%) erosions. Patient satisfaction with the AMS 700CX, AMS Ambicor, and AMS 600-650 was 97%, 81%, and 75%, respectively; dissatisfaction was 0%, 5%, and 6%, respectively. Partner satisfaction with the AMS 700CX, AMS Ambicor, and AMS 600-650 was 91%, 91%, and 75%, respectively; dissatisfaction was 0%, 5%, and 6%, respectively. Erections were more natural (harder) than before with the AMS 700CX, AMS Ambicor, and AMS 600-650 in 91%, 85%, and 88%, respectively; hardness was the same as before in 9%, 15%, and 13%, respectively; no erections were less hard than before.
CONCLUSIONS: Postoperative complications differed from those reported in the literature, while patient satisfaction rates were roughly similar. The reporting of specific data for different implant types, plus the use of standardized assessment tools for patient satisfaction is significant as in the future, it will allow comparison of data between centers performing penile prosthesis implants using these devices.

Entities:  

Mesh:

Year:  2008        PMID: 18410306     DOI: 10.1111/j.1743-6109.2008.00819.x

Source DB:  PubMed          Journal:  J Sex Med        ISSN: 1743-6095            Impact factor:   3.802


  27 in total

Review 1.  A practical overview of considerations for penile prosthesis placement.

Authors:  Landon Trost; Philip Wanzek; George Bailey
Journal:  Nat Rev Urol       Date:  2015-12-01       Impact factor: 14.432

2.  Comparison of infrapubic versus transcrotal approaches for inflatable penile prosthesis placement: a multi-institution report.

Authors:  L W Trost; A G Boonjindasup; W J G Hellstrom
Journal:  Int J Impot Res       Date:  2014-10-23       Impact factor: 2.896

3.  Long-term outcomes of AMS Spectra® penile prosthesis implantation and satisfaction rates.

Authors:  F Akdemir; E Okulu; Ö Kayıgil
Journal:  Int J Impot Res       Date:  2017-04-20       Impact factor: 2.896

4.  Satisfaction assessment with malleable prosthetic implant of Spectra (AMS) and Genesis (Coloplast) models.

Authors:  A R Casabé; N Sarotto; C Gutierrez; A J Bechara
Journal:  Int J Impot Res       Date:  2016-08-25       Impact factor: 2.896

5.  Conservative Therapy is an Effective Option in Patients With Localized Infection After Penile Implant Surgery.

Authors:  Mohamad Habous; Mohammed Farag; Ben Williamson; Osama Laban; Saad Mahmoud; Osama Abdelwahab; Mohamed Elkhouly; Usama Kamil; Saleh Binsaleh; Raanan Tal; David Ralph; John P Mulhall
Journal:  J Sex Med       Date:  2016-05-06       Impact factor: 3.802

6.  Penile prostheses.

Authors:  Carlo Bettocchi; Fabrizio Palumbo; Marco Spilotros; Silvano Palazzo; Gabriele A Saracino; Pasquale Martino; Michele Battaglia; Francesco P Selvaggi; Pasquale Ditonno
Journal:  Ther Adv Urol       Date:  2010-02

7.  Neutrophil to Lymphocyte Ratio as a predictor of early penile prosthesis implant infection.

Authors:  Deniz Bolat; Yusuf Kadir Topcu; Ozgu Aydogdu; Suleyman Minareci; Cetin Dincel
Journal:  Int Urol Nephrol       Date:  2017-03-16       Impact factor: 2.370

8.  Assessment of erectile and ejaculatory function after penile prosthesis implantation.

Authors:  Jang Ho Bae; Phil Hyun Song; Hyun Tae Kim; Ki Hak Moon
Journal:  Korean J Urol       Date:  2010-03-19

9.  Comparison of patient satisfaction rates for the malleable and two piece-inflatable penile prostheses.

Authors:  Hakan Kılıçarslan; Yurdaer Kaynak; Kaan Gökcen; Burhan Coşkun; Onur Kaygısız
Journal:  Turk J Urol       Date:  2014-12

10.  Long-term sequelae of inflatable penile prosthesis: clinical characteristics of patients who develop distal cylinder tip extrusion.

Authors:  Jorge L Fuentes; Yooni A Yi; Michael T Davenport; Rachel L Bergeson; Ellen E Ward; Allen F Morey
Journal:  Transl Androl Urol       Date:  2020-02
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.