Flow impairment during protected carotid artery stenting: impact of filter device design.

Marco Roffi1, Matthias Greutmann, Urs Schwarz, Thomas F Luscher, Franz R Eberli, Beatrice Amann-Vesti.   

Abstract

PURPOSE: To investigate the impact of filter design on blood flow impairment in the internal carotid artery (ICA) among patients undergoing carotid artery stenting (CAS) using filter-type emboli protection devices (EPD).
METHODS: Between July 2003 and March 2007, 115 filter-protected CAS procedures were performed at an academic institution in 107 consecutive patients (78 men; mean age 68 years, range 38-87). The Angioguard, FilterWire EZ, and Spider filters were used in 68 (59%), 32 (28%), and 15 (13%) of cases, respectively. Patient characteristics, procedural and angiographic data, and outcomes were prospectively entered into an electronic database and reviewed retrospectively along with all angiograms.
RESULTS: Flow impairment while the filter was in place was observed in 25 (22%) cases. The presumptive reason of flow impairment was filter obstruction in 21 (18%) instances and flow-limiting spasm at the level of the filter in 4 (4%). In all cases, flow was restored after retrieval of the filter. Flow obstruction in the ICA occurred more frequently with Angioguard (22/68; 32.3%) than with FilterWire EZ (2/32; 6.2%) or Spider (1/15; 6.7%; p = 0.004). No flow occurred in 13 (19%) procedures, all of them protected with Angioguard; no patient treated with other devices experienced this event (p = 0.007). Two (8.0%) strokes occurred in procedures associated with flow impairment, while 1 (1.1%) event was observed in the presence of preserved flow throughout the intervention (p = 0.11).
CONCLUSION: Flow impairment in the ICA during filter-based CAS is common and related to the type of filter used.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18254671     DOI: 10.1583/07-2252.1

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  9 in total

1.  Blood flow changes caused by distal filter protection and catheter aspiration in the internal carotid artery during carotid stenting: evaluation using carotid Doppler sonography.

Authors:  T Sorimachi; K Nishino; K Morita; S Takeuchi; Y Ito; Y Fujii
Journal:  AJNR Am J Neuroradiol       Date:  2010-11-04       Impact factor: 3.825

2.  Computational modeling of distal protection filters.

Authors:  Gail M Siewiorek; Ender A Finol
Journal:  J Endovasc Ther       Date:  2010-12       Impact factor: 3.487

Review 3.  The Angioguard embolic protection device.

Authors:  Gail M Siewiorek; Mark K Eskandari; Ender A Finol
Journal:  Expert Rev Med Devices       Date:  2008-05       Impact factor: 3.166

4.  Carotid angioplasty and stenting without protection devices: safety and efficacy concerns--single center experience.

Authors:  O Y Mansour; J Weber; W Niesen; M Schumacher; A Berlis
Journal:  Clin Neuroradiol       Date:  2011-05-12       Impact factor: 3.649

5.  The association of clinical variables and filter design with carotid artery stenting thirty-day outcome.

Authors:  G M Siewiorek; R T Krafty; M H Wholey; E A Finol
Journal:  Eur J Vasc Endovasc Surg       Date:  2011-04-29       Impact factor: 7.069

6.  Distal protection filter device efficacy with carotid artery stenting: comparison between a distal protection filter and a distal protection balloon.

Authors:  Minoru Iko; Masanori Tsutsumi; Hiroshi Aikawa; Yoshihisa Matsumoto; Yoshinori Go; Kouhei Nii; Gorou Abe; Iwae Ye; Yasuyuki Nomoto; Kiyoshi Kazekawa
Journal:  Jpn J Radiol       Date:  2012-11-20       Impact factor: 2.374

7.  Treatment outcomes of carotid artery stenting with two types of distal protection filter device.

Authors:  Minoru Iko; Hiroshi Aikawa; Yoshinori Go; Kanji Nakai; Masanori Tsutsumi; Iwae Yu; Taichiro Mizokami; Kimiya Sakamoto; Ritsuro Inoue; Takafumi Mitsutake; Ayumu Eto; Hayatsura Hanada; Kiyoshi Kazekawa
Journal:  Springerplus       Date:  2014-03-08

Review 8.  Selective-versus-Standard Poststent Dilation for Carotid Artery Disease: A Systematic Review and Meta-Analysis.

Authors:  O Petr; W Brinjikji; M H Murad; B Glodny; G Lanzino
Journal:  AJNR Am J Neuroradiol       Date:  2017-03-16       Impact factor: 3.825

9.  Comparison of Flow Impairment during Carotid Artery Stenting Using Two Types of Eccentric Filter Embolic Protection Devices.

Authors:  Kouhei Nii; Masanori Tsutsumi; Hitoshi Maeda; Hiroshi Aikawa; Ritsuro Inoue; Ayumu Eto; Kimiya Sakamoto; Takafumi Mitsutake; Hayatsura Hanada; Kiyoshi Kazekawa
Journal:  Neurol Med Chir (Tokyo)       Date:  2016-06-16       Impact factor: 1.742

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.